Boscarol OB 500 Series User Manual

Medical suction unit fa/fm/liner

Advertisement

Quick Links

Rev. 10-2018 USMAN500 - UK
MEDICAL SUCTION UNIT BOSCAROL
OB500
FA / FM / LINER
USER MANUAL
:
MANUFACTURED BY

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the OB 500 Series and is the answer not in the manual?

Questions and answers

Summary of Contents for Boscarol OB 500 Series

  • Page 1 Rev. 10-2018 USMAN500 - UK MEDICAL SUCTION UNIT BOSCAROL OB500 FA / FM / LINER USER MANUAL MANUFACTURED BY...
  • Page 2 Rev. 10-2018 USMAN500 - UK Rev. 04-2010 USMAN2012-IT 2 - 28...
  • Page 3: Table Of Contents

    Rev. 10-2018 USMAN500 - UK Rev. 04-2010 USMAN2012-IT INDEX INDEX ........................................3 SYMBOLS .........................................4 WARNINGS, PRECAUTIONS AND IMPORTANT INFORMATION ........................6 IMPORTANT INFORMATION ..................................7 OB500 SUCTION UNIT ....................................8 DESCRIPTION AND INTENDED USE STATEMENT ..........................8 CONTRA INDICATION FOR USE ................................8 CONTROLS, INDICATIONS AND CHECK PANEL ............................8 CONTROL MODULE ....................................9 SUCTION UNIT ....................................9 ELECTRICAL CONNECTIONS ................................10...
  • Page 4: Symbols

    Rev. 10-2018 USMAN500 - UK Rev. 04-2010 USMAN2012-IT SYMBOLS S.1 Symbols used on the device and recalled in the user manual Double insulation. The device is doubly electrically isolated from the user and the work environment (there is no ground conductor or connection equipotential point). Applied part type BF Only use the device in the indicated temperature range.
  • Page 5 Rev. 10-2018 USMAN500 - UK Rev. 04-2010 USMAN2012-IT Authorized Representative in the European Community, if the manufacturer is not resident therein. Expiration date Do not use if the packaging is not intact Catalogue ID code Sterile device. Sterilization method: ethylene oxide Sterile device.
  • Page 6: Warnings, Precautions And Important Information

    WARNINGS, PRECAUTIONS AND IMPORTANT INFORMATION Read carefully This User Manual has been prepared using simple, easy-to-understand language. If you have difficulty interpreting the above, contact the manufacturer for further clarification info@boscarol.it Tel.: +39 0471 932893 raq@boscarol.it • Read all instructions carefully before using the unit. Careful, proper use will ensure smooth operation and protect both patients and operators alike.
  • Page 7: Important Information

    If the substances aspirated have entered the suction unit, it must be discarded. For Oscar Boscarol srl, the safety of its employees and authorized service centre staff is important. The suction units will not be demolished according to the WEEE Directive (Waste Electrical and Electronic Equipment) if the suction unit is contaminated and there is the risk of infection (application of It.
  • Page 8: Ob500 Suction Unit

    Rev. 10-2018 USMAN500 - UK Rev. 04-2010 USMAN2012-IT OB500 SUCTION UNIT INSPECT THE SUCTION UNIT AND ALL PARTS BEFORE USE. DO NOT ATTEMPT TO USE IF ANY PARTS ARE DAMAGED OR MISSING. Model BSU402: 1. Suction unit 2. Control module 3.
  • Page 9: Control Module

    Rev. 10-2018 USMAN500 - UK Rev. 04-2010 USMAN2012-IT Control module The regulation device is supplied already assembled and it is made in plastic material (UL white ABS). The chassis comes with holes for screw fixing onto the wall. Vacuum regulation may occur through the proper knob.
  • Page 10: Electrical Connections

    Rev. 10-2018 USMAN500 - UK Rev. 04-2010 USMAN2012-IT WARNING Open the cover and check the free spaces before the fixing of the square bracket. Don not use screws more than 20mm long! Electrical connections Refer to the electric international norms (for SELV DC voltage), for the installation of the suction unit on the emergency vehicles.
  • Page 11: Test Of The Unit

    Rev. 10-2018 USMAN500 - UK Rev. 04-2010 USMAN2012-IT Test of the unit The operations described in this section allow the user to verify the correct functioning of the suction unit. The test should be performed at least once a day and before use in the field: Daily checklist: •...
  • Page 12: Reusable Jar With Disposable Liner Serres

    Rev. 10-2018 USMAN500 - UK Rev. 04-2010 USMAN2012-IT to replace the filter once a month. The material it is made should be damaged even under specific environmental conditions (humidity, heat, cold) Do not use the suction unit without the protection filter or jar! Reusable jar with disposable Liner SERRES®...
  • Page 13: Power Supply

