Input Voltage Configuration Warning - Garner TS-1 Series Operation And Installation Manual

Continuous degaussing system
Hide thumbs Also See for TS-1 Series:
Table of Contents

Advertisement

Input Voltage Configuration
The TS-1 model is equipped with a voltage selection block allowing the TS-1 to be
configured for use in any country.
To check or change the voltage configuration of the TS-1, follow the steps listed below:
All trouble shooting and wiring checks must be performed by a
qualified electrician or authorized plant personnel.
If you are installing the TS-1 for the first time, or if the TS-1 has moved to a new
location, confirm the input power setting of the TS-1 is configured to match the power at
your location**.
Step 1
Step 2
Step 3
Step 4
**If the power at your location is prone to voltage sag, the TS-1 may/will report a
"CHARGE ERROR" fault. This is an operational feature designed to ensure the
degausser will only operate when an optimal charge condition is present. If the
charge error fault persists when the power is known to be stable, call customer
support.
TS-1 Operation and Installation Manual
WARNING
Changing the Voltage Configuration
Un-plug the TS-1 from the power mains.
Loosen and remove the Power Selection Panel
from the rear panel. (4) T-20 Torx Screws
Move the (brown) power selection wire (located on
the right side of the fuse block) to the appropriate
tap/fuse on the right side of the fuse block. The
picture below shows the brown wire on the
120VAC connector.
Replace the Power Selection Panel.
Page 6