Nonin LifeSense Vet LS1-10R Operator's Manual

Veterinary vital signs monitor capnograph/pulse oximeter
Table of Contents

Advertisement

Quick Links

Operator's Manual
LifeSense
Vet
®
LS1-10R
Veterinary Vital Signs Monitor
Capnograph/Pulse Oximeter
English

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the LifeSense Vet LS1-10R and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Nonin LifeSense Vet LS1-10R

  • Page 1 Operator’s Manual LifeSense ® LS1-10R Veterinary Vital Signs Monitor Capnograph/Pulse Oximeter English...
  • Page 2 Consult Instructions for Use. Nonin makes no claim for use of the product other than those uses specified herein and disclaims any liability resulting from other uses. Observe all warnings, cautions, and notes. Unauthorized use, copying, or distribution of this manual is prohibited without written consent from Nonin Medical, Inc.
  • Page 3: Table Of Contents

    Contents Indications for Use ..................1 Contraindications....................1 Warnings ........................ 1 Cautions ......................... 2 Guide to Symbols ..................5 Introduction....................6 About LifeSense Vet....................6 About Capnography....................6 About Pulse Oximetry..................... 6 Operator Requirements ..................7 Displays and Controls................... 8 Monitor Front Views....................
  • Page 4 Contents (Continued) Changing Settings....................26 Alarms ........................26 Alarm Function....................26 Alarm Silence.....................26 High Priority Alarm.....................27 Low Priority Alarm....................27 Disable Alarms....................28 Maintenance and Inspection ...............29 Battery Operation ....................29 Charging the Battery..................29 Checking Battery Capacity ................29 Battery Message....................29 Battery Care.......................29 Maintenance......................30 Ensuring Optimal Performance................30 Cleaning the Sensor ..................30 Cleaning the Monitor..................30 Calibration ......................30...
  • Page 5 Figures Figure 1. Start-up Screens ..................8 Figure 2. Operating Screen ..................10 Figure 3. Trend Screen .................... 13 Figure 4. Rear View of Monitor................. 14 Figure 5. Right Side of Monitor................. 16 Figure 6. Replacing the Moisture Trap/Filter ............21...
  • Page 6 Tables Table 1. Start-up Screen Icons and Display Descriptions.......... 9 Table 2. Device and Operating Screen Icons and Display Descriptions....10 Table 3. Trend Screen Icons and Display Descriptions ........... 13 Table 4. Rear View Features and Descriptions ............14 Table 5.
  • Page 7: Indications For Use

    (see “Accessories”). Only use power supplies that are either supplied with LifeSense Vet or specified by Nonin (see “Accessories”). Prior to connecting LifeSense Vet to the power supply and the power outlet, be sure to verify the voltage and frequency rating on the power supply are compatible with the outlet.
  • Page 8: Cautions

    If you are unsure how to reach your distributor, please call Nonin for your distributor’s contact information. Always turn off and unplug the monitor prior to cleaning the monitor or changing the Nonin-branded PureLight veterinary sensor, or moisture trap and/or filter.
  • Page 9 (33 – 43 mmHg or 4.4 – 5.7 kPa) and no airway or circuit air leak has been detected, an internal air leak is possible. Replace the moisture trap and perform the zero point calibration procedure. If the problem persists contact Nonin Technical Service and discontinue use of the LifeSense Vet monitor.
  • Page 10 Cautions (Continued) Certain pharmacologic agents used to sedate or anesthetize animals may have cardiovascular effects that can adversely affect the performance of the pulse oximeter by reducing the perfusion to the sensor site. Examples of commonly used agents that may have this type of effect on certain animal species are Detomidine HCl and Xylazine HCl.
  • Page 11: Guide To Symbols

    Charging indicator. This indicator is green when the monitor is connected to a power outlet. DC input. Used for connecting the power supply. Serial interface port. l O l O l 2 Input used for connecting a Nonin-branded PureLight SpO sensor. NONIN SpO Indoor use only Class II equipment, double insulated...
  • Page 12: Introduction

