® CAPINTEC, INC. CAPTUS TABLE OF CONTENTS PREFACE SAFETY ..........................1-1 SYMBOL DEFINITIONS ....................1-1 WARNING AND CAUTION LABELS .................1-2 CAUTIONS AND NOTES ....................1-5 GENERAL SAFETY TIPS....................1-6 FUNCTIONAL & TECHNICAL DESCRIPTION ..............2-1 GENERAL..........................2-1 TECHNICAL DESCRIPTION.....................2-1 Detector .........................2-2 Multichannel Analyzer....................2-2 TECHNICAL SPECIFICATIONS ..................2-2 Thyroid Uptake Stand ....................2-2...
Page 4
® CAPINTEC, INC. CAPTUS Lab Tests – Urine......................2-9 Lab Tests – Blood ....................2-10 Staff Bioassay......................2-10 GENERAL OPERATING OVERVIEW..................3-1 TURN ON SYSTEM ......................3-1 LOG-ON PASSWORD ......................3-1 SYSTEM FUNCTIONS......................3-2 PROBE SCREEN ......................3-3 WELL SCREEN.........................3-3 HOME Key ........................3-4 Changing from PROBE to WELL Functions ..............3-4 Changing from WELL to PROBE Functions ..............3-4...
Page 5
® CAPINTEC, INC. CAPTUS Staff Data ........................3-13 SYSTEM SETUP........................4-1 INSTALLATION .........................4-1 ENVIRONMENT REQUIREMENTS ..................4-1 Operational ........................4-1 Storage ..........................4-1 POWER REQUIREMENTS ....................4-2 TURN ON PROCEDURE ....................4-2 GENERAL OPERATIONAL SETUP ..................4-3 ACCEPTANCE TESTING....................4-3 SETTING UP FOR THYROID UPTAKES OR WIPE TESTS ..........4-4 SETTING UP FOR LAB TESTS (URINE &...
Page 6
® CAPINTEC, INC. CAPTUS THYROID UPTAKE SETUP ....................7-1 COUNTING METHODS ....................7-1 PROTOCOLS ........................7-4 Adding a Protocol ......................7-6 Editing a Protocol......................7-11 Deleting a Protocol ......................7-11 Printing Protocol Information ..................7-11 THYROID UPTAKE GENERAL INSTRUCTIONS..............8-1 ACCESSING THYROID UPTAKE STUDIES ..............8-1 ENTERING PATIENT DEMOGRAPHICS .................8-2 SELECTING PROTOCOL ....................8-3...
Page 7
® CAPINTEC, INC. CAPTUS UPTAKE MEASUREMENT .....................11-9 VIEWING RESULTS......................11-9 WIPE TEST SETUP ......................12-1 INTRODUCTION ......................12-1 WIPE DESCRIPTIONS....................12-2 Entering Data for a Wipe Location ................12-2 Adding a Location ......................12-3 Editing or Deleting a Location ..................12-4 TRIGGER LEVEL/COUNT TIME..................12-4 GENERIC CONVERSION FACTORS ................12-6 ADDITIONAL NUCLIDES ....................12-7...
Page 8
® CAPINTEC, INC. CAPTUS MAKING BIOASSAY MEASUREMENTS................17-5 PRINTING STAFF BIOASSAY RESULTS ..............17-7 MULTICHANNEL ANALYZER SETUP ................18-1 INTRODUCTION ......................18-1 MCA SETUPS .........................18-1 Adding a Setup ......................18-3 Editing a Setup ......................18-6 Deleting a Setup ......................18-6 Printing All the Setups ....................18-6 MULTICHANNEL ANALYZER OPERATION ..............19-1 MCA MEASUREMENTS ....................19-1...
CAPTUS PREFACE Thank you for purchasing the CAPINTEC line of Thyroid Uptake Measurement Systems. Every effort has been made to insure that the information in this document is complete, accurate, and up-to-date. CAPINTEC assumes no responsibility for the results of errors beyond its control.
Page 10
Plug the unit into an outlet on a circuit different from that which the affected device is connected. If this fails to correct the problem, please contact Capintec’s Authorized Service Center. INSTALLATION Your system has been shipped from the factory pre-configured and fully tested. Installation consists of placing the components on the stand and securing them in place.
® CAPINTEC, INC. CAPTUS CHAPTER 1 SAFETY ® These warnings and instructions for use form an integral part of the CAPTUS 600 and must therefore be kept available for consultation at all times. Precise compliance with the warnings and instructions is an essential condition for normal use, correct application and thus safety of patients and user.
® CAPINTEC, INC. CAPTUS WARNING AND CAUTION LABELS Located in the center section on both sides of the spring-arm of the floor stand is a warning label, (Figure 1-1), concerning the hazard of removing the collimator when the spring-arm is in the wrong position.
Page 13
® CAPINTEC, INC. CAPTUS Figure 1-3 Located on the left side of the Isolation Transformer is a label (Figure 1-4), with two caution statements pertaining to the main power cord of the Isolation Transformer. The statement, READ INSTRUCTIONS CAREFULLY BEFORE CONNECTING TO OUTLET, directs the user to the owner’s manual before connecting the power.
Page 14
® CAPINTEC, INC. CAPTUS Figure 1-5 ® Located on the connector guard on the back panel of the CAPTUS 600 Readout Unit is a label (Figure 1-6), with the caution statement pertaining to safety of the operator. This statement, DO NOT REMOVE CONNECTOR GUARD, is necessary because of the presence of high voltage on the connectors and on the inside of the Instrument.
However, there is no guarantee that interference will not occur in a particular installation. CAUTION If any printer other than the model supplied by Capintec is used, the safety of the unit may be compromised or Electromagnetic Interference (EMI) may be introduced into other devices ®...
NOTE It is recommended that periodic (every five years) re-calibration of the unit be performed by Capintec’s Authorized Service Center (reference CHAPTER 20: CLEANING AND MAINTENANCE) to guarantee that the instrument's high reliability is maintained. GENERAL SAFETY TIPS •...
® CAPINTEC, INC. CAPTUS CHAPTER 2 FUNCTIONAL & TECHNICAL DESCRIPTION GENERAL ® The CAPTUS 600 Thyroid Uptake / Well System is intended to be used by trained Nuclear Medicine Technologists for performing Thyroid Uptake, Thyroid Bioassay, Wipe Test, Blood Volume Test, Shillings Test and RBC Survival Test.
® CAPINTEC, INC. CAPTUS Detector ® Your CAPTUS 600 system is equipped with a Probe (and possibly a Well), which consists of a NaI (Tl) (thallium-activated sodium iodide) crystal and photomultiplier tube detector assemblies. When a gamma or x-ray from a radioactive source reaches the NaI crystal, it collides with the electrons in the crystal matrix and transfers energy.
® CAPINTEC, INC. CAPTUS System Power Requirements CAUTION If the input voltage to the following items is not within the stated limits, the unit may not function correctly or may be damaged. Line Voltage: 115 or 230 VAC depending on model ordered...
® CAPINTEC, INC. CAPTUS Isolation Transformer ® An isolation transformer is provided with the CAPTUS 600 system to ensure that the leakage current meets EN60601-1-1 Medical Equipment Safety Standard. A detachable 4.6m (15 ft) power cord, with a hospital grade plug, is provided to connect the isolation transformer to the power source.
