Error Number
154
155
156
157
158
159
160
161
162
163
164
165
166
167
168
169
170
171
172
173
174
175
176
177
178
179
K0553 Revision D
Pump monitoring has failed when pumps commanded ON
Pump monitoring has failed when pumps commanded OFF
Pump monitoring has failed when pressure pump commanded ON
Pump monitoring has failed when vacuum pump commanded ON
Pump monitoring has failed when heater commanded ON
Pump Timer Failed
Invalid PWM Channel
QSPI LISR Failed
SCI LISR Failed
RAM validation has failed
Driver Read
RPT diode out of range
ALTERA count = 0, indicating RPT not connected or failed
ALTERA Not complete, indicating RPT input frequency too low
ALTERA overflowed, indicating RPT input frequency too high
RPT count is out of range at 1000mB
Invalid RTC Data
RTC Null Pointer
Real time clock is not incrementing
Real time clock parameter RAM failed
Received Serial data is not identical to transmitted test data
No Received serial data during loopback test
PS apply valve failed - short
PS apply valve failed - open
PS release valve failed - short
PS release valve failed - open