    PureLight veterinary sensor, provided with the system. Use only those accessories and replacement parts recommended by Nonin. Refer to the “Accessories” section for more information. LifeSense Vet has visual and audible alarms when readings are outside the predefined limits.
  • Page 13: Operator Requirements

    Operator Requirements The LifeSense Vet monitor is easy to operate. Each operator should read this manual before using the monitor. LifeSense Vet should only be operated by a qualified professional.
  • Page 14: Displays And Controls

    Displays and Controls A standard LifeSense Vet set-up consists of a monitor, moisture trap with filter, Nonin-branded PureLight Lingual clip sensor (Model 2000SL), 1-meter extension cable (UNI-EXT-1), sampling tube, straight T-connector, manual, and power supply. See “Accessories” for information on optional accessories.
  • Page 15: Table 1. Start-Up Screen Icons And Display Descriptions

    Table 1: Start-up Screen Icons and Display Descriptions Name Description Audible Alarm Disable Pressing this icon turns the audible alarms off. It disables the audible alarms by setting all lower limits to 0. Audible Alarm Enable Pressing this icon turns the audible alarms on. Default if no icon is chosen.
  • Page 16: Figure 2. Operating Screen

    Figure 2: Operating Screen Table 2: Device and Operating Screen Icons and Display Descriptions Name Description LCD Display The LCD monitor displays parameters, graphs, menus and other information. It is also a touch panel from which all the operator-defined settings are made.
  • Page 17 Table 2: Device and Operating Screen Icons and Display Descriptions (Continued) Name Description Audible Alarm Audible alarms alert the operator when readings are outside the preset Pause/Resume limits. The operator can temporarily disable the audible alarm by pressing this button. The audible alarm will remain inactive for 2 minutes unless the operator presses the Audible Alarm Pause/Resume button again.
  • Page 18 Table 2: Device and Operating Screen Icons and Display Descriptions (Continued) Name Description 14. Trend Icon Touch this icon to display the trend screen. The trend screen remains visible until the operator touches the screen. 15. Pulse Oximetry Displays a graph giving information on the oximetry signal Plethysmograph (plethysmograph).
  • Page 19: Figure 3. Trend Screen

    Figure 3: Trend Screen Table 3: Trend Screen Icons and Display Descriptions Name Description Trend HR Displays a trend graph of the pulse rate. This scale is fixed and cannot be changed. Trend SpO Displays a trend graph of the SpO values.
  • Page 20: Monitor Rear View

    Monitor Rear View The moisture trap, filter, equipment label, and gas outlet are located on the back of the LifeSense Vet (figure 4). Names and descriptions of each component are listed in table 4. WARNING: Blocking the gas outlet at the back of the monitor may result in inaccurate readings.
  • Page 21 Table 4: Rear View Features and Descriptions (Continued) Name Description Equipment Label The label contains the model number, serial number, manufacturer, UL mark, and other applicable symbols. See “Guide to Symbols” for descriptions of the different symbols. Every LifeSense Vet device has a unique serial number for identification.
  • Page 22: Monitor Right Side View

    No other sensors may be used. WARNING: Use only Nonin-branded PureLight veterinary pulse oximeter sensors. These sensors are manufactured to meet the accuracy specifications of Nonin pulse oximeters. Using other manufacturers’ sensors can result in improper pulse oximeter performance.
  • Page 23: Using The Lifesense Vet Monitor

    Using the LifeSense Vet Monitor After unpacking the monitor and accessories, LifeSense Vet is ready for use. Ensure the LifeSense Vet battery is fully charged by viewing the status of the battery indicator on the display panel after the power supply is connected to the monitor and the power outlet. CAUTION: To prevent damage to the monitor, operate and store the monitor in an upright position.
  • Page 24: Mounting

    If the pole is mobile, do not attach the monitor to the pole higher than 1.5 meters (5 feet) and do not exceed a total of 2 kilograms (4.5 pounds) of equipment on the pole. Contact Nonin Customer Support to order a mounting bracket and adjustable mounting clamp (see the “Accessories” section).
  • Page 25 2. Connect the Nonin-branded PureLight veterinary pulse oximeter sensor (with the Nonin logo facing toward you) into the NONIN SpO connector on the right side of the LifeSense Vet. 3. Ensure that the sensor is firmly plugged in. 4. Position the appropriate sensor on the animal.
  • Page 26: Sample Line

