XPREZZON™ and qube™
91390, 91393, 94267
070-2112-01 Rev. F |
July 2015
www.spacelabshealthcare.com
O P E R A T I O N S
M A N U A L

Advertisement

Table of Contents
loading

Summary of Contents for Spacelabs XPREZZON

  • Page 1 XPREZZON™ and qube™ 91390, 91393, 94267 070-2112-01 Rev. F | July 2015 www.spacelabshealthcare.com O P E R A T I O N S M A N U A L...
  • Page 2 Printed in U.S.A. Specifications and price change privileges are reserved. Spacelabs Healthcare considers itself responsible for the effects on safety, reliability and performance of the equipment only • assembly operations, re-adjustments, modifications or repairs are carried out by persons authorized by Spacelabs Healthcare, and •...
  • Page 3: Table Of Contents

    To dock the qube monitor ................. 3-10 To undock the qube monitor ................3-11 Display Options Displays Overview......................4-1 XPREZZON Display options..................4-1 qube™ Secondary Display (94267-L15) ..............4-3 Display Adjustments .....................4-4 User Control Buttons ..................... 4-4 Onscreen Display (OSD) Menu ................4-5 To power ON the display..................
  • Page 4 X P R E Z Z O N A N D Q U B E P E R A T I O N S A N U A L Admit, Discharge, and Edit..................5-4 To admit a patient ....................5-5 To admit a patient (using barcode scanner) ..........
  • Page 5 X P R E Z Z O N A N D Q U B E P E R A T I O N S A N U A L To use Full Bed Review (FBR)................6-4 To use Full Bed Review (FBR) when a remote parameter is not shown..................6-4 To turn off Full Bed Review (FBR)..............
  • Page 6 X P R E Z Z O N A N D Q U B E P E R A T I O N S A N U A L To stop a continuous recording................. 8-5 To print Vital Signs....................8-5 Group Recordings ....................
  • Page 7 To select a large numerics format ..............12-15 To turn off the large numerics................. 12-16 To configure the large numerics format............12-16 Remote Config (XPREZZON only) ............... 12-17 To select a subset of parameters ..............12-17 To change the position of a parameter ............12-18 To change the color of a parameter waveform........
  • Page 8 X P R E Z Z O N A N D Q U B E P E R A T I O N S A N U A L 13 Cleaning, Disinfecting, and Sterilization Cleaning Products Not Recommended for Use ..........13-1 Touchscreen Cleaning ....................13-1 Monitors, Modules, Cables, Printers, and Docking Station......13-2 Cleaning/Disinfecting ...................
  • Page 9: Introduction

    Spacelabs Healthcare patient monitoring systems demonstrate an ongoing commitment to innovation that supports decision-making and enhances patient care. Spacelabs monitors are flexible enough to be set for any level of acuity, and they can be precisely and easily adjusted using touchscreen technology.
  • Page 10: Conventions Used In This Manual

    A N U A L Conventions used in this manual Spacelabs uses specific types of indications to draw your attention to how to use the monitor. • Reference links are in blue. Place the mouse over the blue text and click the text, and the topic will be shown.
  • Page 11 Spacelabs Healthcare patient monitors are intended to alert the user to alarm conditions that are reported by Spacelabs Healthcare parameter modules and/or other physiologic monitors via Flexport interfaces. These devices determine a) when an alarm condition is violated;...
  • Page 12: Flexport System Interface

    Spacelabs Healthcare monitors. Flexport interfaces provide current numeric data, alarm information, and selected waveforms. All of the data communicated to the Spacelabs Healthcare monitor is available at network locations, providing alarms and centralized displays. Contact your Spacelabs Healthcare sales representative for additional information.
  • Page 13: Parameter Keys

    Manager System Administration Guide for detailed information on patient preparation for each module and parameter. Warnings: • Use only Spacelabs Healthcare parts and accessories with your Spacelabs Healthcare products. Other parts and accessories may degrade performance or damage the components. Refer to the Spacelabs Healthcare Supplies and Accessories Catalog for the part numbers and descriptions of additional parts and accessories.
  • Page 14 X P R E Z Z O N A N D Q U B E N T R O D U C T I O N P E R A T I O N S A N U A L 1 - 6 www.spacelabshealthcare.com...
  • Page 15: Xprezzon Monitor

    XPREZZON allows for a number of different mounting options and accommodates two single-high parameter modules. Included with each monitor is a DC power supply that provides power for the monitor and up to three Flexport interfaces. The monitor supports single-display configurations and dual-display configurations, depending on options.
  • Page 16: Touchscreen Display

    P E R A T I O N S A N U A L • The XPREZZON should be plugged into AC power when not in use to charge the internal NiMH battery used to retain alarm settings in modules if power loss occurs.
  • Page 17: Basic Monitor Components

