DL9110 / DL9118
6.4 Line Impedance
Supply Voltage
Nominal Test Current
Display Range
Measuring Range
Resolution
Accuracy
PFC Range
6.5 RCD
Supply Voltage
Nominal Test Currents
Test Current Accuracy
Trip Time Ranges
Trip Time Accuracy
Ramp Current Range
Trip Current Measurement
Accuracy
6.6 Voltage/Frequency Measurement
Display Range
Voltage Measuring Range
Resolution
Accuracy
Frequency Range
Frequency Accuracy
195V – 253V, 45Hz – 65Hz
328V - 440V, 45Hz – 65Hz
3A
0.01Ω - 2000Ω
0.1Ω - 1999Ω
0.01Ω maximum
±5% ±5 digits
0kA – 26kA
195V – 253V, 45Hz – 65Hz
10mA, 30mA, 100mA, 300mA,
500mA
-0% +10% at I∆n and 5I∆n
-10% +0% at ½ I∆n
0ms – 2000ms, ½ I∆n
0ms – 300ms, I∆n General
0ms – 500ms, I∆n Selective
0ms – 40ms, 5I∆n
±5% ±2 digits
½ I∆n to 1.1 I∆n
10%
0V – 440V
0V – 440V
1V
±5% ±2 digits
45Hz – 65Hz
±1Hz