Brief instruction V 1.0E
Operating instructions
The power switch is positioned on the rear of the CASY
Model DT tower.
The top right-hand area of the control panel accommodates
the keys MENU, CANCEL, ENTER, NEXT,
enter the measuring and output parameters.
,
Display
The bottom right-hand area accommodates the keys START,
CLEAN, PRINT. used to control the measuring stand and
output the results to the printer or external computer.
used to
On the left are the keys L-CURSOR,
R-CURSOR used to position the evaluation cursors.
Defi nition of key functions:
MENU:
Starts the menu control.
CANCEL:
Cancels a measurement or entry.
ENTER:
Confi rms an entry.
NEXT:
Continue to next number.
:
Increases number/change selection.
:
Increases number/change selection.
START:
Starts a measurement.
CLEAN:
Starts one or several cleaning cycles.
PRINT:
Output to printer and/or serial interface.
L-CURSOR:
Positions left cursor.
:
Shifts cursor to the right.
:
Shifts cursor to the left.
R-CURSOR: Positions right cursor.
,
,

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the CASY DT and is the answer not in the manual?

Questions and answers

Summary of Contents for Roche CASY DT

  • Page 1 CASY Model DT Brief instruction V 1.0E Operating instructions The power switch is positioned on the rear of the CASY The bottom right-hand area accommodates the keys START, Model DT tower. CLEAN, PRINT. used to control the measuring stand and output the results to the printer or external computer.
  • Page 2 Preparing for a measurement Switch on your CASY Model DT (The green power switch is fi tted to the rear panel of CASY Model DT) Replenish the storage container with clean CASY ton. Empty the waste container. After the system successfully performed the self test, press Main Menu DT-2AB-9999 15.04.2008 10:15 the Menu button to open the Main Menu.
  • Page 3 Manual input of the parameters Edit Setup opens the setup screen. When a setup screen is SETUP NUMBER: 00 NAME: Default opened, the input marker is located initially on the Dilution Capillary: 150 μm input fi eld. Sample Volume: 1 x 200 μl The current input position is indicated by a fl...
  • Page 4 3.684e+05 Before switching off the CASY Model DT monitor again the background (see 5.). For life science research only. Not for use in diagnostic procedures. CASY and 454 are trademarks of Roche. Published by Roche Diagnostics GmbH Roche Applied Science...

Table of Contents