    P E R A T I O N S A N U A L Basic Monitor Components The connectors for the monitor include the following: Figure 2-1 Front and back (XPREZZON) Table 2-1 Front (XPREZZON) Description Power On/Off button- Press and hold the Power button for five seconds to turn the monitor off.
  • Page 18 A N D Q U B E O N I T O R P E R A T I O N S A N U A L Table 2-2 Back Panel (XPREZZON) Description Symbol on panel Battery backup cover Equipotential Terminal used for grounding...
  • Page 19: Optional Integrated Recorder

    (Independent of 94267 display) Power Output, 12 volt Optional Integrated Recorder Using the 12-volt power output on the back of the monitor, XPREZZON has an optional integrated recorder that requires no additional power. For more information on the printing functions, refer to Recording/Printing.
  • Page 20 X P R E Z Z O N X P R E Z Z O N M A N D Q U B E O N I T O R P E R A T I O N S A N U A L 2 - 6 www.spacelabshealthcare.com...
  • Page 21: Qube™ Monitor

    • Wireless. Refer to Signal Strength Indicator on page 3-7 for details. Refer to the ™ ™ XPREZZON and qube System Administration Manual to setup the wireless connection. • Battery operation. Refer to Battery Management* on page 3-8 for details.
  • Page 22: Basic Components

    P E R A T I O N S A N U A L Note: This chapter should be used with the following chapters as well as the following manuals included on the Spacelabs Healthcare Bedside, Central, and Telemetry Systems CD-ROM. ™ ™...
  • Page 23 X P R E Z Z O N ™ M A N D Q U B E Q U B E O N I T O R P E R A T I O N S A N U A L Description Symbol Power On/Off button- Press and hold the...
  • Page 24 X P R E Z Z O N ™ M A N D Q U B E Q U B E O N I T O R P E R A T I O N S A N U A L Figure 3-3 Back Panel (qube) Description Symbol on panel...
  • Page 25 X P R E Z Z O N ™ M A N D Q U B E Q U B E O N I T O R P E R A T I O N S A N U A L Description Symbol on panel Ethernet 10/100 Base T Connection...
  • Page 26: Feature Options

    X P R E Z Z O N ™ M A N D Q U B E Q U B E O N I T O R P E R A T I O N S A N U A L Feature Options Table 3-1 qube Feature Options Feature...
  • Page 27: To Enable Or Disable The Display Alarm Lights

    Environmental factors can contribute to waveform gaps over the wireless network. If this issue continues, consult with a biomed technician or a Spacelabs Healthcare field service representative. The wireless signal strength shows between the Admit Patient button and alarm buttons. The signal strength shows a green antenna with vertical bars on a black background.
  • Page 28: Battery Management

    X P R E Z Z O N ™ M A N D Q U B E Q U B E O N I T O R P E R A T I O N S A N U A L When connectivity is lost, the message NETWORK SIGNAL LOST shows.
  • Page 29: Battery Power Level

    • the monitor is not plugged into an AC power source and there are no batteries installed • an unrecoverable error occurs To optimize the notification, configure the power loss alarm delay according to the ™ ™ XPREZZON and qube System Administration Manual. Table 3-3 Battery power level messages and tones **Approximate time Recorder...
  • Page 30: To Dock The Qube Monitor

    X P R E Z Z O N ™ M A N D Q U B E Q U B E O N I T O R P E R A T I O N S A N U A L The Docking Station lets the qube monitor change from a bedside monitor to a transport monitor quickly and simply.
  • Page 31: To Undock The Qube Monitor

    X P R E Z Z O N ™ M A N D Q U B E Q U B E O N I T O R P E R A T I O N S A N U A L Figure 3-5 Docking the qube monitor To undock the qube monitor Hold the qube monitor by its handle and press the release button located in...
  • Page 32 X P R E Z Z O N ™ M A N D Q U B E Q U B E O N I T O R P E R A T I O N S A N U A L 3 - 1 2 www.spacelabshealthcare.com...
  • Page 33: Display Options

    Displays Overview ™ The XPREZZON and the qube have display options described in this chapter. The XPREZZON has three display options. The display features are listed in the table on page 4-2. ™ The qube is a portable monitor and has an option to have a secondary display attached.
  • Page 34 X P R E Z Z O N A N D Q U B E I S P L A Y P T I O N S P E R A T I O N S A N U A L Table 4-1 Display Features (Continued) 94267-L17 94267-L19...
  • Page 35: Qube™ Secondary Display (94267-L15)

    X P R E Z Z O N A N D Q U B E I S P L A Y P T I O N S P E R A T I O N S A N U A L qube™...
  • Page 36: Display Adjustments

    If power fails during a power outage or prolonged generator test, or an unrecoverable error occurs, the monitor makes a high-pitched sound. To optimize ™ the notification, configure the power loss alarm delay according to the XPREZZON ™ and qube System Administration Manual.
  • Page 37: Onscreen Display (Osd) Menu

    X P R E Z Z O N A N D Q U B E I S P L A Y P T I O N S P E R A T I O N S A N U A L Menu + Up at the same time: Press Menu and Up buttons at the same time and hold for five (5) seconds to enable or disable the Key Lock.
  • Page 38: Display Alarm Light