® CAPINTEC, INC. CAPTUS Calibration Source Data Memory for Cs137 (1 source for Probe, 1 source for Well) Multichannel Pulse Height Analyzer 256 Channels 4 Gain Ranges (1, 2, 4, 8 keV per channel) Automatic self-calibration Linearity - within 2% of full scale Differential Linearity - within 2% of full scale A/D Conversion time - 35 µsec - Maximum count rate 25,000 cps (real time)
® CAPINTEC, INC. CAPTUS Probe Detector NaI(TI) flat face crystal / PM Tube: 5.1 x 5.1 cm (2" x 2") Features Monoline construction Pulse Height Resolution - 8% or better for Cs137 Lead Shielded Collimator Complies with ANSI Standard N44.3-1973 Lined with brass to minimize back scatter Well Detector –...
® CAPINTEC, INC. CAPTUS FUNCTIONAL DESCRIPTION Thyroid Uptake ® The CAPTUS 600 Thyroid Uptake / Well System is intended to be used by trained Nuclear Medicine Technologists for performing Thyroid Uptake, Thyroid Bioassay, Wipe Test, Blood Volume (with Cr-51 or I-125) Test, Shillings Test and RBC Survival Test.
® CAPINTEC, INC. CAPTUS One Capsule Method A capsule is measured in a neck phantom. The same capsule is then administered to the patient. At a later time, the patient returns and the Thyroid (neck) counts and thigh counts are measured. The uptake is then calculated as:...
® CAPINTEC, INC. CAPTUS Multi-Capsules In the One Capsule method, the patient’s dose can be more than one capsule. In the Two Capsule method, the patient’s dose and the reference can be more than one capsule. There are four possible choices for Multi-Capsules option: Only one capsule may be used.
® CAPINTEC, INC. CAPTUS Lab Tests – Blood Blood Volume Test Blood Volume determinations involving radioactive tagging are most frequently used in specific disease conditions when the hematocrit may not accurately estimate true blood volume. Such conditions include extensive trauma or burn patients, certain types of anemia, and polycythemia.
14:32 THYROID UPTAKE SYSTEM CAPTUS™-600 REV X.XX COPYRIGHT 1994 ALL RIGHTS RESERVED CAPINTEC, INC. NJ USA Press Any Key to Continue Figure 3-1 Press any key to continue. LOG-ON PASSWORD A password is required to gain access to the program.
® CAPINTEC, INC. CAPTUS ENTER PASSWORD Figure 3-2 ® The password (which CANNOT be changed) for all CAPTUS 600 systems consists of 6 characters. The password is 897NNN where NNN is the last three digits of the serial number. ®...
® CAPINTEC, INC. CAPTUS PROBE SCREEN The Probe Screen is shown in Figure 3-3. PROBE 14:32 STAND Figure 3-3 WELL SCREEN The Well Screen is shown is Figure 3-4. 14:32 W E LL W IP E LA B Figure 3-4...
® CAPINTEC, INC. CAPTUS HOME Key ® Pressing the HOME key when performing a Probe function returns the CAPTUS 600 to the Probe Screen (Figure 3-3). ® Pressing the HOME key when performing a Well function returns the CAPTUS 600 to the Well Screen (Figure 3-4).
® CAPINTEC, INC. CAPTUS FUNCTION Keys The Keys labeled F1, F2, F3, F4 – across the top row at the left side of the keyboard – are function keys. They correspond to the rectangles at the bottom of certain screens (as in Figure 3-3 and Figure 3-4).
® CAPINTEC, INC. CAPTUS Note: The “0” on this screen will say “Return to Probe Screen” or “Return to Well Screen” depending upon Main Screen (Figure 3-3 or Figure 3-4) from which you entered Utilities. KEYBOARD ® The CAPTUS 600 keyboard is used to select functions and to input data (letters, numbers and radionuclide names).
® CAPINTEC, INC. CAPTUS ALPHANUMERIC ENTRIES When a number or alphanumeric value is to be entered, the prompt will appear. After the first character is entered, the cursor (a filled in rectangle) will appear at the next character to be entered.
® CAPINTEC, INC. CAPTUS BLANK ENTRIES While some entries are optional and may be left blank (for example, patient’s date of birth) most are required and must be filled in. If the [OK] key is pressed while a required entry is...
® CAPINTEC, INC. CAPTUS YES or NO Answers Whenever a question is asked, it must be answered Yes or No. Answer the question by pressing Y and ENTER for YES or N and ENTER for NO. You can change a YES to a NO answer, and vice versa, if you have not yet pressed the ENTER key.
Adjustment clockwise of these two controls increases the voltage and counter clockwise decreases the voltage. Note: The setting of these controls should not be changed unless directed by a Capintec Service Representative. Unauthorized adjustment of these two controls will invalidate the calibration of the unit.
® CAPINTEC, INC. CAPTUS DISPLAY CONTRAST ADJUSTMENT This adjustment controls the contrast of the Readout LCD. Perform adjustment until LCD screen displays the best contrast. DATA ENTRY INPUT LIMITS Thyroid Uptake Protocols REGIONS OF INTEREST I123 I131 Tc99 Left of peak 10 –...
® CAPINTEC, INC. CAPTUS Bioassay Protocol Range Default X-RAY / GAMMA RAY (I 131) 2 – 20 % 8.5% X-RAY / GAMMA RAY (I 123) 2 – 100% DISTANCE 15 – 40cm 25 cm Additional Nuclide HALF-LIFE .1 – 9999.9 (for M, H, D, Y) ENERGY 20 –...
® CAPINTEC, INC. CAPTUS CHAPTER 4 SYSTEM SETUP INSTALLATION Your system has been shipped from the factory pre-configured and fully tested. Installation consists of placing the components on the stand and securing them in place as shown in the ®...
Your system will power up and display the CAPTUS 600 Start Up Screen (Figure 4-1) 14:32 THYROID UPTAKE SYSTEM CAPTUS™-600 REV X.XX COPYRIGHT 1994 ALL RIGHTS RESERVED CAPINTEC, INC. NJ USA Press Any Key to Continue Figure 4-1 SYSTEM SETUP November 07...
® CAPINTEC, INC. CAPTUS GENERAL OPERATIONAL SETUP ® There are several things that must be done before using the CAPTUS 600 Thyroid Uptake System for the first time. The following briefly describes these steps and refers the reader to the appropriate section in the owner's manual in order to perform them.
® CAPINTEC, INC. CAPTUS • Chi-Square Test: Reference CHAPTER 6: ACCEPTANCE & QUALITY ASSURANCE TESTS, SECTION: CHI-SQUARE TEST for instructions on how to perform this test. • Minimum Detectable Activity: Reference CHAPTER 6: ACCEPTANCE & QUALITY ASSURANCE TESTS, SECTION: MDA TEST for instructions on how to perform this test.
® CAPINTEC, INC. CAPTUS CHAPTER 5 SYSTEM INITIALIZATION DATE & TIME SETUP Date Entry Format Although dates are displayed as month-day-year, the date is entered in numeric format (2 digits for month, 2 digits for day, 2 digits for year).
Changing the Time and Date Setting ® The Capintec CAPTUS 600 has an internal clock so that the date and time of all procedures can be documented. It is important that Date and Time be set properly before using the ®...
The new date and time will then be displayed for verification. SETTING HOSPITAL OR FACILITY NAME ® All printouts from the Capintec CAPTUS 600 include your hospital or facility name and address. The Hospital Name consists of up to 4 lines of 30 characters each.
® CAPINTEC, INC. CAPTUS 13:06 H O S P IT A L N A M E > N U M Figure 5-5 Input up to 4 lines for the hospital name. Press F1 [NUM] to switch to numeric input and display “LET” for F1.
