Troubleshooting - Yahorng BP-700W Instruction Manual

Bluetooth transmission
Table of Contents

Advertisement

TROUBLESHOOTING

Fault
Even though the batteries are
installed, there is either no
indication or an incorrect
indication on the LCD.
The cuff does not inflate
or the
air pressure cannot rise.
The low battery indication is
shown on the LCD.
The unit cannot take your
blood pressure, and the
LCD shows an error
message or a wrong
result.
Under normal measuring
circumstances a reading
at home is different from
one taken at a clinic or
two readings at home have
different values.
If you have taken all of the above countermeasures, and blood pressure
still cannot be read, then the unit should be replaced.
Remedy
1.
Check and correct the battery polarities.
2.
Remove the batteries and wait for one
minute. Then install the batteries or
replace the batteries.
1.
Check the cuff position, re-fasten the
cuff correctly and measure the
pressure again.
2.
Check the cuff connection to
the monitor.
1.
Replace the batteries.
2.
Once BP700W enter sleep mode
(Bluetooth icon disappear), the measurement
you saved in the device will not be vanished if
you replace the batteries in 18 seconds. This
function provide user enough time to install
new batteries.
1.
Re-fasten the cuff.
2.
Relax yourself and sit down.
3.
Keep the cuff and heart at the same
level during the measurement period.
4.
Keep silent and still during measurement.
5.
If the patient has a severe heart beat
6.
problem, then the blood pressure may
not be read correctly.
1.
The variation is due to the different
environments.
2.
Your blood pressure is changing
according to the physiological or
psychological status of the human body.
3.
Record your blood pressure each day and
consult with your physician.
14

Advertisement

Table of Contents
loading

Table of Contents