Honeywell 705 Series Operating Instructions Manual page 10

H2s sensor and transmitter for electrochemical cells ul certified
Hide thumbs Also See for 705 Series:
Table of Contents

Advertisement

MAN0362.P65 Issue 03, Mar 01
00705M5001
1.3.5 Gassing Point Assembly
The Gassing Point Assembl assembly can be fixed in a
convenient position and permanently connected by suitable
tubing to an inaccessible sensor, thus simplifying the
application of test gas when checking sensor calibration. A
DIN Rail Mounting Assembly (00785-A-0069) is available to
enable five gassing point assemblies to be mounted side by
side.
1.3.6 Hydrophobic Barrier
The Hydrophobic Barrier consists of a layer of stainless
steel mesh with a rubber rim. A hydrophobic barrier of PTFE
is bonded to the stainless steel surface.
This MUST be fitted with the white surface faceing outwards.
1. INTRODUCTION
Stainless Steel
Tube
120 mm Dia
Gassing Tube Cap
Figure 5 Gassing Point Assembly
11
705 Sensor
25mm
Yellow Nylon Block
70mm x 38mm x 19mm

Advertisement