ECHNICAL DATA
• VapoMeter measuring unit
System components
• 2 standard (Ø 11 mm) and 2 small (Ø 4.5 mm) adapters
• User manual
• Carrying case
Accessories
• DMC data collection software
• Receiver unit
• RoomSensor for monitoring ambient RH and temperature
• Set of standard adapters
• Set of small adapters
• Set of in-vitro adapters
Power Supply
• Dimensions
VapoMeter measuring unit
• Weight with batteries
• Protection class
• Power consumption
• Measuring time
• Measurement range
• Water loss in g/m
• Coefficient variation (SD/mean) measured on normal skin typically < 5 %
• Wireless technology
2 x 1.5 V AA LR06 alkaline batteries
175 x 40 x 35 mm
135 g
IP52
operating 120 mW (0.6 mAh/measurement)
with the standard adapter 5 to 16 seconds
with the small adapter 9 to 32 seconds
0-300 g/m
with the standard adapter 3-200 g/m
with the small adapter 10-300 g/m
2
h, accuracy ±4 %
Bluetooth Smart (4.0)
2
h
15
VapoMeter SWL5 User Manual 1.3
10.11.2016
2
h
2
h