1 Description and Preparation
1.1 Description
The
LEVIFLOW®
LFS-SU
designed for non-invasive flow measurements in
single-use
applications
industries. The wetted materials of the sensors are
made out of biocompatible polypropylene (FDA,
USP-VI, BSE/TSE and Animal free), which can be
gamma sterilized.
Fixation Hole
Measurement Path
Label with Serie # and
Flow Direction Arrow
Inlet
Figure 1: Single-use flowsensor (example LFS-06SU)
Before using the flowsensor make yourself familiar
with the following instructions.
1.2 Inspection Prior to Use
The flowsensor should be inspected prior to use for
any damage. Do not use the flowsensor if any
damage is found. Contact Levitronix
return of any suspected flowsensor.
1.3 Traceability for Troubleshooting
For full traceability of the flowsensor the serial
number located on the sensor body (see Figure 1)
shall be used. The serial number is also printed on
the biocompatibility declaration and the calibration
sheet delivered with each sensor.
®
1.4 Levitronix
Configuration Software
For sensor parameter setting and calibration
parameter transfer to the converter (not necessary
for sensors with a parameter chip) the Levitronix
Configuration Software with version V1.30 or
newer is needed. For details regarding installation
of software and establishing RS485 connection
consult the LEVIFLOW
manual (see Doc# PL-4501-00).
PL-4509-01, Rev01, DCO# 16-191
flowsensors
are
of
the
Life
Science
Outlet
Connector
Calibration
Box
Parameter
Tag
®
regarding
®
Configuration Software
Instructions for Use of LEVIFLOW® LFS-SU Flowsensors
Phone for US:
Phone for outside US:
E-Mail:
2 General Warnings and Cautions
Gamma Sterilization
The LFS-SU flowsensors have been tested to be
robust against gamma radiation with a dose up to
40 kGy.
Handle the flowsensors with care, especially after
gamma radiation with doses above 25 kGy, since the
radiation exposure decreases the flexibility of the
flowsensor material.
3 Procedure for Installation
3.1 Mechanical/Electrical Installation
Connect the sensor to the converter together with
the relevant adaptor cables. Take specific care
about the flow direction indicated with the arrow on
the flow sensor (see Figure 1). For details about
mechanical and electrical installation and design of
an adequate hydraulic circuit, consult the user
manual (Doc# PL-4509-00).
3.2 Setting of Sensor Type Parameters
The sensor type (or size) specific parameters can
be set or updated on the converter at the first
installation or whenever the sensor type is
changed.
Note: Sensors which are labeled with "Type: PC"
on
the
calibration
parameter chip and parameters are transferred
automatically to the converter.
Start the LEVIFLOW
open the "Parameter Settings" menu (Figure 2).
Make sure that "Sensor Size", "Full Scale", "Kinem.
Viscosity" and "Liquid Temperature" are set to the
values designated for the used sensor type.
Table 1 table shows the standard values. For other
configuration
specification.
Sensor
Full Scale
Type
[L/min]
LFS-03SU
0.800
®
LFS-06SU
8.000
LFS-10SU
20.000
LFS-15SU
50.000
LFS-20SU
80.000
Table 1: Standard sensor type parameters
www.levitronix.com
888-569 07 18
+1 888-569 07 18
support@levitronix.com
CAUTION
tag
have
an
®
Configuration Software and
consult
the
specific
Kinem.
Liquid
Viscosity
Temperature
[cSt]
[°C]
1
20
1
20
1
20
1
20
1
20
integrated
sensor
1/2