    Rev. 10-2018 USMAN500 - UK Rev. 04-2010 USMAN2012-IT OB500 LINER includes one plastic bracket. The fixing procedure is the same as for OB500 FA. Power supply The external power source of the OB500 should be between 12÷15Vdc. Insert in series to the plus wire an appropriate protection fuse (min.
  • Page 14: Cleaning Instruction

    Rev. 10-2018 USMAN500 - UK Rev. 04-2010 USMAN2012-IT The organic secretions collected by the device’s jar might cause serious infections to the users. For this reason, suitable PPE and disinfection products must always be used, applying what has been established by the competent authorities.
  • Page 15: Replacing The Protection Filter

    Rev. 10-2018 USMAN500 - UK Rev. 04-2010 USMAN2012-IT • Never exceed the value of 60°C for washing or cleaning operations (with the exception of sterilization in a steam autoclave). • Cleaning, disinfection and sterilisation should only be carried out by trained personnel. •...
  • Page 16: Reusable Collection Jar Serres

    Rev. 10-2018 USMAN500 - UK Rev. 04-2010 USMAN2012-IT Dispose of disposable parts and disassembled the jar, soak the parts in cold water and rinse well. Then dip the same parts in hot water at a temperature not exceeding 60 ° C containing a mild, non-alcoholic detergent. After washing, rinse all parts with clean hot water (30-40°C max.) and then dry with a lint free soft cloth.
  • Page 17: Disposal Of Contaminated Parts

    To correctly disinfect and decontaminate the suction unit, we recommend using specific, approved products. These disinfectants must be free of alcoholic and/or abrasive substances. Oscar Boscarol srl can provide specific materials for disinfection of medical equipment, including our suction units. These disinfectants, available in different formats (wipes, spray, liquids), have been laboratory tested and guaranteed to deactivate viruses, bacteria and microorganisms.
  • Page 18: Accessories And Spare Parts

    SPS5092 Red angular joint for OB-j jar – 3 pcs User manual – available on the website www.boscarol.it NOTE The parts listed in the list may be changed without prior notice from the manufacturer and / or technical improvements. Contact the manufacturer for more information.
  • Page 19: Service

    Rev. 10-2018 USMAN500 - UK Rev. 04-2010 USMAN2012-IT SERVICE There are no serviceable components inside the suction unit Do not open the suction unit and do not modify any electrical or mechanical parts. Always contact your service center or the manufacturer. Any even minimal intervention on the device voids the warranty.
  • Page 20 Rev. 10-2018 USMAN500 - UK Rev. 04-2010 USMAN2012-IT Degree of protection against electric shock (IEC60601-1): TYPE BF With respect to protection from electric shock: CLASS II Degree of protection against ingress of liquids (IEC529): IP32d Accordance with general IEC60601-1: Complying with the 3° Edition Dimensions Suction unit’s max dimensions: 188 mm (l) x 188 mm (h) x 90 mm (p)
  • Page 21: The Risks Of Reciprocal Interference With Other Devices

    Rev. 10-2018 USMAN500 - UK Rev. 04-2010 USMAN2012-IT Operating in the rain OB500 Medical Suction Unit is a medical device designed to be used in closed environment. It must be firmly installed before its usage. The device is designed and made up to be safe from fluid substances penetration.
  • Page 22 Rev. 10-2018 USMAN500 - UK Rev. 04-2010 USMAN2012-IT Emissions tests Compliance Electromagnetic environment - guidance The OB500 suction unit uses RF energy only for RF emissions its internal function. Therefore, its RF Group 1 emissions are very low and are not likely cause CISPR11 any interference in nearby electronic equipment.
  • Page 23 Rev. 10-2018 USMAN500 - UK Rev. 04-2010 USMAN2012-IT GUIDELINES AND DECLARATION OF COMPLIANCE FOR MEDICAL DEVICES TEST OF CONDUCTED AND RADIATED ELECTROMAGNETIC IMMUNITY For the purposes of conducting the tests using test levels IEC60601, V =3 and E The BSU suction units are intended for use in the electromagnetic environment specified below. The user of the OB500 suction unit should assure that it’s used in such an electromagnetic environment.
  • Page 24: Guarantee

    The product that during the 24 months warranty is found defective, should be sent to Oscar Boscarol srl with a note describing the defect. The Oscar Boscarol Company will, at their discretion, arrange for repairs or complete replacement within the terms of this guarantee.
  • Page 25: Declaration Of Conformity

    Rev. 10-2018 USMAN500 - UK Rev. 04-2010 USMAN2012-IT DECLARATION OF CONFORMITY We, the manufacturer: OSCAR BOSCAROL SRL Il produttore: Via E. Ferrari, 29 – 39100 BOLZANO – ITALY Tel. +39 0471 932893 – Fax. +39 0257760140 Web: www.boscarol.it - Email: info@boscarol.it...