    X P R E Z Z O N A N D Q U B E I S P L A Y P T I O N S P E R A T I O N S A N U A L Press the Up and Down buttons in the MISC menu to highlight the RESET option.
  • Page 39: Connectors And Mounting

    P E R A T I O N S A N U A L Connectors and Mounting The setup and maintenance information is contained in the XPREZZON (P/N 070- 2155-xx) or qube (P/N 070-2451-xx) service manuals on the Service CD (P/N 084- 0700-xx) Equipment connected to the analog or digital interfaces must comply with the respective IEC standards (e.g.
  • Page 40: Mounting

    X P R E Z Z O N A N D Q U B E I S P L A Y P T I O N S P E R A T I O N S A N U A L Figure 4-4 Display connections for 94267-L15, 94267-L17, and 94267-L19 Touchscreen serial port DVI video in, analog or digital connector...
  • Page 41 X P R E Z Z O N A N D Q U B E I S P L A Y P T I O N S P E R A T I O N S A N U A L Figure 4-5 Mounting for 94267-19 and 94267-L19 Figure 4-6 Mounting for the 94267-L17 display www.spacelabshealthcare.com...
  • Page 42 X P R E Z Z O N A N D Q U B E I S P L A Y P T I O N S P E R A T I O N S A N U A L Figure 4-7 Mounting for the 94267-L15 Never open the displays.
  • Page 43: Basic Operations

    X P R E Z Z O N A N D Q U B E P E R A T I O N S A N U A L Basic Operations The list of basic functions below are done from the monitor before, during, or after monitoring patient vital signs: •...
  • Page 44: Home Screen And Taskbar Icons

    The bedside monitor can (optionally) show four, six or eight parameter zones. Refer to the ™ ™ XPREZZON and qube System Administration Manual for more details on configuring the number of parameter zones. Refer to the Clinical Parameters Operations Manual for details on using individual parameter modules, parameter keys or menu keys.
  • Page 45: Main Taskbar

    X P R E Z Z O N A N D Q U B E A S I C P E R A T I O N S P E R A T I O N S A N U A L Main taskbar.
  • Page 46: Large Numeric Displays

    12-15. Appearance of the parameters in the large numerics are user configurable. Below are the display options. Note: qube has two large numeric display formats and XPREZZON has three formats. Figure 5-2 Large Numeric display formats Depending on the format selected, a fewer number of waveforms are shown. The following steps let you see the waveforms of the parameters that are shown in large numerics.
  • Page 47: To Admit A Patient

    X P R E Z Z O N A N D Q U B E A S I C P E R A T I O N S P E R A T I O N S A N U A L •...
  • Page 48: To Edit Patient Data

    X P R E Z Z O N A N D Q U B E A S I C P E R A T I O N S P E R A T I O N S A N U A L •...
  • Page 49: To View And Print The Patient's Vital Signs

    X P R E Z Z O N A N D Q U B E A S I C P E R A T I O N S P E R A T I O N S A N U A L •...
  • Page 50: Recording/Printing

    X P R E Z Z O N A N D Q U B E A S I C P E R A T I O N S P E R A T I O N S A N U A L Touch an area on the time bar.
  • Page 51: To Make A Recording Of All Currently Shown Patient Parameters

    X P R E Z Z O N A N D Q U B E A S I C P E R A T I O N S P E R A T I O N S A N U A L To make a recording of all currently shown patient parameters From the main taskbar, touch the Printer icon twice.
  • Page 52: To Print A Preconfigured Recording Set

    X P R E Z Z O N A N D Q U B E A S I C P E R A T I O N S P E R A T I O N S A N U A L To print a preconfigured recording set From the main taskbar, touch the Printer icon twice.
  • Page 53: Remote Access

    A N U A L Remote Access Use Remote Access to view or print vital signs of patients that are being monitored on other Spacelabs Healthcare monitors. For additional information on configuration of remote access, refer to Remote Access on page 6-1 or contact your system administrator.
  • Page 54: To Use Full Bed Review (Fbr)

    X P R E Z Z O N A N D Q U B E A S I C P E R A T I O N S P E R A T I O N S A N U A L To use Full Bed Review (FBR) From the main taskbar, touch the Remote icon.
  • Page 55: Standby Mode

    X P R E Z Z O N A N D Q U B E A S I C P E R A T I O N S P E R A T I O N S A N U A L Standby Mode If you need to temporarily stop patient monitoring, you can use the Standby Mode feature.
  • Page 56: Perioperative

    However, visual alarm indicators will continue to show upon END CASE, unless Alarms Off is selected. Remote Configurable Display (XPREZZON only) - A secondary display that is controlled from the monitor's Monitor Setup menu, using the SECONDARY CONFIG.
  • Page 57: Start Case/End Case

    X P R E Z Z O N A N D Q U B E A S I C P E R A T I O N S P E R A T I O N S A N U A L Start Case/End Case When enabled, the monitor shows a key labeled START CASE or END CASE along the bottom of the monitor's display (refer to...
  • Page 58: Calculations