Page 49
® CAPINTEC, INC. CAPTUS U TILITIES 14:32 1. M ULTI CH AN N EL AN ALY ZE R 2. S TAFF BIO AS S AY RE SULTS 3. S ETUP 4. DIAG N O STICS 0. RETURN TO PRO BE SCREEN Figure 5-6 Press 3 to select SETUP.
Page 50
® CAPINTEC, INC. CAPTUS A C T IV IT Y W ILL B E IN C i C i/B q Figure 5-8 The F1 [Ci/Bq] key is a toggle switch. When this key is pressed, the screen changes to Figure 5-9.
CAPINTEC, INC. CAPTUS Cs137 STANDARD SETUP ® All calibration of the Capintec CAPTUS 600 is done with a Cesium-137 standard reference source. Before using this source to calibrate the system, the source calibration data must be ® input in the CAPTUS 600 memory.
Page 52
® CAPINTEC, INC. CAPTUS 14:32 1. C S 137 S TA N D A R D 2. C H I-S Q U A R E P R O T O C O L 0. R eturn to S etup M en u Figure 5-12 Press 1 to select CS137 STANDARD.
® CAPINTEC, INC. CAPTUS CS137 Standard 14:32 PRO BE Source ID :> μ Ci 0.11 Calib Activity : Calib Date Jan 06, 1992 Figure 5-14 For the Source ID, input an identification name for the source. This name can include letters and/or numbers.
Page 54
® CAPINTEC, INC. CAPTUS Press 3 to select SETUP. The message “PASSWORD:” will appear. Type the last three digits ® of your CAPTUS 600 serial number, and press the ENTER key. The Setup menu (Figure 5- 16) will appear. SETUP 14:32 1.
® CAPINTEC, INC. CAPTUS 14:32 CHI-SQUARE PROTOCOL REPETITIONS: >10 30 SEC COUNTING TIME: Figure 5-18 Input the number of repetitions desired (the minimum is 5 and the maximum is 30; at least 10 repetitions is recommended). Input the counter time for each measurement (the minimum is 10 seconds and the maximum is 9999 seconds).
Page 56
® CAPINTEC, INC. CAPTUS To access the Conversion Factors section, press the HOME key, then the UTIL key to display the main Utilities menu (Figure 5-19). UTILITIES 14:32 1. M U LTICHANNEL ANALYZER 2. STAFF BIO ASSAY RESULTS 3. SE TUP 4.
® CAPINTEC, INC. CAPTUS CONVERSION FACTORS 14:32 1. WELL 2. PROBE 3. VIEW/PRINT CONV FACTORS 0. RETURN TO PREVIOUS SCREEN Figure 5-21 Press 3 to view and/or print the current Conversion Factors. Select the appropriate detector by pressing 1 or 2.
® CAPINTEC, INC. CAPTUS For example, to enter Cobalt-57, input C, O, 5, 7 and ENTER; to enter Iodine-131, input I, SP, 1, 3, 1 and ENTER. Press F4 [OK] when the Nuclide Name is correct. Press F1 [DONE] if all CFs have been input or measured.
® CAPINTEC, INC. CAPTUS C F FO R W E LL 13:06 N U C LID E N A M E : T C 99m IN P U T C F: > Figure 5-24 Input the CF in units of dpm/cpm.
Page 60
® CAPINTEC, INC. CAPTUS M EASURE TC99m SO UR CE IN W ELL CO U N TING PE RIO D: >60 SEC STA RT E XIT Figure 5-26 Input the counting time. The correct counting time depends upon the strength of the source.
Page 61
® CAPINTEC, INC. CAPTUS 200 CTS FS 4335. cpm O K ? (Y O R N) 200 keV Figure 5-27 If the data is OK, press Y. You will then be asked to measure background. After you accept the background, a summary screen (Figure 5-28) will appear. The example is for a measurement of Co57.
® CAPINTEC, INC. CAPTUS CHAPTER 6 ACCEPTANCE & QUALITY ASSURANCE TESTS ® To insure proper operation of the CAPTUS 600, the following tests should be performed at the indicated intervals. ACCEPTANCE TESTS The following tests must be performed in the following order before the initial use of the ®...
® CAPINTEC, INC. CAPTUS Quarterly Tests The Chi-Square Test, as described in the CHI-SQUARE TEST section on page 6-5, may be performed every time Calibration Test is performed but should be performed at least quarterly. The Chi-Square Test is performed by repeatedly measuring the Cs137 standard using the region of interest determined in the test just performed.
Page 65
® CAPINTEC, INC. CAPTUS TEST 14:32 1. WELL 2. PROBE 0. RETURN TO PREVIOUS SCREEN Figure 6-2 To test the Well, press 1. To test the Probe, press 2. Place the source in the correct geometry for the detector. (i.e., in the center of the Flat Probe;...
Page 66
If the Calibration Test fails, perform an Autocalibration function as described in the AUTO CALIBRATION TEST section on page 6-5. Repeat the Calibration Test. If it fails another time, contact Capintec’s Authorized Service Center for service. ACCEPTANCE & QUALITY ASSURANCE TESTS...
If the Chi-Square falls outside of this range, first repeat the test making sure that the background and positioning are kept constant. If the values are still out of range, contact Capintec’s Authorized Service Center for assistance. You always have the option of performing the Autocalibration Test, even when the Test shows that calibration is within established limits.
Page 68
® CAPINTEC, INC. CAPTUS The system also allows you to request the Auto Cal procedure, even when it is not indicated by the Test function. Press the HOME key, then the UTIL key. The main Utilities menu (Figure 6-7) will appear.
Page 69
® CAPINTEC, INC. CAPTUS CALIBRATION 14:32 1. WELL 2. PROBE 0. RETURN TO PREVIOUS SCREEN Figure 6-9 Press 1 to select WELL or 2 to select PROBE. After making this selection, the appropriate Calibration Screen will appear. Figure 6-10 is an example of the Probe Calibration Screen. If data for a cesium standard was entered in the Cs137 Standard Setup, the source name will appear on the display.
Page 70
® CAPINTEC, INC. CAPTUS ® The CAPTUS 600 will repeatedly adjust the High Voltage to the detector until the Cs137 peak is adjusted to 662 ±8 keV. The counting period may repeat several times until the system is calibrated. 2000 Cts FS...
® CAPINTEC, INC. CAPTUS MDA TEST This is a test of Minimum Detectable Activity for the specific nuclide selected by the user. The test can be performed for the Probe or Well. A nuclide is chosen for the test (the default is Tc99m) and the appropriate energy range (400 keV or 1600 keV) is automatically selected.
Page 72
® CAPINTEC, INC. CAPTUS 14:32 MDA TEST 1. WELL 2. PROBE 0. RETURN TO PREVIOUS SCREEN Figure 6-14 To test the Well, press 1 . To test the Probe, press 2 . After Well or Probe is chosen, the Nuclide Entry Screen (Figure 6-15) will appear.
Page 73
® CAPINTEC, INC. CAPTUS MDA TEST 13:06 WELL COUNTING PERIOD: >6O SEC PRECISION : 3 ST. DEV. START PREC Figure 6-16 Input the desired counting period (60 sec is the minimum allowed). Press F1 [PREC] to change the precision from 1 to 9 standard deviations (the default value is 3).
® CAPINTEC, INC. CAPTUS DIAGNOSTICS ® The CAPTUS 600 has the ability to check system functions using a Diagnostics Utility. To access the Diagnostics Utility, press the HOME key and then the UTIL key. The main Utilities Menu (Figure 6-17) will appear.
