ResMed S8 AutoSet Spirit II Clinical Manual page 32

Hide thumbs Also See for S8 AutoSet Spirit II:
Table of Contents

Advertisement

Menu
Parameter
Erase Data?
Date
Time
Menu Type
Language*
Servicing
Run Hours*
SN*
PCB*
SW*
BR
*These parameters may appear in one or both patient menus and the clinical menu.
Note: When Leak Alert is set to ON, SmartStart/Stop is automatically set to
OFF. SmartStart/Stop cannot be used with Leak Alert because, if a high leak
occurs, SmartStop may stop treatment before the Leak Alert signal is activated.
Settings for Mask Types
The following table shows the setting that should be selected for each mask
type.
Settings
ULTRA
MIR FULL
ACTIVA
SWIFT
Description
Allows the clinician to erase all data stored in the
device (except for Run Hours).
Settings are not affected.
Sets the current date.
Sets the current time.
Sets the menu that the patient will see: standard
(STD) or detailed (ADV).
Sets the display language.
Displays the total number of machine hours.
(View only)
Displays the device serial number.
(View only)
Displays the printed circuit board number.
(View only)
Displays the current software version.
Sets the baud rate. This value is reset to 19200
every time the S8 AutoSet Spirit II is powered up.
Mask
Ultra Mirage II Nasal Mask
Mirage Liberty Full Face Mask
Mirage Quattro Full Face Mask
Ultra Mirage Full Face Mask
Mirage Activa Nasal Mask
Mirage Swift Nasal Pillows System
Mirage Swift II Nasal Pillows System
Navigating the S8 AutoSet Spirit II Menus
26