    2. Connect the Luer lock fitting to the straight T-connector, turn to tighten. 3. Connect the Luer lock fitting to the moisture trap, turn to tighten. WARNING: Use only Nonin-recommended accessories and replacement parts. CAUTION: The sample line is a single-use, disposable component. Use a new sample line for each animal.
  • Page 27: Trend Screen

    4. To remove the moisture trap and replace the filter, reverse the three steps as above. Figure 6: Replacing the Moisture Trap/Filter Trend Screen The trend screen displays up to 4 hours of trending data for pulse rate, SpO , respiration, and EtCO .
  • Page 28: Getting Started

    Use only Nonin-branded PureLight veterinary pulse oximeter sensors (see “Accessories”). These sensors are manufactured to meet the accuracy specifications of Nonin veterinary pulse oximeters. Using other manufacturers’ sensors can result in improper pulse oximeter performance.
  • Page 29: Disconnect The Animal

    The audible alarm function activates approximately 2 minutes after start up, unless activated by the operator before then. The monitor is now ready for use. The animal can stay connected to the monitor for as long as needed. CAUTION: Inspect the pulse oximeter sensor application site periodically to ensure correct sensor alignment and skin integrity.
  • Page 30: Settings And Alarms

    Settings and Alarms Touch Panel Display All adjustments and settings are made using the LifeSense Vet touch panel display. Each specific parameter is adjusted by using the up/down arrows on the display bar Factory Default Settings LifeSense Vet recalls and displays the factory default settings (table 6) upon start-up. At the start- up screen, the operator can select from 3 different default settings (only if alarms are activated on the first start-up screen).
  • Page 31: Alarm Limits

    Alarm Limits All parameters have built in limits that cannot be exceeded. Pulse Limits – Upper limit: 450 BPM – Lower limit: 0 BPM Limits – Upper limit: 100 % – Lower limit: 0 % Respiration Limits – Upper limit: 93 RPM –...
  • Page 32: Changing Settings

    Changing Settings All settings follow the same procedure to increase or decrease an alarm limit. • The up arrow on the right side of a displayed parameter bar is used to increase an alarm limit. • The down arrow on the left side of a displayed parameter bar is used to decrease an alarm limit.
  • Page 33: High Priority Alarm

    High Priority Alarm A high priority alarm calls for immediate action from the operator. An alarm (table 7) occurs if any of the parameters are outside the operator-defined limits (or default alarm limits if operator-defined limits have not been set). High priority alarms are both audible and visual: •...
  • Page 34: Disable Alarms

    Warm up delay and stabilizing measurements. System NO OXIMET No communication from the pulse oximetry unit. Possibly due to a sensor error. System NO CAPNO No communication from capnography unit. Contact Nonin Technical Service. System BATT LOW Battery is almost depleted. System DISP ERROR Touch panel display is not working properly.
  • Page 35: Maintenance And Inspection

    The battery, made of Lithium Ion (Li-Ion) rechargeable cells, is integral to the device and cannot be replaced by anyone other Nonin Technical Service. The battery warranty is 1 year. For optimal performance, the battery should be replaced once per year to limit the amount of Li...
  • Page 36: Maintenance

    Refer to individual sensor Instructions for Use for details. Cleaning the Monitor Clean the Nonin LifeSense Vet monitor with a soft cloth moistened with isopropyl alcohol. Allow the monitor to dry completely after cleaning. CAUTION: Do not sterilize, autoclave, or use caustic or abrasive cleaning agents on the LifeSense Vet monitor or sensors.
  • Page 37: Recommended Inspections And Functional Check

