satron VCT Calibration Manual page 2

Advertisement

ENTER:
Press ENTER to move to lower level in a menu or to accept a com-
mand or parameter value.
UP ARROW:
Use the UP arrow key to move up on the current menu level or to
increase the selected parameter value.
DOWN ARROW:
Use the DOWN arrow key to move down on the current menu level or
to decrease the selected parameter value.
ESC:
Press the ESC to move back/higher in menu or cancel current action.
Basic settings
press the ESC- button to
enter the menu
select CONFIGURA-
TION and press the
ENTER-button
press the [
] -button and
select MAOUTPUT and
press the ENTER-button
select LRV (mA-output
lower range value 4mA)
and press the ENTER-but-
ton
Place the decimal separa-
tor with the [ ][ ] and
press the ENTER-button
insert lower range value
g
g
g
(4mA) with the[
][ ]
and ENTER- buttons and
press the ENTER-button,
until upper separator
reaches the right end
of display. press the
ENTER-button to store
the lower range value for
mA-output
Collect sample
press the [
]-button and
select URV (mA-out upper
range value 20mA) and
press the ENTER-button
Place the decimal separa-
tor with [
] [
and
press the ENTER-button
insert upper range value
(20mA) with the[ ][ ]
and ENTER- buttons and
press the ENTER-button,
until upper separator
reaches the right end
of display. press the
ENTER-button to store
the upper range value for
mA-output
press the [
] -button
and select DAMPING
(time constant for mA-out
damping) and press the
ENTER-button
Laboratory
values
press the ESC- button to
press the [
select CALIBRATION and
press the ENTER-button
set the time constant with
p
p
p
p
press the [
p
select SAMPLES and press
the [
] [
]-buttons
and press the ENTER
press the ENTER-button
to store the time constant
for mA-output damping
select witht he [
buttons the desired sample
press the [
]-button and
point to which laboratory
select ALARMTYP and
value will be inserted and
press the ENTER-button
press the ENTER-button
Place the decimal separa-
tor with [
press the ENTER button
Set alarm current with the
insert the laboratory value
[
] [
]-buttons (3.7
or 22.5mA) and press the
with the [
ENTER-button
ENTER- buttons and press
the ENTER-button, until
press the ENTER-button
upper separator reaches
to store the alarm current
value
press the ESC-button
to return to the main
measuring screen
Start-up cali-
2-calibration with water
bration
and one sample point
press the ESC- button to
press the ESC- button to
press the [
enter the menu
select CALIBRATION and
press the ENTER-button
press the [
] -button and
press the [
select NEWSAMPLE and
select CALIBRATE and
press the ENTER-button
press the ENTER-button
press the [
select the SAMPLEH2O
to be the 1st calibration
press the ENTER-button
to activate sampling
the screeen will blink
SAMPLING text during
sampling process. Press
the ENTER-button when
sample has been taken to
end sampling
(SAMPL 01..09) and press
the sample time stamp,
ENTER-button.the display
average and min and max
rolls the new calculated
cs-values during the sam-
pling process are shwon on
press ENTER to store val-
display. Press ENTER to
ues or press ESC to cacel
store the sample or press
press ESC-button to return
ESC to cancel
1-Point
calibration
press the ESC- button to
enter the menu
enter the menu
] -button and
press the [
select CALIBRATION and
press the ENTER-button
press the [
select CALIBRATE and
press the ENTER-button
] -button and
press the [
and select the sample
the ENTER-button
point (SAMPL 01...09) for
1-point calibration and
press the ENTER-button
] [
]
press ESC-button (1-point
calibration, no second
point). the display rolls the
new calculated OFFSET,
GAIN values. press
ENTER-button to store
ESC-button to cacel.
] [
and
press ESC-button to return
to the main measuring
screen
y
] [
] and
the right end of display
press the ESC-button
to return to the main
measuring screen
2-Point
2-point calibration with
calibration
two sample points
press the ESC- button to
enter the menu
press the [
enter the menu
select CALIBRATION and
press the ENTER-button
] -button and
press the [
select CALIBRATE and
press the ENTER-button
] -button and
select with the [ ][ ]
-buttons the fi rst sample
point for calibration
(SAMPL 01...09) and press
the ENTER-button
]-button and
point and press the
ENTER-button
select with the [
] -buttons the second
sample point for calibra-
tion (SAMPL 01...09) and
press the ENTER-button.
press the[
] button
the display rolls the new
and select the second
calculated OFFSET, GAIN
point for calibration
values. press ENTER-but-
ton to store ESC-button
to cacel.
press ESC-button to return
OFFSET, GAIN values.
to measuring screen
to measuring screen
OFFSET
adjustment
press the ESC- button to
enter the menu
] -button and
press the [
] -button and
select CALIBRATION and
press the ENTER-button
] -button and
select RECIPE and press
]-button
the ENTER-button
select OFFSET and press
the ENTER-button
place the decimal sepa-
rator with the [
buttons and press ENTER
insert the OFFSET value
with the [
] [
ENTER- buttons and press
the ENTER-button, until
upper separator reaches
the right end of display
press the ENTER-button
to store the OFFSET value
press ESC-button to return
to main measuring screen
GAIN
adjustment
press the ESC- button to
] -button and
enter the menu
] -button and
press the [
] -button and
select CALIBRATION and
press the ENTER-button
select RECIPE and press
the ENTER-button
][
press the [
and select GAIN and press
the ENTER-button
place the decimal sepa-
rator with the [
buttons and the ENTER
insert the GAIN value
with the [
] [
] and
ENTER- buttons and press
the ENTER-button, until
upper separator reaches
the right end of display
press the ENTER-button
to store the GAIN value
press ESC-button to return
to main measuring screen
TIME and
p
DATE settings
] [
]
press the ESC- button to
enter the menu
] and
select CONFIGURA-
TION and press the
ENTER-button
press the [
] -button
and select SYSTCONF and
press the ENTER-button
press the [
] -button and
select SETCLOCK and
press the ENTER-button
insert date with [ ][
]
-buttons (dd.mm.yyyy),
press ENTER-button to
move from dd->mm-
>yyyy and press the
ENTER-button
insert time with [
][
]
-buttons (hh.mm.ss), press
ENTER-button to move
hh->mm->ss and press the
ENTER-button
press ESC-button to return
] -button
to main measuring screen
p
] [
]