Figure 6-19 If any of the test results indicate “FAIL” rather than “PASS”, it indicates a system malfunction. If this happens, perform the Diagnostics Utility again. If the test(s) fail again, contact Capintec’s Authorized Serve Center for assistance. CAUTION! ®...
The results of the RAM, PROM, and EEPROM tests. If the PROM test fails, the PROM values will be printed. Any test that fails indicates a system malfunction and warrants a call to Capintec’s Authorized Service Center. 6-14 ACCEPTANCE & QUALITY ASSURANCE TESTS...
Page 77
® CAPINTEC, INC. CAPTUS Page 1 November 07 ACCEPTANCE & QUALITY ASSURANCE TESTS 6-15...
Page 78
® CAPINTEC, INC. CAPTUS Page 2 Page 3 6-16 ACCEPTANCE & QUALITY ASSURANCE TESTS November 07...
® CAPINTEC, INC. CAPTUS CHAPTER 7 THYROID UPTAKE SETUP COUNTING METHODS Counting Methods refer to the Counting Time Base and Regions of Interest for each nuclide. The values are the same for all entered protocols. ® The CAPTUS 600 offers two counting time bases: REAL time and LIVE time. The ®...
Page 80
® CAPINTEC, INC. CAPTUS Note: Selection 0 will display “RETURN TO WELL SCREEN” if the Utilities menu is accessed from the Well main screen. Press 3 to select SETUP. The message “PASSWORD:” will appear. Type the last three digits ®...
Page 81
® CAPINTEC, INC. CAPTUS THYROID UPTAKE SETUP 14:32 COUNTING TIME BASE: LIVE Counting stops when the measured Live Time reaches the specified counting period BASE Figure 7-4 The F1 [BASE] key is a toggle switch that changes the Counting Time Base between two choices: LIVE and REAL.
® CAPINTEC, INC. CAPTUS The Region Of Interest Screen will be displayed for each nuclide. The default values are 20% to the left and 20% to the right for all the nuclides. The range limits are: I123: TO LEFT: from 10% to 40%...
Page 83
® CAPINTEC, INC. CAPTUS Press 3 to select SETUP. The message “PASSWORD:” will appear. Type the last three digits ® of your CAPTUS 600 serial number, and press the ENTER key. The Setup menu (Figure 7- 8) will appear. SETUP 14:32 1.
® CAPINTEC, INC. CAPTUS 14:32 14:32 TU P R O TO C O LS TU P R O TO C O LS > M A IN P R O TO C O L 2N D P R O TO C O L...
Page 85
® CAPINTEC, INC. CAPTUS Press F1 [NUM] to switch to numeric input and display “LET”. Press F1 [LET] to switch back to letter input and display “NUM”. When the protocol name is correct, press F4 [OK]. The Calculation Method Screen (Figure 7-13) will appear.
Page 86
® CAPINTEC, INC. CAPTUS Press F1 [NUCLD] to select the nuclide. The nuclide will cycle from “I131” to “Tc99m” to “I123” and back to “I131”. Press F2 [NUMBR] to select the number of uptakes in the study. The choices are “1”, “2”, “3”...
Page 87
® CAPINTEC, INC. CAPTUS If the Calculation Method is One Capsule or Two Capsules, the Multi-Capsules option is now chosen as shown in Figure 7-16. 14:32 TU P R O TO C O LS N A M E : P R O T 1...
Page 88
® CAPINTEC, INC. CAPTUS Note: This is only the expected time of the uptake measurement for this protocol. The actual time the study is performed will determine how results are calculated. The actual time will appear on all reports. ®...
® CAPINTEC, INC. CAPTUS Editing a Protocol A protocol may be changed (edited) when it is not in use. If there is an active patient with the selected protocol, a message will appear and you will only be allowed to view the protocol, not edit it.
® CAPINTEC, INC. CAPTUS CHAPTER 8 THYROID UPTAKE GENERAL INSTRUCTIONS The instructions in this section pertain to all of the Calculation Methods. Further instructions for each method are in separate sections. The section describes: • Accessing Thyroid Uptake Studies, •...
® CAPINTEC, INC. CAPTUS Press F1 [T/U]. The Thyroid Uptake Patient Selection Screen (Figure 8-2 or Figure 8-3) will appear. If no patients have been entered, the list is blank and only the F1 [NEW] key is displayed. TH Y R O ID U P TA K E...
® CAPINTEC, INC. CAPTUS To exit from the Patient Selection Screen, press BSP or HOME to return to the main Probe Screen (Figure 8-1). SELECTING PROTOCOL If more than one protocol has been entered, the Select Protocol Screen (Figure 8-5) will appear.
® CAPINTEC, INC. CAPTUS If the Thyroid Uptake Protocol Information, displayed in Figure 8-6, is not correct, press the N key, to answer “NO”, then press the ENTER key. The Select Protocol Screen (Figure 8-5) will re-appear. Once the patient information is entered it can still be changed. Refer to the VIEWING PATIENT INFORMATION section on page 8-5 for instructions.
® CAPINTEC, INC. CAPTUS After the second measurement has been made, the result of the second uptake and the time for the next uptake will be displayed as shown in Figure 8-9. 13:11 THYROID UPTAKE >SMITH, JOH 24 WE 11:49...
® CAPINTEC, INC. CAPTUS THYROID UPTAKE 14:32 DASILVA, AND PROTOCOL: 1 CAPSULE μ CAPSULE : TU 11:09 250.0 UPTAKE 1: 24% UPTAKE 2: TU 15:09 EDIT GRAPH MEAS Figure 8-10 Press F3 [EDIT] to edit patient data. Press F2 [DEL] to delete patient data Press F1 [GRAPH] to view the graphic display as shown in Figure 8-20.
® CAPINTEC, INC. CAPTUS If the prompt is not at the first uptake, the previous uptake can be re-measured. If the “ONE CAPSULE” method has been chosen and the capsule has been measured but not administered, it can be re-measured.
Page 98
® CAPINTEC, INC. CAPTUS 14:32 THYROID UPTAKE DASILVA, AND PROTOCOL: 1 CAPSULE MEASURE NECK AT: 25 CM COUNTING REPETITION # 1 / 2 EXIT START Figure 8-13 Note: Figure 8-13 calls for “COUNTING REPETITION #1”. Some protocols call for repetitions of each measurement so that results can be averaged. In such protocols, all measurements except room background counts are repeated.
® CAPINTEC, INC. CAPTUS If the final count value does appear to be correct, press Y and ENTER. The Measure Thigh Screen (Figure 8-16) will appear. 14:32 THYROID UPTAKE DASILVA, AND PROTOCOL: 1 CAPSULE THIGH MEASURE AT: 25 CM COUNTING REPETITION # 1 / 2...
Page 100
® CAPINTEC, INC. CAPTUS Press F1 [SKIP] to skip the measurement (i.e. perform it later if possible or choose not to perform it). Figure 8-18 will appear. 14:32 TH Y R O ID U P TA K E P R E -D O S E M E A S U R E M E N T: S K IP...
® CAPINTEC, INC. CAPTUS PRINTING RESULTS After the uptake measurements have been completed, a results summary screen for the chosen method will be displayed as shown in Figure 8-20. The results of all the uptakes thus far performed for this patient will be displayed in graphic form: DASILVA, AND ..........
® CAPINTEC, INC. CAPTUS CHAPTER 9 THYROID UPTAKE: ONE CAPSULE METHOD GENERAL PROCEDURE The general procedure for performing a TU study via the One Capsule Method is: • Enter Patient Demographics • Choose One Capsule Protocol • Perform Pre-Dose Measurement (if it is in protocol) •...
