Specifications - 3M Ranger 247 Operator's Manual

Irrigation fluid warming system
Hide thumbs Also See for Ranger 247:
Table of Contents

Advertisement

Check the 3M
TM
Ranger
TM
system website to ensure you have the most recent version of this document.
English

Specifications

Physical Characteristics
warming unit
4.5 in. (11 cm) high x 7.5 in. (19 cm) wide x 10 in. (25 cm) long;
wt.: 7 lb. 7 oz. (3.4 kg)
certifications
IEC/EN 60601-1; CAN/CSA-C22.2, No.601.1
classification
MEDICAL — GENERAL MEDICAL EQUIPMENT AS TO ELECTRICAL SHOCK, FIRE AND
MECHANICAL HAZARDS ONLY IN ACCORDANCE WITH UL 60601-1; CAN/CSA-C22�2,
No�601�1; ANSI/AAMI ES60601-1:2005 CSA-C22�2 No� 60601-1:08; Control No�4HZ8
Classified under IEC 60601-1 Guidelines (and other national versions of the Guidelines) as Class I, Type B,
Ordinary equipment, Continuous operation. Classified by Underwriters Laboratories Inc. with respect to electric
shock, fire and mechanical hazards only, in accordance with IEC/EN 60601-1 and in accordance with Canadian/
CSA C22.2, No. 601.1. Classified under the Medical Device Directive as a Class IIb device.
Electrical characteristics
input voltage
100-120 or 220-240 VAC
operating frequency
100-120 VAC, 50/60 Hz
220-240 VAC, 50/60 Hz
maximum heating power
900 W
fuse
2 x T10A-H (250V) for 100-120 VAC
2 x T6.3A-H (250V) for 220-240 VAC
fuse type
Time delay, high breaking
Leakage Current
Meets leakage current requirements in accordance with UL/IEC 60601-1.
Environmental Conditions
operating temperature range
15° to 40°C (59° to 104°F)
storage temperature range
-20° to 45°C (-4° to 113°F)
operating humidity
10 to 85% RH, noncondensing
atmospheric pressure range
50 kPa to 106 kPa
14
3M
Ranger
Irrigation Fluid Warming System
TM
TM
Temperature characteristics
set point temperature
41°C
over-temperature alert
48°C
under-temperature alert
33°C
over-temperature cutoff
50°C
www.rangerfluidwarming.com reorder #202456B