Cuff Sizes; Error Messages - Yahorng BP-700W Instruction Manual

Bluetooth transmission
Table of Contents

Advertisement

CUFF SIZES

Cuff sizes indicate the diameter of the upper arm.
Medium and Large cuff sizes are available.
Medium Cuff Size
Large Cuff Size

ERROR MESSAGES

Error
Cause
Message
Low voltage or internal
circuit problem
Abnormal situation
during inflation
Artificial interference and
noise
Abnormal measurement
Over safety pressure
Hardware broken
Low battery
Metric
English
23 to 33 cm
9" to 13"
33 to 43 cm
13" to 17"
Remedy
Remove and re-install the batteries. Measure again.
If the monitor still does not work, return the unit
for repair.
1.
Cuff is not fastened well, please
re-fasten again according to the
instructions.
2.
Cuff broken.
3.
Pump or venting valve failure,
replace the device.
1.
User moved during the measurement.
2.
Incorrect measurement procedure
and/or posture.
3.
User is too tense and not
relaxed.
When measuring, the value will increase and be
shown on the LCD.
1.
If the value is not over 300mmHg,
measure again. If it still does not
work, and user has severe heart beat
problem or other medical condition,
please contact a physician.
2.
If the value is over 300mmHg,
send the device for replacement.
1.
Air pipe blocked.
2.
Inflation controller failure.
3.
Pressure sensor fault.
Re-measure, otherwise contact
manufacturer for replacement.
Counsult with Service center for repair
Replace the batteries.
13

Advertisement

Table of Contents
loading