® CAPINTEC, INC. CAPTUS ENTERING CAPSULE DATA The One Capsule Data Entry Screen (Figure 9-1) will appear. K E TC H U M , LIN P R O TO C O L: P R O T 1 N U C LID E : I 131 R X ID :>...
® CAPINTEC, INC. CAPTUS Press the F3 [DONE] key if you are not ready to measure the capsule(s). The screen returns to the TU Patient Selection Screen, with the new patient added to the list. Press the F4 [MEAS] key to measure the capsule(s).
Page 106
® CAPINTEC, INC. CAPTUS dt: 4% 5683. cpm 400 keV CAPSULE ABORT Figure 9-4 If the system does not find a peak at the expected energy, the following message will appear: MEASUREMENT ERROR Press Any Key to Continue Check to make sure that the correct capsules are in the Probe, or if any background activity could be causing interference, and press any key to perform the measurement again.
Page 107
® CAPINTEC, INC. CAPTUS Screen (Figure 9-3) will re-appear with the last line indicating the next repetition (ex. Counting Repetition # 2 / 2). The prompt will not appear. After the last repetition, the screen changes to the Measure Background Screen (Figure 9-6).
® CAPINTEC, INC. CAPTUS 5683. cpm (Y OR N) 400 keV BACKGROUND Figure 9-8 If the total count, displayed in the upper left corner, is not OK, press the N key to answer “NO” and press ENTER. The Measure Background Screen (Figure 9-6) will re-appear allowing the background measurement to be repeated.
® CAPINTEC, INC. CAPTUS Input the time and date of administration if it is different from the default of “now”. The number of capsules in the dose depends on the Multi-Capsule option: • “NO”: The number of capsules in the dose is 1 and cannot be changed.
® CAPINTEC, INC. CAPTUS CHAPTER 10 THYROID UPTAKE: TWO CAPSULE METHOD GENERAL PROCEDURE The general procedure for performing a Thyroid Uptake study via the Two Capsule Method • Enter Patient Demographics • Choose Two Capsule Protocol • Perform Pre-Dose Measurement (if it is in protocol) •...
Page 114
® CAPINTEC, INC. CAPTUS N A M E , P A T P R O T O C O L: 2 C A P A D M IN IS T E R E D R X ID :> μ C i...
® CAPINTEC, INC. CAPTUS Press F3 [DONE] if it is not time to make a Thyroid Uptake measurement. Press F4 [MEAS] to make a Thyroid Uptake measurement. MEASURING THE REFERENCE CAPSULE(S) In the Two Capsule Method, each uptake begins by measuring the reference capsule(s) in a phantom.
Page 116
® CAPINTEC, INC. CAPTUS dt: 4% 5683. cpm 400 keV REFERENCE ABORT Figure 10-4 If the system does not find a peak at the expected energy, the following message will appear: MEASUREMENT ERROR Press Any Key to Continue Check to make sure that the correct capsule(s) are in the Probe, or if any background activity could be causing interference, and press any key to perform the measurement again.
Page 117
® CAPINTEC, INC. CAPTUS Screen (Figure 10-3) will re-appear, with the last line indicating the next repetition (ex. Counting Repetition # 2 / 2). The prompt will not appear. After the last repetition, the screen changes to the Measure Background Screen (Figure 10-...
® CAPINTEC, INC. CAPTUS 5683. cpm (Y OR N) 400 keV BACKGROUND Figure 10-8 If the total count, displayed in the upper left corner is not OK, press the N key to answer “NO” and press ENTER. The Measure Background Screen (Figure 10-6) will re-appear allowing the background measurement to be repeated.
® CAPINTEC, INC. CAPTUS VIEWING RESULTS After the neck and thigh have been measured, the Results Screen (Figure 10-9) will appear. μ Ci DO SE (D) : 200.0 NECK (N) : 3647 cpm THIG H (B) : 2107 cpm THYRO ID (T: N-B) : 1540 cpm...
Page 120
® CAPINTEC, INC. CAPTUS Figure 10-11 10-8 THYROID UPTAKE: TWO CAPSULES METHOD November 07...
® CAPINTEC, INC. CAPTUS CHAPTER 11 THYROID UPTAKE: STANDARD CAPSULE METHOD GENERAL PROCEDURE The general procedure for performing a TU study via the Standard Capsule Method is: • Enter TU Standard Data • Measure TU Standard once a day •...
® CAPINTEC, INC. CAPTUS 14:32 TU STANDARD COUNTING PERIOD: >120 SEC STANDARD ID μ Ci CALIB ACTIVITY CALIB DATE :JAN 30, 1994 CALIB TIME :12:34 Figure 11-5 Input the desired Counting Period. The default of 120 seconds is displayed. The minimum value is 10 seconds and the maximum value is 999 seconds.
® CAPINTEC, INC. CAPTUS This question, with the appropriate Standard Name in the second line, will appear on the screen: DO YOU WANT TO DELETE STANDARD NAME Y OR N If a TU Standard is being used in a TU study, then it cannot be deleted. If an attempt is made...
Page 125
® CAPINTEC, INC. CAPTUS Input the distance that will be used for the TU studies. The distance entered here cannot be changed when a patient is being measured. When the distance is correct, place the capsule in the phantom and press F4 [START]. The Standard Count Screen (Figure 11-7) will appear.
® CAPINTEC, INC. CAPTUS If the measurement is accepted, the TU Standard Done Screen (Figure 11-9) will appear. 14:32 TU S TA N D A R D I 131 S T A N D A R D ID : A T: 25.0 C M...
® CAPINTEC, INC. CAPTUS Select the desired Standard by moving the cursor next to the name of the Standard and press the ENTER key. If the Standard nuclide is not the same as the protocol nuclide, an error message will appear and you will be able to select again.
Page 128
® CAPINTEC, INC. CAPTUS N A M E , P A T P R O TO C O L: S T A N D A D M IN IS TE R E D R X ID :> μ C i...
® CAPINTEC, INC. CAPTUS If the Standard has not been measured “today”, the Standard Measurement procedure will be performed and then the uptake procedure will be entered. Press HOME when the Measure Neck Screen appears if it is not time for an uptake.
Page 130
® CAPINTEC, INC. CAPTUS DASILVA, AND ............................................hours AGAIN DONE Figure 11-15 Press PRINT [P] to print a report of this TU study. Refer to Figure 11-16 for an example of the printout. Press F1 [AGAIN] if an error was made in the measurement and the measurement needs to be repeated.
Page 131
® CAPINTEC, INC. CAPTUS Figure 11-16 November 07 THYROID UPTAKE: STANDARD CAPSULE METHOD 11-11...
® CAPINTEC, INC. CAPTUS CHAPTER 12 WIPE TEST SETUP INTRODUCTION Several items must be set up in order to do Wipe Testing. To access the Wipe Setup, press the HOME key and then the UTIL key. The main Utilities menu (Figure 12-1) will appear.
® CAPINTEC, INC. CAPTUS Press 7 to select WIPE AND MCA SETUP. The Wipe and MCA Setup menu (Figure 12-3) will appear. WIPE AND MCA SETUP 14:32 1. WIPE DESCRIPTIONS 2. USER KEY ASSIGNMENTS 3. ADDITIONAL NUCLIDES 4. GENERIC CONVERSION FACTORS 5.
® CAPINTEC, INC. CAPTUS Determine which wipe area the location fits into. Examples of locations and areas are: WORK AREA: Hot Lab, Sink UNRESTRICTED AREA: Patient Lavatory RECEIVING: Incoming Packages SEALED SOURCES: Co60 Reference Source Depending upon which area has been selected, a screen similar to Figure 12-5 will appear.