    X P R E Z Z O N A N D Q U B E A S I C P E R A T I O N S P E R A T I O N S A N U A L Calculations The Calculator icon on the Trends toolbar lets you access the Drug Calculator window or Physiologic Calculations window.
  • Page 59: Data Shuttle

    X P R E Z Z O N A N D Q U B E A S I C P E R A T I O N S P E R A T I O N S A N U A L Data Shuttle Data Shuttle®...
  • Page 60: To Retrieve Patient Data At The Destination Monitor

    X P R E Z Z O N A N D Q U B E A S I C P E R A T I O N S P E R A T I O N S A N U A L To retrieve patient data at the destination monitor Insert the Command module into the destination monitor.
  • Page 61: Remote Access

    A N D Q U B E P E R A T I O N S A N U A L Remote Access Spacelabs Healthcare networking features support seamless data exchange across the medical enterprise, addressing the need for continuous information management. Overview The Remote Access feature lets you view data or receive alarm notifications from any bedside monitor, at any other bedside or central monitor.
  • Page 62: Network Basics

    X P R E Z Z O N A N D Q U B E E M O T E C C E S S P E R A T I O N S A N U A L Figure 6-1 Remote Access Window Network Basics Networking lets you communicate between beds in a single care unit or to review information from another care unit.
  • Page 63: Remote View

    X P R E Z Z O N A N D Q U B E E M O T E C C E S S P E R A T I O N S A N U A L Under Select Subnet, touch the subnet that you want to view remotely. The Select Bed options shows.
  • Page 64: Full Bed Review

    X P R E Z Z O N A N D Q U B E E M O T E C C E S S P E R A T I O N S A N U A L Touch the red X in the top right corner of the Remote window to close the window.
  • Page 65: To Turn Off Full Bed Review (Fbr)

    X P R E Z Z O N A N D Q U B E E M O T E C C E S S P E R A T I O N S A N U A L To turn off Full Bed Review (FBR) Touch Home.
  • Page 66: To Use Alarm Watch To View Remote Alarms On Other Beds On The Network

    X P R E Z Z O N A N D Q U B E E M O T E C C E S S P E R A T I O N S A N U A L To use Alarm Watch to view remote alarms on other beds on the network From the main taskbar, touch the Remote icon.
  • Page 67: To Access Remote Calculations Window

    X P R E Z Z O N A N D Q U B E E M O T E C C E S S P E R A T I O N S A N U A L You can view calculated hemodynamic and oxygenation data for any patient on the network.
  • Page 68 X P R E Z Z O N A N D Q U B E E M O T E C C E S S P E R A T I O N S A N U A L 6 - 8 www.spacelabshealthcare.com...
  • Page 69: Alarms

    Note: The bedside monitor is the primary monitoring and alarming device for Spacelabs-monitored parameters. When a parameter value on a bedside monitor exceeds an alarm limit, that monitor initiates an alarm. User notification of an alarm has five possible characteristics: visual, audible (tone), recording, alarm relay, and alarm watch.
  • Page 70 X P R E Z Z O N A N D Q U B E L A R M S P E R A T I O N S A N U A L Warnings: • Alarm conditions for which you want to be alerted must be set to ON or enabled at the location where you want to be alerted —...
  • Page 71: Alarm Limits

    X P R E Z Z O N A N D Q U B E L A R M S P E R A T I O N S A N U A L Alarm Limits Default alarm limit settings activate when the monitor is powered ON, a module is inserted, or a parameter is enabled.
  • Page 72: Bedside Alarms

    X P R E Z Z O N A N D Q U B E L A R M S P E R A T I O N S A N U A L Bedside Alarms Priority Tone Type Tone Duration Visual Flash Rates High...
  • Page 73: Alarm Icons

    X P R E Z Z O N A N D Q U B E L A R M S P E R A T I O N S A N U A L Alarm Icons The Alarms Pause, Audio, and Alarm Acknowledge icons show on the bottom right corner of the home screen (refer to Figure 7-1 on page 7-5).
  • Page 74: Audio Pause

    X P R E Z Z O N A N D Q U B E L A R M S P E R A T I O N S A N U A L Wireless, Network, and Battery icons. Refer to page 3-7 and page 3-7. Note: Some of the options may not be available.
  • Page 75: Alarm Pause

    X P R E Z Z O N A N D Q U B E L A R M S P E R A T I O N S A N U A L Alarm Pause Use this control to pause all alarms for a specified period of time. The alarms paused period is determined by your system administrator.
  • Page 76: To Open The Alarm History Window

    X P R E Z Z O N A N D Q U B E L A R M S P E R A T I O N S A N U A L Figure 7-2 Alarm History window To open the Alarm History window Touch the Alarm History bar on the home screen.
  • Page 77: To View A Waveform In The Alarm History Window (For Ecg And Invasive Pressure Alarms Only)

