Immunity Test
Power
frequency
magnetic field
GB/T 17626.8
Conducted RF
GB/T 17626.6
Radiated RF
GB/T 17626.3
Where p is the maximum output power rating of the transmitter in watts (W) d – is the
minimum separation distance in (m). The field strength of the fixed RF transmitter is
determined by an electromagnetic site survey* of the electromagnetic field. Each frequency
range b should be lower than the current frequency.** Interference may occur win the
22
vicinity of equipment marked with the following symbol:
IEC 60601
Compliance
Test Level
Level
3 A/m
3 A/m
50 Hz
50 Hz
60 Hz
60 Hz
3V rms
3 V/m
15 KHz to
80 MHz
3V/m
80 MHz
3 V/m
to 2.5 GHz
Electromagnetic
Environment Guidance
Power frequency magnetic field
should be at levels characteristic
of a typical location in a typical
commercial or hospital environment.
Portable and mobile RF
equipment should not be used
close to any part of this portable
mesh nebulizer less than
recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter.
Recommended Separation
Distance:
d = 1.2√P 150 kHz ~ 80 MHz
d = 1.2√P 80 MHz ~ 800 MHz
d = 2.3√P 800 MHz ~ 8.5 GHz