Final Inspection - Vancare VanGo 600 Manual

Mobile lift
Table of Contents

Advertisement

Final inspection

Check to ensure that no parts have been left in the packaging.
Inspect the lift for signs of wear and damage.
Check all four castor wheels and caster wheel locks.
Check all connections and fixtures including screws and bolts.
Check the emergency stop function by depressing the emergency stop, and then
pressing either the up or down button. If nothing happens when the up or down
buttons are pressed, the emergency stop is functioning properly.
Grasp the handset, press the up button and run the lift arm all the way up. Then, press
the down button and run the lift arm all the way down.
Test base-width adjustment function.
Test base-width adjustment function on the VanGo 400 by pushing down on the
respective pedals for widening and narrowing the base.
Test lift function by lifting a person (not a patient) using an approved sling. At the same
time, check the emergency lowering function with someone on the lift. See section on
Emergency lowering.
If the lift is functioning correctly, connect the charger and check to ensure that the
charging lamp on the control box lights up.
NOTE!
Before the lift is used for the first time, it must be charged for at least 4 hours. See
section on charging batteries.
Keep the manual where it is accessible to users of the product.