WINNCARE FRANCE TABLE OF CONTENTS INTRODUCTION ....................................- 3 - Necessary information for patients and their families ............- 3 - Instructions for use ........................... - 3 - Indications ............................- 4 - Contraindications ........................... - 4 - Precautions ............................- 4 - Warning..............................
WINNCARE FRANCE INTRODUCTION ECESSARY INFORMATION FOR PATIENTS AND THEIR FAMILIES This product helps to prevent and treat pressure ulcers. Why have you been prescribed this support? Your health condition reduces your mobility and exposes you to the risk of pressure ulcers.
WINNCARE FRANCE NDICATIONS According to the opinion of the National Commission for the Evaluation of Medical Devices and Health Technologies (CNEDiMTS) dated 22 December 2009 and in accordance with the expert opinions of clinicians: Prevention and support for treatment of stage 1 to 4 pressure ulcers (according to medical advice) for patients up during the day, confined to bed for over 15 hours and “...
WINNCARE FRANCE Products in the AXTAIR motorized air range are therapeutic supports in the context of the CEI standard 60601-2-52 regarding the evaluation of protection against mechanical danger from Electro-Mechanical devices (EM) and EM systems, particularly against trapping of the patient in non-mobile parts. The requirements in figures 201.107, 201.108 and table 201.101 involving Axtair type therapeutic mattress are excluded.
WINNCARE FRANCE ONTENT OF PACKAGING FOR MATTRESS TOPPER AND MATTRESS TYPE SUPPORTS 1 rolled support in a transport bag 1 dirty/clean label 1 identification label 1 integrated compressor in a rolled support 1 electrical power supply cable with 2 attachment systems to long sides of a medical bed ...
Page 8
WINNCARE FRANCE Connect the rapid inflation pump to the CPR. Keep the CPR and pump in the same axis in order to avoid arching the CPR by the weight of the pump. Incorrect use could cause a water tightness defect for the CPR.
Page 9
WINNCARE FRANCE Use: Shows that the compressor is switched on Flashes to indicate inflation of the support After it goes off, the patient can be installed The on LED shows that the keyboard is locked: it is not possible to change mode.
WINNCARE FRANCE Key press: stops sound alarm In the event of a medium priority alarm, the sound alarm is reactivated after 3 minutes. The following conditions of use must be followed: Temperature: between +15°C and + 40°C Hygrometry level: between 30% and 95% ...
WINNCARE FRANCE INSTALLING THE SUPPORT Disconnect the compressor from the mains Open the CPR valve to deflate the support. Complete removal of air from the support can be facilitated by using the inflation/deflation kit connected to the CPR, and obstructing the power supply pipe., Position the compressor in the middle of the support.
WINNCARE FRANCE WARNING: Keep the device away from humidity, direct sunlight, continuous heat source. Prevent the risk of shock or alteration from sharp objects. NOTE: The MANUFACTURER will provide, upon request, the circuit schemas, list of components, descriptions, calibration instructions or any other relevant information to MAINTENANCE STAFF to perform authorized repairs in accordance with the contract that binds the requester to the manufacturer ASKLÉ...
WINNCARE FRANCE UPPORTS On the support, two zippers inside the cover at the head and foot of the support are used to separate the elements quickly. All the cells and the maintenance sheet remain together and must be manually disinfected: use a cloth soaked in a detergent solution or CE* labelled surface detergent/disinfectant at usage concentrations recommended by the manufacturer.
WINNCARE FRANCE ALARMS REPAIR The following table shows all the alarms for the device, their triggering conditions, their priority as well as possible repair. Description Priority Trigger Possible repair Condition Time frame 1 - Re-connect the compressor Alarm LED flashing 2 - Check the status of the and mains present Electrical fault:...
WINNCARE FRANCE WARNING: After accidental deflation of the support due to an extended power cut, at re- inflation, ensure that the patient has no limbs stuck between the support and the bed barriers. MAINTENANCE - REVISION REQUENCY OF REVISIONS it is recommended revising the AUTOMORPHO PLUS compressor at least every 2 years of use. To facilitate the management and scheduling of these revisions running time counters show the compressor usage time.
WINNCARE FRANCE Thank you for your public citizen contribution to protecting the environment. GUARANTEE The compressor and its supports are guaranteed for 2 years against any manufacturing defect from the date of purchase and under the recommended usage conditions outlined in the usage leaflet. This guarantee cannot substitute legal guarantees.
WINNCARE FRANCE TECHNICAL CHARACTERISTICS CHARACTERISTICS OF: MATTRESS TOPPER SUPPORTS References Supports from KITS Supports from KITS VAXT4/AUTO-P VAXT4/CIC100-P VAXT4/CIC-P VAXT4/AUTO100-P VAXT4/AUTO120-P Validated user weight – – 165 kg 165 kg Support weight 8 Kg 10 kg Support size 195 x 87 x 17 cm 195 x 87 x 17 cm 195 x 97 x 17 cm 195 x 117 x 17 cm...
The Axtair AUTOMORPHO PLUS requires special precautions with respect to CEM and it must be installed and serviced according to CEM information provided upon request from ASKLE SANTE and available on their website. The use of accessories, and cables other than those supplied and specified by ASKLE SANTE can result in increased emissions from the Axtair AUTOMORPHO PLUS or reduced immunity, affecting its operation, as well as its performance.