Download Print this page

SMART 433-CMU Installation And Use Instructions Manual page 6

Wireless central monitor and nurse call paging system

Advertisement

CordLess® Transmitter (
a
Plug a CordLess® Transmitter into a CordLess®
TM-02
b
1 2 3 4 5 6 7 8 9 1 0
11 12 13 14 15 16 17 18 19 20
21 22 23 24 25 26 27 28 29 30
WIRELESS CALL SYSTEM
433-CMU
ON/OFF Switch
c
TM-03
CordLess®
Transmitter
TM-03
433-EXT Door/Window Exit Alarm (optional) Programming Instructions
a
1 2 3 4 5 6 7 8 9 1 0
11 12 13 14 15 16 17 18 19 20
21 22 23 24 25 26 27 28 29 30
WIRELESS CALL SYSTEM
433-CMU
b
433-EXT
WARNING: If you do not wait for the Central Monitor to make a melodic tone before pressing the next Nurse Call Button or other
component, each component will be programmed to the same number on the Central Monitor. In this situation, the Central Monitor will
speak one number for the component, but show a different number to represent the room. When multiple components are programmed to
the same number on the Central Monitor, THE NUMBER SPOKEN WILL NOT CORRESPOND WITH THE NUMBER DISPLAYED. This may be
useful when multiple devices are being used in the same room, but not for any other situation.
6
) (optional) Programming Instructions
TM-03
Sensor Pad or Floor Mat. Repeat this step for
every CordLess Pad and Transmitter
you plan to use with the Central
Monitor system.
Press and hold "Record/
CALL REVIEW
Add" button until you
VOLUME
LANGUAGE
hear 2 dings and a
REPLACE/ADD
RECORD/ADD
melodic tone.
RESET
Turn on the transmitter. The
ON/OFF switch located on the
side of the Transmitter.
Press and hold "Record/
CALL REVIEW
Add" button until you hear
VOLUME
LANGUAGE
2 dings and a melodic
REPLACE/ADD
RECORD/ADD
tone.
RESET
Touch the two parts of the Exit
Alarm together and then the Central
Monitor will display and speak the
number the Exit Alarm has been
programmed to.
d
e
1 2 3 4 5 6 7 8 9 1 0
11 12 13 14 15 16 17 18 19 20
21 22 23 24 25 26 27 28 29 30
WIRELESS CALL SYSTEM
433-CMU
f
1 2 3 4 5 6 7 8 9 1 0
11 12 13 14 15 16 17 18 19 20
21 22 23 24 25 26 27 28 29 30
WIRELESS CALL SYSTEM
433-CMU
c
1 2 3 4 5 6 7 8 9 1 0
11 12 13 14 15 16 17 18 19 20
21 22 23 24 25 26 27 28 29 30
WIRELESS CALL SYSTEM
433-CMU
d
1 2 3 4 5 6 7 8 9 1 0
11 12 13 14 15 16 17 18 19 20
21 22 23 24 25 26 27 28 29 30
WIRELESS CALL SYSTEM
433-CMU
Press on the sensor pad or step on floor
mat. The Central Monitor will speak and
display a number. The number shown is the
Cordless Transmitter's identification number
on the Central Monitor.
Wait for the Central Monitor
CALL REVIEW
to make a melodic tone. Then,
VOLUME
LANGUAGE
repeat steps "c" & "d" with the
REPLACE/ADD
RECORD/ADD
next Monitor Adaptor or follow
programming directions for
RESET
another component that you
would like to program.
Press the "Reset" button to
CALL REVIEW
exit programming mode.
VOLUME
LANGUAGE
REPLACE/ADD
RECORD/ADD
RESET
Wait for the Central Monitor to
CALL REVIEW
make a melodic tone. Then, repeat
VOLUME
LANGUAGE
step "c" with the next Exit Alarm
REPLACE/ADD
RECORD/ADD
or follow programming directions
for another component that you
RESET
would like to program.
Press the "Reset" button to
CALL REVIEW
exit programming mode.
VOLUME
LANGUAGE
REPLACE/ADD
RECORD/ADD
RESET

Advertisement