® CAPINTEC, INC. CAPTUS Input the “ROOM” and “DESCRIPTION” using letters and/or numbers. Press F1 [NUM] to switch to numeric input and display “LET” for F1. Press F1 [LET] to switch back to letter input and display “NUM” for F1.
Page 137
® CAPINTEC, INC. CAPTUS TRIGGER LEVEL/ 14:32 COUNT TIME 1. BACKGROUND 2. WORK AREAS 3. UNRESTRICTED AREAS 4. RECEIVING 5. SEALED SOURCES 6. GENERAL 7. PRINT ALL 0. RETURN TO WIPE SETUP MENU Figure 12-7 If BACKGROUND is selected, the trigger level is in cpm (cps if in Becquerel mode). If counting rate is greater than the trigger level during a background measurement, the message “BACKGROUND HIGH”...
® CAPINTEC, INC. CAPTUS in dpm, not cpm. This means only peaks for which a Conversion Factor (CF) is entered in memory will be compared to this trigger level. ® Counting Time determines how long the CAPTUS 600 will count during a Wipe Test.
® CAPINTEC, INC. CAPTUS Input new values if so desired. Press F4 [OK] to store the data. Press PRINT [P] to print a report of the Generic Conversion Factors. Press F1 [DONE] to exit and return to the Wipe and MCA Setup menu (Figure 12-3).
Page 140
® CAPINTEC, INC. CAPTUS ADDITIONAL NUCLIDES 13:06 NUCLIDE NAME: FE55 HALF LIFE : ENERGY : UNIT Figure 12-11 Press F1 [UNIT] to change the half-life unit from Minutes to Hours to Days to Years. Input the half-life of the nuclide and the energy of the main photopeak of the nuclide.
® CAPINTEC, INC. CAPTUS USER KEY ASSIGNMENT USER keys are active in the graphic display for Wipe Test and MCA. If a USER Key has a nuclide assigned to it, it works the same as a Pre-Set Nuclide key (e.g. Tc99m key).
Page 142
® CAPINTEC, INC. CAPTUS 13:06 ASSIGN NUCLIDE TO USER KEY #1 NUCLIDE NAME:> DONE Figure 12-14 Input the nuclide name. Press F4 [OK] to accept the name. Press F1 [DONE] to keep the previous assignment. 12-10 WIPE TEST SETUP November 07...
® CAPINTEC, INC. CAPTUS CHAPTER 13 WIPE TEST INTRODUCTION Before performing Wipe Tests, the Calibration Test must have been performed for the calendar day. Reference CHAPTER 6: ACCEPTANCE & QUALITY ASSURANCE TESTS for instructions on performing the Calibration Test. The room background must be measured in the Well for the 400 keV and 1600 keV scales once a day.
Page 144
® CAPINTEC, INC. CAPTUS 14:43 MEASURE BACKGROUND MAKE SURE NO SOURCES ARE NEARBY AFTER MEASUREMENT-- PRESS SAVE OR PRINT TO ACCEPT STAR Figure 13-2 If you need to measure background, remove all sources from well (or drilled crystal) and make sure no sources are nearby. Press the F4 [START] key.
® CAPINTEC, INC. CAPTUS The Function keys shown in Figure 13-3 operate as follows: • F1 [COUNT] Repeats the measurement without saving the background. • F2 [RANGE] Toggles the display between the spectra of the 1600 keV full scale range and the 400 keV full scale range.
Page 146
® CAPINTEC, INC. CAPTUS 13:06 RECEIVING WORK AREAS 12:50 W-01 HOT LAB SINK W-02 HOT LAB DOOR KNOB W-03 CAMERA TABLE NO WIPES IN THAT CATEGORY Press Any Key to Continue SELCT DONE Figure 13-7 Figure 13-6 Press the PRINT [P] key to obtain a printed summary of the wipes for the selected area.
® CAPINTEC, INC. CAPTUS Significant counts is defined as greater than one square root of the total counts detected in the overflow channel of the stored background count. WIPE RESULTS The Wipe Results Activity Screen (Figure 13-9) will appear. W-01 HOT LAB...
® CAPINTEC, INC. CAPTUS • F3 [GRAPH] Switches to the spectrum graph display, which allows detailed observation of the spectra taken on the 400 or 1600keV range. Reference Figure 13-10. • F4 [NEXT] Stores the data for this measurement in memory and proceeds to Figure 13-6 for the next wipe location.
® CAPINTEC, INC. CAPTUS Scale/Cursor Mode At any time when viewing the Wipe Screen, the system is either in Scale Mode or Cursor Mode. The S key toggles between these two modes. The default is Cursor Mode. This means the number to the right of the F4 key determines how many channels the cursor will jump with each press of the ARROW keys.
Page 150
® CAPINTEC, INC. CAPTUS Figure 13-11 13-8 WIPE TEST November 07...
® CAPINTEC, INC. CAPTUS INTERPRETING WIPE RESULTS The data shown in Figure 13-9 and Figure 13-12 are the results of a measurement with only Cs137. W-01 HOT LAB SINK Trig Level: 2000 HIGH CF: 1600 87.75 1.05 kdpm HIGH CS137 193.5...
® CAPINTEC, INC. CAPTUS CHAPTER 14 LAB TESTS SETUP INTRODUCTION ® The Lab Tests chapters describe protocols programmed into the CAPTUS 600 for use with certain lab tests based upon radiopharmaceuticals. These tests are: CATEGORY TEST NAME PHARMACEUTICAL URINE Schilling Test...
® CAPINTEC, INC. CAPTUS ® The CAPTUS 600 has the ability to store demographics for 50 patients in memory. This is the total number of patients and they can be distributed in any combination throughout the tests, except in the RBC Survival where only two tests may be recorded simultaneously.
® CAPINTEC, INC. CAPTUS 14:32 LA B TE S TS U R IN E B L O O D Figure 14-2 Press the F1 [URINE] key to access the Urine Test: Schilling Test. Press the F2 [BLOOD] key to access the three Blood Tests: Blood Volume (Cr51), Blood Volume (I125), RBC Survival.
® CAPINTEC, INC. CAPTUS To add a patient, press the F1 [NEW] key. To edit patient demographics, move the cursor next to the patient name and press the F2 [EDIT] key. Figure 14-5 will appear with the current data filled in.
Page 157
® CAPINTEC, INC. CAPTUS 14:32 S C H ILL IN G TE S T P A TIE N T, O N E M E A S U R E S TA N D A R D E X IT S T A R T Figure 14-6 Place the requested Standard into the Well and press the F4 [START] key to begin counting.
Page 158
® again. If you are certain the correct Standard or sample was in the Well and the CAPTUS 600 cannot measure it, contact the Capintec’s Authorized Service Center for assistance. When Background is being measured, no peak is expected. CALCULATING...
® CAPINTEC, INC. CAPTUS CHAPTER 15 LAB TESTS: URINE TEST INTRODUCTION This chapter describes the Schilling Test Urine Test. In the Main Lab Tests Screen, press the F1 [URINE] key. The Schilling Test Screen (Figure 15-1) will appear. 14:32 S C H ILLIN G TE S T...
Page 160
® CAPINTEC, INC. CAPTUS 14:32 SCHILLING TEST PATIENT, ONE COUNTING PERIOD: 300 SEC DILUTION FACTOR : 1000 ALIQUOT VOLUME : 4.0 m l URINE VOLUME : 1000 m l Figure 15-2 Input the correct values. The fields are defined as follows: •...
