Download Print this page

FLS C300.001TC Instruction Manual

Stainless steel conductivity sensor

Advertisement

Quick Links

1. General data
These special sensors are particularly suitable for conductivity measurement in
applications where stainless steel is requested as material in contact with fluid.
Sensors are completely sealed and consist on:
_ a double coaxial electrodes conductivity cell,
_ a Pt1000 temperature sensor with 3 wires connections,
_ cable connection wires are marked by different colours.
Technical features
2.
Wetted Materials
Sensor Body: Stainless Steel AISI 316L
O-ring
: EPDM
Temperature@Pressure: 100°C @ 7 bars (installed in PP fitting);
Cable
Cell constant: 0,01 ; 10,0
3. Installation
Conductivity sensor should be installed properly to gain best performances.
It should obey these rules:
_ sample in contact with sensor should be representative of fluid to measure;
_ liquid should continuously flow throw the cell;
_ flow rate should avoid cavitation phenomenas;
_ cell mounting should be so that an air bubbles presence is avoided on electrodes
surface;
_ solid sediments shouldn't accumulate into space between electrodes dedicated to
measurement;
_ it should verify that temperature and pressure conditions are obeyed.
C300.001TC, C300.10TC
Stainless Steel Conductivity Sensor
INSTRUCTION MANUAL
:
100°C @ 34 bars (installed in SS fitting);
120°C @ 13 bars (installed in SS fitting).
: 3 meters of shielded cable with 5 wires
1
EN 09-10

Advertisement

loading

Summary of Contents for FLS C300.001TC

  • Page 1 C300.001TC, C300.10TC Stainless Steel Conductivity Sensor INSTRUCTION MANUAL EN 09-10 1. General data These special sensors are particularly suitable for conductivity measurement in applications where stainless steel is requested as material in contact with fluid. Sensors are completely sealed and consist on:...
  • Page 2 It’s important to avoid to interrupt cable along its whole length. 5. Maintanance This type of sensor doesn’t require particular operations for maintenance except for the eventual cleaning of the electrodes. Dimensional Data Table Model n° “A” “B” “C” C300.001TC 2” 5,25” 3,5” C300.10TC 0,5” 9,1” 7,6”...

This manual is also suitable for:

C300.10tc