    X P R E Z Z O N A N D Q U B E L A R M S P E R A T I O N S A N U A L To view a waveform in the Alarm History window (for ECG and invasive pressure alarms only) Touch an alarm event to show a compressed 30 second waveform with 3 second markers.
  • Page 78: To Acknowledge An Alarm

    X P R E Z Z O N A N D Q U B E L A R M S P E R A T I O N S A N U A L To acknowledge an alarm Touch the Alarm Acknowledge icon, shown on the right. The Audio Off icon shows in the parameter zone (refer to Figure 7-3),...
  • Page 79: Alarm Audio

    • Check alarm volumes periodically! Turning the volume off or too low may defeat the audible alarm function. A minimum volume selection is described ™ ™ in the XPREZZON and qube System Administration Manual. • The two lowest volume settings produce less than 45 dBA at one meter from the internal speaker.
  • Page 80: To Adjust The Volume Levels For High, Medium, Low Priority Alarms, Remote Alarm, Alarm Watch, And Key Tones

    X P R E Z Z O N A N D Q U B E L A R M S P E R A T I O N S A N U A L To adjust the volume levels for High, Medium, Low Priority Alarms, Remote Alarm, Alarm Watch, and Key tones Touch Monitor Setup.
  • Page 81: Alarm Watch

    The message AW OFF also shows on the central monitor for each bed not selected for Alarm Watch. For more details on Alarm Watch Setup, ™ ™ refer to the XPREZZON and qube System Administration Manual. Alarm Watch notifications distribute across wired or wireless ethernet networks.
  • Page 82: Alarm Recording

    Configuration Manager (MCM). To enable or disable alarm recordings for those parameters that are not available in the MCM, you will need to access the Clinical ™ Setup tab. For more information about Clinical Setup, refer to the XPREZZON ™ qube System Administration Manual.
  • Page 83: Recording/Printing

    • Print non-waveform data such as tabular trends, hemodynamic and oxygenation calculations, and alarm limits. Printing Configurations Each Spacelabs Healthcare, network-connected, patient monitor is capable of sending recordings to either of two network printers. These can be configured in several ways.
  • Page 84: Configuration 2

    X P R E Z Z O N A N D Q U B E E C O R D I N G R I N T I N G P E R A T I O N S A N U A L Under high-load conditions, the following factors are taken into consideration when a print request occurs: •...
  • Page 85: Printing Priorities

    X P R E Z Z O N A N D Q U B E E C O R D I N G R I N T I N G P E R A T I O N S A N U A L Printing Priorities The following list defines printing priorities from the highest to lowest: •...
  • Page 86: Selecting Recording Duration

    Recordings requested from the monitor are either 12 or 20 seconds in Duration ™ ™ (seconds). To configure, refer to Printer on page 12-9 or the XPREZZON and qube System Administration Manual. Selecting Recording Destination The Recording Destination menu lets you direct: •...
  • Page 87: To Make A Recording Of All Currently Shown Patient Parameters

    X P R E Z Z O N A N D Q U B E E C O R D I N G R I N T I N G P E R A T I O N S A N U A L To make a recording of all currently shown patient parameters From the main taskbar, touch the Printer icon twice.
  • Page 88: Printer Controls

    X P R E Z Z O N A N D Q U B E E C O R D I N G R I N T I N G P E R A T I O N S A N U A L Printer Controls If a printer is installed, the Printer Controls button shows in the lower left section of the main window.
  • Page 89: Trends

    X P R E Z Z O N A N D Q U B E P E R A T I O N S A N U A L Trends Trends provide clinicians with a tool to quickly assess the patient’s current clinical status.
  • Page 90 X P R E Z Z O N A N D Q U B E R E N D S P E R A T I O N S A N U A L Measurement data can be grouped into categories. For example, a category named Vitals might consist of the measurements HR, RR, and NIBP (Systolic/Diastolic/ Mean).
  • Page 91: Trends Toolbar

    X P R E Z Z O N A N D Q U B E R E N D S P E R A T I O N S A N U A L Timeline—Used to access all of the data available for the current patient.
  • Page 92: To Open The Trends Window

    X P R E Z Z O N A N D Q U B E R E N D S P E R A T I O N S A N U A L To open the Trends window From the main taskbar, touch Trends. The Trends window shows.
  • Page 93: To Set The Interval For Tabular Trends In The Current View

    X P R E Z Z O N A N D Q U B E R E N D S P E R A T I O N S A N U A L To set the interval for tabular trends in the current view From the Trends toolbar, touch Interval.
  • Page 94: Trends Setup

    X P R E Z Z O N A N D Q U B E R E N D S P E R A T I O N S A N U A L Trends Setup Several options are available for customizing trends displays. •...
  • Page 95: To Open The Trends Setup Window

    X P R E Z Z O N A N D Q U B E R E N D S P E R A T I O N S A N U A L Down arrow icon moves the selected parameter to a lower position on the list.
  • Page 96: Tabular Trends Setup