Page 161
® CAPINTEC, INC. CAPTUS After accepting the Co57 Standard measurement, the system will then prompt the user to measure the room background. Remove any samples from the Well and count room background. Next, measure the urine aliquot sample. After accepting the Urine Sample measurement, the Schilling Test Result Screen (Figure 15- 4) will appear.
Page 162
® CAPINTEC, INC. CAPTUS Figure 15-5 15-4 LAB TESTS: URINE TEST November 07...
® CAPINTEC, INC. CAPTUS CHAPTER 16 LAB TESTS: BLOOD TESTS INTRODUCTION This section describes the three blood tests: 1. Blood Volume (Cr51) 2. Blood Volume (I 125) 3. RBC Survival (Cr51) To access the Blood Test section of the program, press the F2 [BLOOD] key in the Main Lab Tests screen.
Page 164
® CAPINTEC, INC. CAPTUS 14:32 RBC VOLUME PATIENT, ONE DOSE HEMATOCRIT: 50 PATIENT HEMATOCRIT: 51 WEIGHT : 100.0 Figure 16-2 Input the data in the fields and press ENTER after each entry. The fields are defined as follows: • Dose Hematocrit Hematocrit of the tagged RBC suspension determined from the remainder of the suspension of the injection into the patient.
Page 165
® CAPINTEC, INC. CAPTUS • Plasma Standard 4 ml solution made from 1 ml of the plasma from the centrifuged tagged RBC suspension diluted with water to 100 ml • Room Background No samples in the Well • Whole Blood Sample 4 ml of blood withdrawn from the patient ten to twenty minutes post injection.
Page 166
® CAPINTEC, INC. CAPTUS Plasma Volume (ml) Whole Blood Volume (ml) Cell Volume (ml) To print the results press the PRINT [P] key. To return to the Main RBC Volume (Cr51) Test Screen, press the F4 [DONE] key. If you press the F4 [DONE] key before printing, you will be reminded that the data will be deleted.
Page 167
® CAPINTEC, INC. CAPTUS Figure 16-5 November 07 LAB TESTS: BLOOD TESTS 16-5...
® CAPINTEC, INC. CAPTUS BLOOD VOLUME (I125) The Blood Volume (I125) protocol is designed to be used in determining plasma volume. From the Main Blood Tests Screen (Figure 16-1), press 2 to select BLOOD VOLUME (I125). If patient demographics have not yet been input, do so according to the directions in CHAPTER 14: LAB TESTS: SETUP.
Page 169
® CAPINTEC, INC. CAPTUS Input the Injection Time in 24 hour format (example 1:00 PM as 1300) and press ENTER. Input the Counting Period, in seconds, for measurements. Press the F4 [OK] key. You will be prompted to count the following samples in the Well.
Page 170
® CAPINTEC, INC. CAPTUS Cell Volume Calculated Hematocrit Whole Blood Volume To print the results press the PRINT [P] key. To return to the Main Plasma Volume (I125) Test Screen press the F4 [DONE] key. If you press the F4 [DONE] key before printing, you will be reminded that the data will be deleted.
Page 171
® CAPINTEC, INC. CAPTUS Figure 16-9 November 07 LAB TESTS: BLOOD TESTS 16-9...
® CAPINTEC, INC. CAPTUS RED BLOOD CELL SURVIVAL This protocol is used to count samples taken as part of a procedure to measure Red Cell Survival time. The protocol is set up so that samples are counted as they are withdrawn. The ®...
Page 173
® CAPINTEC, INC. CAPTUS To make a measurement from the RBC Survival Patient Selection Screen, move the cursor to the desired patient and press the F4 [MEAS] key. Figure 16-11 will appear. RBC SURVIVAL 14:32 PATIENT, ONE HEMATOCRIT :> COUNTING TIME...
Page 174
® CAPINTEC, INC. CAPTUS displaying a message that reads “Data not Saved OK? (Y or N)”. If you escape by pressing the HOME key, the data is also lost. For measurements after the first one, Figure 16-13 will appear after room background is...
Page 175
® CAPINTEC, INC. CAPTUS Figure 16-14 November 07 LAB TESTS: BLOOD TESTS 16-13...
The CAPTUS 600 uses one bioassay protocol, which is user-defined. ® Before using the Capintec CAPTUS 600 system to perform Bioassay measurements on the ® staff, the staff member names, ID numbers and initials should be stored in the CAPTUS memory (if not, the results can't be saved), as well as the parameters for the Bioassay protocol.
® CAPINTEC, INC. CAPTUS SETUP 14:32 1. Ci/Bq 2. TEST DATA 3. THYROID UPTAKE 4. BIO ASSAY 5. AUTO CALIBRATION 6. ACTIVITY CONVERSION FACTORS 7. WIPE AND MCA SETUP 8. OTHER 0. RETURN TO UTILITIES MENU Figure 17-2 Press 4 to select BIOASSAY. The Bioassay Setup menu (Figure 17-3) will appear.
® CAPINTEC, INC. CAPTUS BIOASSAY STAFF DATA BIOASSAY STAFF DATA 14:32 14:32 > SEELEY SOPCHACK NAIM LEMISZKO EDIT DONE DONE Figure 17-4 Figure 17-5 Press the F1 [ADD] key. The Staff Data Screen (Figure 17-6) then appears. BIOASSAY STAFF DATA...
® CAPINTEC, INC. CAPTUS Use the ARROW keys to position the cursor at the line you need to change, input the revised data, and then press the ENTER key. When the information on the screen is correct, press the F4 [OK] key.
® CAPINTEC, INC. CAPTUS Input the desired Counting Period; the maximum accepted value is 999 seconds, and the minimum accepted value is 1 second. X-RAY/GAMMA RATIO These factors are necessary because the I125 X-Ray peak overlaps with the I131 X-Ray ®...
Page 182
® CAPINTEC, INC. CAPTUS 14:32 B IO A S S A Y IN P U T IN ITIA LS : B K G M E A S Figure 17-8 At this point, you may enter the initials for the staff member to be measured. You may then measure room background before measuring the staff member's neck.
® CAPINTEC, INC. CAPTUS At this point, you may abort the measurement by pressing the F1 [EXIT] key. If you are to continue with the measurement, first position the staff member in front of the Probe at the distance prescribed in Bioassay protocol. Press the F4 [START] key.
Page 184
® CAPINTEC, INC. CAPTUS 14:32 STAFF BIOASSAY REPORT FOR ALL STAFF MEMBERS Figure 17-11 If you want a report for only one staff member, press F1 [ONE]. Figure 17-12 will appear. 14:32 STAFF BIOASSAY REPORT FOR ONE STAFF MEMBER Figure 17-12 To return to the option of printing a report for all staff members, press F1 [ALL].
Page 185
® CAPINTEC, INC. CAPTUS 14:32 STAFF BIOASSAY SELECT > DOE SMITH Figure 17-13 Use the ARROW keys to move the prompt to the desired Staff Member and press ENTER. A report of the data for the chosen Staff Member will be printed.
® CAPINTEC, INC. CAPTUS CHAPTER 18 MULTICHANNEL ANALYZER SETUP INTRODUCTION ® The CAPTUS 600 is equipped with a general-purpose Multichannel Analyzer (MCA). The features of the MCA include: • Optional Automatic Background Subtract • Total/Net Reporting • Automatic Peak Search •...
Page 188
® CAPINTEC, INC. CAPTUS Press 3 to select SETUP. The message “PASSWORD:” will appear. Type the last three digits ® of your CAPTUS 600 serial number, and press the ENTER key. The Setup menu (Figure 18-2) will appear. SETUP 14:32 1.