    “Capnography Specifications” section. NOTE: If the reading is out of range, an internal air leak is possible. Replace the moisture trap and repeat the calibration procedure. If the out of range reading continues, contact Nonin Technical Service. Recommended Inspections and Functional Check 1.
  • Page 38 WARNING: Never allow liquids to enter into or to be spilled onto the monitor. If liquid has penetrated the monitor, it must be checked by Nonin Technical Service. CAUTION: Never open the monitor housing/case. By opening the case you will...
  • Page 39: Troubleshooting

    The fault conditions are either operator- or system-generated. The table below lists common messages, descriptions and advice on actions to take. If the problem persists, contact Nonin Technical Service. Message Description...
  • Page 40: Troubleshooting

    The monitor is not Recharge the monitor with the power functioning. This indicates supply. that a problem has occurred, If the problem persists, contact Nonin possibly due to interference Technical Service. or loss of power. Low EtCO alarm All alarms for low EtCO Check animal status.
  • Page 41: Accessories

    The following list of accessories can be ordered from Nonin or your distributor. Nonin may update the accessories list at any time. It is the purchaser’s responsibility to ask for the current list when ordering.
  • Page 42: Technical Information

    WARNING: The use of accessories, sensors, and cables other than those specified in this manual may result in increased emission and/or decreased immunity of this device (see “Accessories”). WARNING: Only use power supplies that are either supplied with LifeSense Vet or specified by Nonin (see “Accessories”).
  • Page 43: System Specifications

    System Specifications Power Data Power Supply: 100 – 240 VAC 50 – 60 Hz Power Consumption: 3.6 W with battery operation 9 W with power supply Input: 12 VDC, 720 mA Battery Data Type: Lithium Ion (Li-Ion) internal battery, non-field replaceable, rechargeable Battery Capacity Approximately 8 hours Charging Time: Approximately 17 hours, or 2 hours for each hour of use...
  • Page 44: Pulse Oximeter Specifications

    Red: 660 nanometers, 0.8 mW max. average PureLight Sensors: Infrared: 910 nanometers, 1.2 mW max. average Saturation Accuracy (A 70 % to 100 % ±2% of full scale using Nonin clip sensor Capnography Specifications Respiration Range: 3 to 60 RPM...
  • Page 45: Manufacturer's Declaration

    Manufacturer’s Declaration See the following tables for specific information regarding this device’s compliance to IEC 60601-1-2. Table 10: Electromagnetic Emissions Emissions Test Compliance Electromagnetic Environment—Guidance RF Emissions Group 1 This device uses RF energy only for its internal function. Therefore, its RF emissions are very low CISPR 11 and are not likely to cause any interference in nearby electronic equipment.
  • Page 46: Table 11. Electromagnetic Immunity

    Table 11: Electromagnetic Immunity Electromagnetic Immunity Test IEC 60601 Test Level Compliance Level Environment—Guidance This device is intended for use in the electromagnetic environment specified below; the user of this device should ensure that it is used in such an environment. Electrostatic Discharge ±6 kV contact ±6 kV contact...
  • Page 47: Table 12. Guidance And Manufacturer's Declaration-Electromagnetic Immunity

    Table 12: Guidance and Manufacturer’s Declaration—Electromagnetic Immunity IEC 60601 Electromagnetic Immunity Test Compliance Level Test Level Environment—Guidance This device is intended for use in the electromagnetic environment specified below; the user of this device should ensure that it is used in such an environment. Portable and mobile RF communications equipment should be used no closer to any part of the device, including cables, than the recommended separation distance calculated from the equation applicable to2 the frequency of the transmitter.
  • Page 48: Table 13. Recommended Separation Distances

    Table 13: Recommended Separation Distances This device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. Customers or users of this device can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communication equipment (transmitters) and the device as recommended below, according to maximum output power of the communications equipment.
  • Page 49: Service, Support, And Warranty

    Nonin shall repair or replace any LifeSense Vet found to be defective in accordance with this warranty, free of charge, for which Nonin has been notified by the purchaser by serial number that there is a defect, provided said notification occurs within the applicable warranty period. This warranty shall be the sole and exclusive remedy by the purchaser hereunder for any LifeSense Vet delivered to the purchaser which is found to be defective in any manner, whether such remedies be in contract, tort, or by law.

Table of Contents