    X P R E Z Z O N A N D Q U B E R E N D S P E R A T I O N S A N U A L Tabular Trends Setup You use the Tabular tab of the Trends Setup window: •...
  • Page 97: To Add Measurements To A Category

    X P R E Z Z O N A N D Q U B E R E N D S P E R A T I O N S A N U A L To add measurements to a category From the Trends Setup window, touch the Tabular tab.
  • Page 98: To Rename A Graph

    X P R E Z Z O N A N D Q U B E R E N D S P E R A T I O N S A N U A L Note: Adjusting the setup in the Trends menu is temporary: it is just for the current patient.
  • Page 99: To Set The Scale For A Measurement

    X P R E Z Z O N A N D Q U B E R E N D S P E R A T I O N S A N U A L To set the scale for a measurement From the Trends Setup window, touch the tab for the graph that you want to adjust the properties for.
  • Page 100 X P R E Z Z O N A N D Q U B E R E N D S P E R A T I O N S A N U A L 9 - 1 2 www.spacelabshealthcare.com...
  • Page 101: Calculations

    X P R E Z Z O N A N D Q U B E P E R A T I O N S A N U A L Calculations Although showing waveforms and measurements can provide critical information to the clinician, adequate assessment and diagnosis sometimes requires a broader view.
  • Page 102: To Open The Calculations Window

    X P R E Z Z O N A N D Q U B E A L C U L A T I O N S P E R A T I O N S A N U A L •...
  • Page 103: To Select A Calculation Type To View Or Edit

    X P R E Z Z O N A N D Q U B E A L C U L A T I O N S P E R A T I O N S A N U A L To select a calculation type to view or edit From the toolbar at the top of the Calculations window, touch the key for the type of calculation you want to view: Hemodynamic (far left key) or...
  • Page 104: To Refresh Calculations Data

    X P R E Z Z O N A N D Q U B E A L C U L A T I O N S P E R A T I O N S A N U A L To refresh calculations data Touch the Update key.
  • Page 105: Input Values For Hemodynamic Calculations

    X P R E Z Z O N A N D Q U B E A L C U L A T I O N S P E R A T I O N S A N U A L Input Values for Hemodynamic Calculations The following table shows the input values that are available for hemodynamic calculations.
  • Page 106: Oxygenation Calculations

    X P R E Z Z O N A N D Q U B E A L C U L A T I O N S P E R A T I O N S A N U A L Variable (Label) Equations Units...
  • Page 107: Output Values For Oxygenation Calculations

    X P R E Z Z O N A N D Q U B E A L C U L A T I O N S P E R A T I O N S A N U A L Label Name Units...
  • Page 108: Rate And Dose Values

    X P R E Z Z O N A N D Q U B E A L C U L A T I O N S P E R A T I O N S A N U A L Figure 10-2 Drug Calculator window Input fields Keypad for entering new values...
  • Page 109: To Access The Drug Calculator From The Calculations Window

    X P R E Z Z O N A N D Q U B E A L C U L A T I O N S P E R A T I O N S A N U A L To access the Drug Calculator from the Calculations Window From the Calculations window’s taskbar, touch the Calculator icon.
  • Page 110 X P R E Z Z O N A N D Q U B E A L C U L A T I O N S P E R A T I O N S A N U A L 1 0 - 1 0 www.spacelabshealthcare.com...
  • Page 111: Dynamic Network Access (Dna)

    A N U A L Dynamic Network Access (DNA) Dynamic Network Access™ (DNA™) is a Spacelabs Healthcare proprietary version of the Citrix ICA Client that lets the clinical user run remote applications on the local monitor. The DNA feature is a purchasable option.
  • Page 112: To Open The Dna Application Window

    X P R E Z Z O N ( D N A ) A N D Q U B E Y N A M I C E T W O R K C C E S S P E R A T I O N S A N U A L Figure 11-1 Connection View and Program Neighborhood View windows The DNA Settings key is found in the Connection View tab, and helps the user enter...
  • Page 113: To Refresh The List Of Applications

    X P R E Z Z O N A N D Q U B E Y N A M I C E T W O R K C C E S S ( D N A ) P E R A T I O N S A N U A L Note: You cannot launch a second application from the Program Neighborhood View...
  • Page 114: To Reconnect To A Disconnected Session

    X P R E Z Z O N ( D N A ) A N D Q U B E Y N A M I C E T W O R K C C E S S P E R A T I O N S A N U A L To reconnect to a disconnected session From the Program Neighborhood View tab of the DNA application window,...
  • Page 115: Monitor Setup

    • Remote Config tab: Lets you adjust parameter colors and order for the remote display. Refer to Remote Config (XPREZZON only) on page 12-17. ™ ™ For information on Privileged Access button, refer to the XPREZZON and qube System Administration Manual. www.spacelabshealthcare.com 1 2 - 1...
  • Page 116: Alarm Audio