® CAPINTEC, INC. CAPTUS MULTICHANNEL ANALYZER 14:32 SETUP > SETUP 200 DONE EDIT Figure 18-4 Setups that have been entered will be listed. If no setups have been entered, only the F1 [ADD} and F4 [DONE] keys will be displayed.
Page 190
® CAPINTEC, INC. CAPTUS MULTICHANNEL ANALYZER 13:06 SETUP ENERGY SCALE : 200 keV PEAK COUNTS COUNTING TIME :> 60 SEC X CUT OFF CHANNEL COUNTING TIME BASE: REAL BASE ENRGY Figure 18-6 The MCA counting can stop when the counting time is reached or when the Peak Counts (i.e.
Page 191
® CAPINTEC, INC. CAPTUS MULTICHANNEL ANALYZER 13:06 SETUP ENERGY SCALE : 200 keV PEAK COUNTS COUNTING TIME :> 60 SEC CUT OFF CHANNEL COUNTING TIME BASE: REAL PEAK ENRGY Figure 18-7 Press F2 [PEAK] to scroll Peak Counts through the possible values (from 100 to 1024k) When all entries are correct, press F4 [OK].
® CAPINTEC, INC. CAPTUS MULTICHANNEL ANALYZER 13:06 SETUP DONE Figure 18-9 Press PRINT [P] to print a summary of the Setup. Press F4 [DONE] when entering the setup is finished. You will be returned to the list of existing setups.
® CAPINTEC, INC. CAPTUS CHAPTER 19 MULTICHANNEL ANALYZER OPERATION MCA MEASUREMENTS The Multichannel Analyzer can be used for measurements in the Well or Probe. Press the HOME key and then the UTIL key. The main Utilities menu (Figure 19-1) will appear.
® CAPINTEC, INC. CAPTUS Press the key corresponding to the desired input. Figure 19-3 will appear. 14:32 M U LTIC H A N N E L A N A LY ZE R 1. M E A S U R E S O U R C E 2.
Page 195
® CAPINTEC, INC. CAPTUS MULTICHANNEL ANALYZER 13:06 BACKGROUND SETUP > 200 KEV 400 KEV 800 KEV 1600 KEV Figure 19-4 Use the ARROW keys to move the prompt to the desired range. Press F1 [SEL] to select a range. The range will be highlighted and the prompt will move to the next range. If the prompt points to an already selected range, F1 will change to [DESEL] and pressing it will de-select the range.
® CAPINTEC, INC. CAPTUS 100 Cts FS 256 keV TOTAL lt: 10.0 sec 400 keV COUNT RANGE SAVE SC/CR Figure 19-6 The function keys on this screen operate as follows: • F1 [COUNT] Re-counts the background without saving the current measurement.
Page 197
® CAPINTEC, INC. CAPTUS MULTICHANNEL ANALYZER 13:06 SETUP ROI NUMBER: 1 FROM CHANNEL: > CHANNEL: START Figure 19-7 Press F4 [START] to begin the measurement. If the background has not been measured for the chosen energy scale during the current...
Page 198
® CAPINTEC, INC. CAPTUS 500 Cts FS 250 keV CC:21 TOTAL lt: 9.9sec 400 keV COUNT SC/CR Figure 19-8 Cursor Vertical Full Scale (Counts) Cursor Position Counts at Cursor Indicates whether Total or Net Counts are being displayed. Indicates the Live Time of the measurement.
Page 199
® CAPINTEC, INC. CAPTUS To use the ROI Function: 1. Press the F3 [ROI] key – The Normal MCA cursor will become a full size cursor extending to the full height of the display. This is the ROI cursor. 2. Move the cursor to one side of the desired ROI with the ARROW keys and press the ENTER key.
® CAPINTEC, INC. CAPTUS Activity of a Nuclide The feature is only available if Calibration Test has been performed during the current calendar day. The counts in a Region of Interest corresponding to a given nuclide can be obtained by pressing one of the Pre-Set Nuclide keys ( Cr51 , Co57 , Ga67 , Tc99m , In111 , I131 , I123 , I125 , Tl201 ), a USER key which has had a nuclide assigned to it or the NUCL key.
It is recommended that periodic (every five years) re-calibration of the CAPTUS 600 be performed by Capintec’s Authorized Service Center to guarantee the instrument’s high reliability is maintained. Contact the Capintec’s Authorized Service Center in Pittsburgh for servicing or re-calibration at 1-800-227-6832. CLEANING INSTRUCTIONS CAUTION DISCONNECT THE POWER BEFORE CLEANING.
If the unit fails to pass Auto Calibration Test or any other test, the user should not attempt to perform any adjustments to the system. In this event, please contact Capintec’s Authorized Service Center for further assistance.
• excessive play between the Collimator pin (3) and the Collimator Mounting Block. If there is excessive play or the knob(s) cannot be tightened, contact Capintec’s Authorized Service Center for a information on replacement items, instructions or on-site service. Figure 20-1...
CAPTUS DISPOSAL The following items should be taken into consideration before disposing. These items should be disposed of in accordance with local and national regulations. Please contact Capintec, Inc. or an authorized disposal company to decommission your equipment. Figure 20-2...
® CAPINTEC, INC. CAPTUS FUSE SERVICING CAUTION For continued protection, replace only with same type and rating of fuse(s) Power Line Fuses The power line fuses are located on the front side of the Isolation Transformer Unit directly below the main power cord connector. The type and rating of these 2 fuses are listed on the label that is located on the top of the Isolation Transformer Unit.
® CAPINTEC, INC. CAPTUS Readout Unit Fuses Two fuses are located in the power entry module next to the power cord connector on the back panel of the Readout Unit. The type and rating of these fuses are listed on the label located on the back panel of the Readout Unit.
CAPINTEC, INC. CAPTUS ACCESSORIES AND REPLACEMENT PARTS The following accessories and replacement parts are available from Capintec. Call Capintec at 1-800-227-6832 for answers to your questions or to place an order. • Printer Paper: Fan fold paper, 1000 sheets per box....... 9282-0007 •...
Page 211
® CAPINTEC, INC. CAPTUS RBC Survival Measurement, 3-12 Nuclides Staff Data, 3-13 Additional, 12-7 Thyroid Uptake Measurements, 3-11 Thyroid Uptake Protocols, 3-11 Wipe, 3-12 liners, 1-5 Offset Voltage adjustment, 3-10 Live Time, 3-9, 7-1 One Capsule Method, 2-8 log-on password, 3-1...
Page 212
® CAPINTEC, INC. CAPTUS Counting Methods, 7-1 General Instructions, 8-1 Measurements RBC Survival Test, 2-10 input limits, 3-11 Measurement input limits, 3-12 Measuring Neck and Thigh, 8-7 Readout Unit, 2-4 Nuclides, 2-7 Real Time, 3-9, 7-1 One Capsule Method, 2-8...
Page 213
® CAPINTEC, INC. CAPTUS General Procedure, 10-1 Activity for Generic Conversion Factor, Measuring Reference Capsule(s), 10- 13-7 Activity of a Nuclide, 13-7 Pre-Dose Measurement, 10-1 areas, 12-3 Viewing Results, 10-7 Background, 12-5, 13-1 view patient information, 8-5 too high, 13-2...
Page 215
NEW PRODUCT WARRANTY Conditions and Limitations CAPINTEC warrants our products to be free from defects in material and workmanship for a period of 12 months after delivery to the original buyer (unless indicated otherwise). Our obligation under this warranty is limited to...