MedPro SONAIR Instructions Manual page 25

Ultrasonic mesh nebulizer

Advertisement

NOTE:
- At 80 MHz and 800 MHz, the higher frequency range applies.
- These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
* Field strengths from fixed transmitters, such as base stations for radio (cellular/
cordless) telephones and land mobile radios, amateur radio, AM and FM radio
broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess
the electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the location in which this
device is used exceeds the applicable RF compliance level above, this device should
be observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as re-orienting or relocating this device. ** Over the
frequency range 150 kHz to 80 MHz, field strengths should be less than 3V/m.