    X P R E Z Z O N A N D Q U B E O N I T O R E T U P P E R A T I O N S A N U A L Alarm Audio Tones are the sounds associated with an alarm condition as well as the sound made each time a key on the display is touched.
  • Page 117: To Enable And Adjust Volume Levels For High, Medium, Low Priority Alarms, Remote Alarm, Alarm Watch, And Key Tones

    X P R E Z Z O N A N D Q U B E O N I T O R E T U P P E R A T I O N S A N U A L To enable and adjust volume levels for High, Medium, Low Priority Alarms, Remote Alarm, Alarm Watch, and Key tones Touch Monitor Setup.
  • Page 118: Audio Off Reminder

    X P R E Z Z O N A N D Q U B E O N I T O R E T U P P E R A T I O N S A N U A L Audio Off Reminder By default, the control to turn off audio is not available to the clinician.
  • Page 119: Alarm Limit Review

    X P R E Z Z O N A N D Q U B E O N I T O R E T U P P E R A T I O N S A N U A L Alarm Limit Review The Alarm Limit Review tab is where you can view or print the alarm settings for all parameters.
  • Page 120: Local Config

    X P R E Z Z O N A N D Q U B E O N I T O R E T U P P E R A T I O N S A N U A L Local Config Use the settings on the Local Config tab to: •...
  • Page 121 X P R E Z Z O N A N D Q U B E O N I T O R E T U P P E R A T I O N S A N U A L The following table shows the default priority and color of parameter waveforms. Parameter Type Description Priority...
  • Page 122: To Change The Position Of A Parameter

    X P R E Z Z O N A N D Q U B E O N I T O R E T U P P E R A T I O N S A N U A L To change the position of a parameter Touch Monitor Setup.
  • Page 123: To Restore The Factory Defaults For A Configuration

    X P R E Z Z O N A N D Q U B E O N I T O R E T U P P E R A T I O N S A N U A L To restore the factory defaults for a configuration Note: The Factory Default button is not available if the current configuration is the same as the factory default configuration.
  • Page 124: To Set Print Duration

    X P R E Z Z O N A N D Q U B E O N I T O R E T U P P E R A T I O N S A N U A L Figure 12-6 Printer tab To set print duration Touch Monitor Setup.
  • Page 125: To Setup Print Vital Signs

    X P R E Z Z O N A N D Q U B E O N I T O R E T U P P E R A T I O N S A N U A L Figure 12-7 Print Vitals Signs Setup To setup Print Vital Signs Touch Monitor Setup.
  • Page 126: Setup

    • Enable or disable the patient data logger (PDL) output. Note: The PDL control shows only if the monitor is preconfigured with this option and ™ when the serial port is set to PDL. For more information refer to the XPREZZON ™ and qube System Administration Manual.
  • Page 127: To Enable Or Disable The Clock Display

    X P R E Z Z O N A N D Q U B E O N I T O R E T U P P E R A T I O N S A N U A L • Enable or disable minor graticules and adjust how they show. Minor graticules are small, dashed line segments that show between full-scale graticule lines when scaled pressures show.
  • Page 128: To Set Minimum Alarm Level That Activates The Alarm Light

    X P R E Z Z O N A N D Q U B E O N I T O R E T U P P E R A T I O N S A N U A L To set minimum alarm level that activates the alarm light Touch Monitor Setup.
  • Page 129: Numerics

    Numerics Use the Numerics tab to enable, disable, and change the large numerics format. The list of parameters may vary with the software version. XPREZZON has three large numeric formats, and qube has two formats. Figure 12-9 Numerics tab, qube and XPREZZON...
  • Page 130: To Turn Off The Large Numerics

    X P R E Z Z O N A N D Q U B E O N I T O R E T U P P E R A T I O N S A N U A L To turn off the large numerics Touch Monitor Setup.
  • Page 131: Remote Config (Xprezzon Only)

    E T U P P E R A T I O N S A N U A L Remote Config (XPREZZON only) This tab has similar setup to the Local Config tab but it is used for setting up a remote configurable display.
  • Page 132: To Change The Position Of A Parameter

    X P R E Z Z O N A N D Q U B E O N I T O R E T U P P E R A T I O N S A N U A L Touch Select. Touch all of the parameters that you want to open on the remote display.
  • Page 133: To Rename A Configuration

    X P R E Z Z O N A N D Q U B E O N I T O R E T U P P E R A T I O N S A N U A L To rename a configuration Touch the tab that corresponds to the configuration whose name you want to change.
  • Page 134: To Copy Local Parameter Settings To The Remote Monitor

    X P R E Z Z O N A N D Q U B E O N I T O R E T U P P E R A T I O N S A N U A L Under Clock, touch On or Off. To copy local parameter settings to the remote monitor Touch Monitor Setup.
  • Page 135: Cleaning, Disinfecting, And Sterilization

    X P R E Z Z O N A N D Q U B E P E R A T I O N S A N U A L Cleaning, Disinfecting, and Sterilization Cleaning Products Not Recommended for Use Warnings: •...
  • Page 136: Monitors, Modules, Cables, Printers, And Docking Station

