Technical Information

NOTE: This product complies with ISO 10993-1, Biological Evaluation of Medical Devices Part 1:
Evaluation and Testing.
CAUTION: A functional tester cannot be used to assess the accuracy of a pulse
!
oximeter monitor or sensor.
CAUTION: All parts and accessories connected to the serial port of this device must
!
be certified according to at least IEC Standard EN 60950, IEC 62368-1, or UL 1950
for data-processing equipment.
CAUTION: Portable and mobile RF communications equipment can affect medical
!
electrical equipment.

Manufacturer's Declaration

Refer to the following table for specific information regarding this device's compliance to
IEC 60601-1-2.
Emissions Test
This device is intended for use in the electromagnetic environment specified below. The user of this device
RF Emissions
CISPR 11
RF Emissions
CISPR 11
Harmonic Emissions
IEC 61000-3-2
Voltage Fluctuations/Flicker
Emissions
IEC 61000-3-3
34
Table 7: Electromagnetic Emissions
Compliance
should ensure that it is used in such an environment.
Group 1
Class B
N/A
N/A
Electromagnetic Environment—Guidance
This device uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in
nearby electronic equipment.
This device is suitable for use in all establishments,
including domestic and those directly connected to
the public low-voltage power supply network that
supplies buildings used for domestic purposes.
Technical Information