D. Pause or Stop Treatment Manually: During treatment or current ramp
up, turn the CURRENT knob counterclockwise to reduce current output
to 0.0 mA, or press the CURRENT knob for automatic ramp down. The
display will flash "PAUSE", and the dose delivered prior to the pause is
retained in memory.
EXAMPLE: After decreasing current to 0.0 mA display shows:
NOTES:
1.
During a pause, electrodes
may be disconnected, replaced or removed. Dose and/or current
may also be adjusted.
2.
To restart the treatment, turn the CURRENT knob clockwise to the
desired current level.
3.
Turn the ON/DOSE/START
knob clockwise to start.
The unit will automatically
recalculate the treatment
time, ramp up and
continue treatment.
E.
Stop Treatment Automatically: After the preset dose is reached, the
current automatically ramps down to 0.0 mA.
After a 40 mA-minute treatment is complete the unit beeps and the display
shows:
NOTES:
1.
After Automatic current ramps down, turn the current knob to
silence the beep, or depress the ON/DOSE/START knob to turn
OFF.
2.
Turn CURRENT knob counterclockwise to 0.0 mA anytime to stop
treatment, or depress the ON/DOSE/START knob.
F.
After Treatment:
Remove electrodes: Disconnect the lead wire clips and remove the
electrodes from the patient. Discard both the drug electrode and the
dispersive pad. They cannot be reused.