    Sodium hypochlorite solution (1:10 dilution of household chlorine bleach in water) • Phenolic germicidal detergent (1% aqueous solution) • Glutaraldehyde (2.4%) (Cidex) • Isopropyl alcohol (70% solution) Caution: Questions and concerns about cleaning issues should be directed to Spacelabs Healthcare Technical Support. 1 3 - 2 www.spacelabshealthcare.com...
  • Page 137: To Clean The Exterior Of Displays, Monitors, Modules, And Cables

    • Over time, repeated use of a chlorine bleach solution may cause some colors to fade. • Tape adhesive can be removed with Spacelabs Healthcare adhesive tape remover pads (P/N 392196-001). • In normal use of the docking station and qube monitor, the docking station connectors are not exposed.
  • Page 138: Noninvasive Blood Pressure Cuffs

    X P R E Z Z O N A N D Q U B E L E A N I N G I S I N F E C T I N G P E R A T I O N S A N U A L A N D T E R I L I Z A T I O N...
  • Page 139: A Appendix A - Symbols

    A N U A L Appendix A — Symbols The following list of international and safety symbols describes all symbols used on Spacelabs Healthcare products. No one product contains every symbol. Note: Graphic elements of certain keys and symbols may vary between product lines.
  • Page 140 X P R E Z Z O N A — S A N D Q U B E P P E N D I X Y M B O L S P E R A T I O N S A N U A L TRENDS Key RECORD Key...
  • Page 141 X P R E Z Z O N A — S A N D Q U B E P P E N D I X Y M B O L S P E R A T I O N S A N U A L List of Rooms Printer...
  • Page 142 X P R E Z Z O N A — S A N D Q U B E P P E N D I X Y M B O L S P E R A T I O N S A N U A L ON/OFF Standby...
  • Page 143 X P R E Z Z O N A — S A N D Q U B E P P E N D I X Y M B O L S P E R A T I O N S A N U A L Alarm, General Alarm Reset...
  • Page 144 X P R E Z Z O N A — S A N D Q U B E P P E N D I X Y M B O L S P E R A T I O N S A N U A L Clock/Time Setting Key Slow Run...
  • Page 145 X P R E Z Z O N A — S A N D Q U B E P P E N D I X Y M B O L S P E R A T I O N S A N U A L Waveform/Parameter Keys Return to Prior Menu...
  • Page 146 X P R E Z Z O N A — S A N D Q U B E P P E N D I X Y M B O L S P E R A T I O N S A N U A L Video Output, Primary Video Output, Secondary...
  • Page 147 X P R E Z Z O N A — S A N D Q U B E P P E N D I X Y M B O L S P E R A T I O N S A N U A L Gas Sampling Port Gas Return Port...
  • Page 148 Both Direct and Alternating Current Functional Earth Ground Fuse Equipotentiality Terminal Direct Current Input Power. Use only Spacelabs Power Supply (P/N 119-0527-xx). AC/DC Input Loop Filter Audio Output, Speaker IEC 60601-1 Type B equipment. The unit displaying this symbol contains an adequate degree of protection against electric shock.
  • Page 149 X P R E Z Z O N A — S A N D Q U B E P P E N D I X Y M B O L S P E R A T I O N S A N U A L IEC 60601-1 Type BF equipment.
  • Page 150 X P R E Z Z O N A — S A N D Q U B E P P E N D I X Y M B O L S P E R A T I O N S A N U A L NIBP Cuff, Neonatal 3 NIBP Cuff, Neonatal 4...
  • Page 151 X P R E Z Z O N A — S A N D Q U B E P P E N D I X Y M B O L S P E R A T I O N S A N U A L NIBP Cuff, Neonatal 2 Size (4 to 8 cm) NEONATAL 2...
  • Page 152 X P R E Z Z O N A — S A N D Q U B E P P E N D I X Y M B O L S P E R A T I O N S A N U A L Keep Dry Indoor Use Only...
  • Page 153 X P R E Z Z O N A — S A N D Q U B E P P E N D I X Y M B O L S P E R A T I O N S A N U A L Sad Face PVC-Free (Polyvinyl Chloride)
  • Page 154 Does not contain hazardous substances — Europe Does not contain hazardous substances — China Batch Code Nellcor Oxisensor II Compatible Novametrix Compatible Spacelabs TruLink Compatible Nellcor OxiMax Compatible Spacelabs Compatible UL recognized component in Canada and United States Nellcor OxiMax Compatible...
  • Page 155 X P R E Z Z O N A — S A N D Q U B E P P E N D I X Y M B O L S P E R A T I O N S A N U A L ABBREVIATIONS USED AS SYMBOLS ARE SHOWN BELOW.
  • Page 156 X P R E Z Z O N A — S A N D Q U B E P P E N D I X Y M B O L S P E R A T I O N S A N U A L Nitrous Oxide NIBP...

This manual is also suitable for:

Qube913909139394267

Table of Contents