Eaton Crouse-Hinds Z130 Instruction Manual

Eaton Crouse-Hinds Z130 Instruction Manual

Mtl zirconia oxygen analyser

Advertisement

Quick Links

Instruction manual
MTL gas analysers & systems
Z130
MTL zirconia oxygen analyser
February 2017
INM MTL 130-0200 Rev 4

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Crouse-Hinds Z130 and is the answer not in the manual?

Questions and answers

Summary of Contents for Eaton Crouse-Hinds Z130

  • Page 1 Instruction manual February 2017 MTL gas analysers & systems INM MTL 130-0200 Rev 4 Z130 MTL zirconia oxygen analyser...
  • Page 2 DECLARATION OF CONFORMITY A printed version of the Declaration of Conformity has been provided separately within the original shipment of goods. However, you can find a copy of the latest version at - http://www.mtl-inst.com/certificates INM MTL 130-0200 Rev 4...
  • Page 3: Table Of Contents

    CONTENTS DECLARATION OF CONFORMITY ............ii INTRODUCTION .
  • Page 4 THIS PAGE IS LEFT INTENTIONALLY BLANK INM MTL 130-0200 Rev 4...
  • Page 5: Introduction

    INTRODUCTION General Description The Z130 is a microprocessor controlled oxygen analyser based on a Zirconia oxygen sensor. It provides a range of features and a performance without parallel for an analyser of this type and cost. Standard features include auto-ranging over a span of 100% to 0.01ppm, user programmable alarm levels, hysteresis and analogue output.
  • Page 6: Specification

    SPECIFICATION Display • Multi digit LCD - character height 12.7mm Display ranges • Display range 0.01ppm to 100%, auto ranging Display Resolution • From 100% to 103% • From 10.0% to 99.0% 0.1% • From 1.00% to 9.99% 0.01% • From 0.100 to 0.999% 0.001% •...
  • Page 7: Analogue Output - Isolated

    2.12 Analogue output - isolated • 0 to 5 volts – minimum load 10k ohms or • 4 to 20mA – maximum load 500 ohms NOTE Programmable for full scale values of between 1ppm and 100% oxygen and zero scale values of between 0ppm and 90% however The full scale value must always be higher than zero scale value 2.13 Alarm outputs...
  • Page 8: Installation

    INSTALLATION Unpacking and visual checking Take all normal precautions when opening packages. In particular, avoid the use of long bladed cutters. Check for any sign of damage. Check that all pipe connections have captive seal nuts - search packing if any are missing. Calibration overview The instrument should be installed where free air ventilation around the whole case is provided.
  • Page 9: General Advice Sample

    3.3.2 General advice Sample The instrument is supplied with one of several sample system types specified by the user on order. A label on the instrument shows the pressure limits and other data for the particular one fitted. This should be taken into account before applying a sample. WARNING The case should not be exposed to water jets or drips.
  • Page 10: Electrical Connections

    Electrical connections 3.4.1 Measuring low levels of oxygen Power connections should be made to the instrument using the IEC lead supplied. 3.4.2 RS232 Connection NOTE See also section 8: Communication Protocol DCE (instrument) DTE (PC) signal description Signal gnd Data received by instrument Data transmitted from instrument 3.4.3 Alarm and Analogue Output connections...
  • Page 11: Switching On

    Switching on When the analyser is fully connected it may be switched on. The cell heater will begin to warm up. During this time the display will flash “HE xxxxx and the concentration display will make high and low excursions. Once the correct temperature is reached, the “HR” will end and the display will stabilise.
  • Page 12 Figure 3 - 130 series menu INM MTL 130-0200 Rev 4...
  • Page 13: Calibration

    CALIBRATION CAUTION Various procedures associated with calibration maintenance affect the outputs of the instrument. Any of these outputs that are being used for control (or the associated control loop) should be disabled before commencing. Calibration overview The Z130 is a very stable analyser with a drift rate of better than 2% of reading per month. How often the calibration is checked or verified depends on the quality regime being operated at the installation site.
  • Page 14: Calibration Procedure

    only ppb levels of combustibles. Calibration procedure NOTE: see Figure 3 All calibration screens operate on the same principle :- • The display shows the present reading. This may be changed by pressing “Edit” and then using the “ ” or “ ” arrows to adjust the value. •...
  • Page 15: Maintenance

    MAINTENANCE Calibration is the only routine operation that is necessary on a regular basis. The Z130 is a very stable analyser with a drift rate of better than 2% of reading per month. How often the calibration is checked or verified depends on the quality regime being operated at the installation site.
  • Page 16: Technical Description Of Sensor

    TECHNICAL DESCRIPTION OF SENSOR The zirconia oxygen sensor (see sketch below) is an impervious tube-shaped zirconia (zirconium oxide) element with a closed end coated externally and internally with porous metal electrodes, typically platinum. At high temperatures, typically above 400°C, the zirconia becomes an oxygen ion conductor, which results in a voltage being generated between the electrodes dependent upon the differences between the partial pressures of the oxygen in the sample and the oxygen in a reference gas (generally air).
  • Page 17: Communication Protocol

    COMMUNICATION PROTOCOL Communication parameters are 9600 Baud, 8 bits, no parity, 1 stop bit and no handshaking. The instrument is DTE. The general form of the protocol is: The protocol is a single command and response protocol. 2. A command consists of an “address” and an “action” . A received valid address is always responded to even if the action is invalid.
  • Page 18: Table 1 - Common Groups

    Table 1 - Common groups Send Function Reply (verbose) Reply (terse) Limits (where applicable) Notes C9 Load def=0 C9=0 0 or 1 Load defaults C8 Sens 2 os=0.00 C8 =0.00 Cell offset C7 Sens 2 K=1 C7 =45.0 Cell slope C6 Sens 2 H cal =100% C6 =0 AxCy...
  • Page 19: Table 1 - Common Groups

    Table 1 - Common groups (continued) Send Function Reply (verbose) Reply (terse) Limits (where applicable) Notes P9 Terse=0 P9=1 0 or 1 1 to set instrument into terse mode – 0 to set verbose P8 A2 Mode=1 P8 =1 0 to 3 0, 1, 2, 3 in terse - Off, High, Low, status in verbose P7 A2 Hyst=1.0%...
  • Page 20: Table 2 - Error Messages

    Table 2 - Error messages Error messages take the form ‘? xx’ where ‘xx’ is a numeric code as explained below. Code Error Description Buffer overflow More than 30 characters were received without message terminator (<CR><LF>). Any subsequent characters will begin a new message. Timeout 10 seconds has elapsed since the last character was received without message terminator.
  • Page 21 THIS PAGE IS LEFT INTENTIONALLY BLANK INM MTL 130-0200 Rev 4...
  • Page 22 THIS PAGE IS LEFT INTENTIONALLY BLANK INM MTL 130-0200 Rev 4...
  • Page 23 THIS PAGE IS LEFT INTENTIONALLY BLANK INM MTL 130-0200 Rev 4...
  • Page 24 MTL Instruments GmbH, Heinrich-Hertz-Str. 12, 50170 Kerpen, Germany UNITED ARAB EMIRATES Cooper Industries/Eaton Corporation Tel: +49 (0)22 73 98 12 - 0 Fax: +49 (0)22 73 98 12 - 2 00 Office 205/206, 2nd Floor SJ Towers, off. Old Airport Road, E-mail: csckerpen@eaton.com...