Pediatric-Pak User Instructions

Specifications

} Waveform: Biphasic Truncated Exponential
} PAD impedance range: 0 – 176Ω
} Energy Accuracy: 50J ±15%
Impedance, Ω
Energy, J
25
47.5
50
51.3
75
52.1
100
51.8
125
52.4
samaritan PAD
Waveform Voltages
Waveform Duration
Resistance
(Volts)
(Ohms)
V1
Tilt %
T1
25
514
55.6
7.8
50
671
50.4
8.8
75
751
47.1
10
100
813
44.3
10.8
125
858
41.4
11.5
} Sample Nominal Energy Delivered to a 50Ω load
Age
50TH Percentile
50J Energy Dose
(Yrs)
Weight* (Kg)
In Joules Per Kg
1
10.3
2
12.7
3
14.3
4
16.0
5
18.0
6
21.0
7
23.0
8
25.0
* Doses are based on CDC growth charts for the 50th percentile body weight of boys.
National Center for Statistics in collaboration with the National Center for Chronic
Disease Prevention and Health Promotion (2000).
(msec)
T3
5.4
6
6.6
6.8
7.3
4.9
4.0
3.5
3.2
2.8
2.4
2.2
2.0
6
Only samaritan PADs with the label
are suitable for use with the Pediatric-Pak. If the samaritan PAD you are
using does not have this label, use the Adult Pad-Pak if no alternatives are available.
Electrode Pads
} Conductive/Gel Surface Area: 100 cm
2
(200 cm
} Overall Size: 100 mm x 130 mm
} Standard Pad Placement: Anterior-Posterior or Anterior-Lateral
Environmental
} Operating Temperature: 32°F to 122°F (0°C to 50°C)
} Standby Temperature: 32°F to 122°F (0°C to 50°C)
} Relative Humidity: 5% to 95% (non-condensing)
} Atmospheric Pressure: 0 feet to 15,000 feet, 760 mm Hg to 439 mm Hg above sea level
} Shipping Temperature Range: 14°F to 122°F (-10°C to 50°C)
NOTE: 1. 'Shipping' temperatures are permitted for up to 48 hours. (If device has been stored below 14°F (-10°C) it should
be returned to an ambient temperature of between 32°F to 122°F (0°C to 50°C) for at least 24 hours before use.)