Page 3
One Instruments, Inc. be replaced or repaired and return freight prepaid. In no case shall the liability of Met One Instruments, Inc. exceed the purchase price of the product. This warranty may not apply to products that have been subject to misuse, negligence, accident, acts of nature or that have or modified other than by Met One Instruments, Inc.
GENERAL INFORMATION 1.1. Description The 085 sensor is an extremely accurate microprocessor controlled relative humidity and temperature sensor. The relative humidity sensor responds to the full range of 0 to 100% humidity. Response is linear with negligible hysteresis or temperature dependence.
1.2. 085 Options The 085 options are defined by dash numbers as follows: 085-(Shield Option) 1.2.1. Radiation Shield Compatibility Option: 6 = Use with Model 076B or 077 radiation shield. This sensor comes with 8-inch long pigtail wires for connection to the radiation shield junction box. Refer to the radiation shield manual for sensor and cable connections.
Sensors not installed in a radiation shield should be mounted in a representative location having good airflow and shaded from sunlight or other heat radiation sources that would affect measurement of relative humidity or temperature. 3.1. Mounting Instructions Figure 2 Radiation Shield Installation 085-9800 Manual Rev A.doc Page 9...
Pin D Black Temperature Common Pin E Temperature Signal Pin F No Connection No Connection No Connection White/Brown Shield Figure 3 Electrical Connector View looking at connector pins. (Pins are also identified on connector). Page 10 085-9800 Manual Rev A.doc...
The sensor’s relative humidity output is referenced to saturated water vapor pressure above liquid water. When the air temperature is below freezing, the sensor’s maximum theoretical measurement range is limited as follows: Air Temperature Maximum RH (Deg C) 085-9800 Manual Rev A.doc Page 11...
085 sensor and a standard temperature sensor (NIST thermometer or RTD). The mass can be used in air, or can be partially submerged in ice or an ice bath. The 085 sensor should not be submerged. Reference...
4.4. Temperature Table 4.4.1. Temperature to Resistance Table (Celcius) Model 085-X Temperature vs. Sensor Resistance TEMP DEG C RCAL TEMP DEG C RCAL 110236 26155 104464 25436 99187 24739 94344 24064 89882 23409 85760 22775 81939 22159 78388 21561 75079...
0.1% resistor (Met One Instruments part number 805020) is connected in series with excitation voltage. This resistor is provided as part of all Met One Instruments translators designed for use with this sensor and is included with Met One Instruments data loggers configured for this sensor. For applications with other equipment, contact Met One Instruments to obtain this special resistor.
4.8.1. Sensor Element The RH sensor element is not user-replaceable (replacement requires sensor recalibration). If the element becomes damaged, please send the 085 sensor to Met One Instruments for repair. Please obtain shipping instructions and a return authorization (RA) number before returning any unit.
-30° C to +50° C (-22° F to 122° F) Accuracy ±0.15° C (0.27° F) Output: Resistive AutoMet Auto ID: General Input Power 10 to 18 VDC @ < 5 mA Dimensions Length: 8.5 in (21.59 cm) Diameter: 0.75 in (1.91 cm) 085-9800 Manual Rev A.doc Page 19...