VisionAire 2 & 3

Advertisement

Table of Contents
loading

Summary of Contents for CHART VisionAire 2

  • Page 1 User Manual VisionAire 2 & 3...
  • Page 2: User Controls & System Status Indicators

    User Controls & System Status Indicators Sym- Definition Symbol Definition Read user manual before operation. ON (power switch on) See user manual for instructions. No Smoking Icon: Do not smoke near OFF (power switch off) unit. No serviceable parts inside. Do not open Warnings and Cautions cover.
  • Page 3: Table Of Contents

    Español: Una versión multilingüe del manual está disponible a través de su proveedor de equipo. Français: Une version multilingue du manuel est disponible par l’intermédiaire de votre fournisseur de matériel. Deutsche: Eine mehrsprachige Version des Handbuchs ist in Ihrer Geräte-Anbieter VisionAire 2 & 3 User Manual — 3...
  • Page 4: Indications For Use

    Important! Safety Guidelines WARNING: NO MODIFICATION OF THIS EQUIP- MENT IS PERMITTED Safety Instructions are defined as fol- lows: WARNING: SMOKING WHILE USING OXYGEN IS WARNING: IMPORTANT SAFETY INFORMATION THE NUMBER ONE CAUSE OF FIRE, INJURY, AND FOR HAZARDS THAT MIGHT CAUSE SERIOUS DEATH.
  • Page 5 COULD ADVERSELY AFFECT THE PERFORMANCE OF THE DEVICE. TOO MANY PLUGS INTO ONE OUTLET CAN RESULT IN AN OVERLOAD TO THE ELECTRICAL PANEL CAUSING THE BREAKER/FUSE TO ACTIVATE OR FIRE IF THE BREAKER OR FUSE FAILS TO OPERATE. VisionAire 2 & 3 User Manual — 5...
  • Page 6 AVERTISSEMENT: CET APPAREIL PRODUIT CAUTION: Do not position the unit so that it DE L’OXYGÈNE À CONCENTRATION ÉLEVÉE, is difficult to access the power cord. FAVORISANT UNE COMBUSTION RAPIDE. NE CAUTION: The concentrator should be located PAS PERMETTRE DE FUMER OU DES FLAMMES as to avoid smoke, pollutants or fumes.
  • Page 7 Note: AirSep offers the OxySafe as an optional accessory. This total length of 25 ft. (7.6 m) max. is intended to be used in conjunction with the Visionaire 2/3 concentrator. For customers in regions requiring compliance Note: Ensure the cannula is fully inserted and secure. You...
  • Page 8: Introduction

    Note: There is never a danger of depleting the oxygen in a room when you use your VisionAire 2 or VisionAire 3 unit. 8 — MN165-1 M...
  • Page 9: Concentrator Components

    Concentrator Components Review the figures below and become familiar with the important parts of your VisionAire 2 or VisionAire 3 Oxygen Concentrator. Front of Concentrator Top Handle Hour Meter Oxygen Outlet Side Handle(s) Circuit Breaker Reset Button Flowmeter Adjustment Knob...
  • Page 10 Humidifier Bottle (optional) Humidifier Bottle Humidifier Bottle Fitting Recess Humidifier Bottle Humidifier Bottle Tubing Back of Concentrator Oxygen Outlet Humidifier Bottle Oxygen Tubing/Cannula Power Cord Humidifier Bottle (optional): Humidifies the oxygen Power Cord: Allows connection of the unit to an electrical outlet.
  • Page 11 Relief Cord Wrap Velcro Dust Filter Polyester Humidifier Strap Yarn, Rubber Humidifier Top – Acrylonitrile Butadiene Bottle Styrene (ABS) Bottle – Polypropylene (PP) Diffuser -Polyvinyl Chloride (PVC) Cannula Polyvinyl Chloride (PVC) (Tubing) VisionAire 2 & 3 User Manual — 11...
  • Page 12: Operating Instructions

    Operating Instructions Starting the Concentrator Review the following information before you operate your oxygen concentrator. WARNING: “NO SMOKING – OXYGEN IN USE” SIGNS MUST BE PROMINENTLY DISPLAYED Note: The concentrator releases warm air out the bottom IN THE HOME, OR WHERE OXYGEN IS IN USE. of the unit which can permanently discolor temperature sensi- PATIENTS AND THEIR CAREGIVERS MUST BE tive flooring surfaces such as vinyl.
  • Page 13: Care And Cleaning

    Use a mild household cleaner applied with a damp cloth or sponge to clean the exterior of the concentrator, and then wipe all surfaces dry. Do not allow any liquids to enter the concentrator. VisionAire 2 & 3 User Manual — 13...
  • Page 14: Troubleshooting

    If your concentrator fails to operate properly, consult your Equipment Provider, and refer to the troubleshooting chart on the following pages for probable causes and solutions. Note: Do not attempt any maintenance other than the possible solutions listed within the manual Note: If the unit has not been used for an extended time period, it needs to operate for several minutes before power failure alarm can become activated.
  • Page 15: Product Specifications

    0-7 kPa (0-1 psig back pressure). Maximum outlet pressure : 55 kPa. ( 7.98 psig) (3) Operating outside of these operational specifications can limit the concentrators ability to meet Oxygen Concentration specification at higher liter flow rates. VisionAire 2 & 3 User Manual — 15...
  • Page 16 Medical equipment needs special precautions regarding EMC and need to be installed and put into service according to the EMC information provided in this section. Guidance and Manufacturer’s Declaration—Electromagnetic Immunity The unit is intended for use in the electromagnetic environment specified below. The customer or the user of the unit should assure that it is used in such an environment.
  • Page 17 VisionAire. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m. VisionAire 2 & 3 User Manual — 17...
  • Page 18 Recommended separation distances between portable and mobile RF communications equipment and the unit The unit is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the unit can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the unit as recommended below, according to the maximum output power of the communications equipment.
  • Page 19: Disposal

    Please refer to your Equipment Provider and the VisionAire Service Manual. Degree of safety of application in the presence of flammable gases: Equipment not suitable for such application. Mode of operation: Continuous duty. VisionAire 2 & 3 User Manual — 19...
  • Page 20 Worldwide: www.chartindustries.com/Respiratory-Healthcare As of the date of this publication, VisionAire, AirSep and their related logos Copyright © 2016 Chart Industries. CAIRE Inc. reserves (as applicable) are trademarks registered by CAIRE Inc. the right to discontinue its products, or change the prices, Lysol ®...