Hydrogen Cyanide Passive Sampler
Introduction
The SKC Hydrogen Cyanide (HCN) Passive Sampler, specifi ed in OSHA Method
1015, provides accurate sampling of hydrogen cyanide from 0.44 to 20 ppm for
either 15-minute or 8-hour samples. The uptake rate for the HCN Passive Sampler
is 28.4 ml/min. The sampler design allows the user to load 600 mg of soda lime
sorbent (supplied in a vial) into the sampler housing in the fi eld immediately before
sampling and transfer the sorbent from sampler to vial in the fi eld immediately
after sampling as specifi ed in the method. Samples are desorbed with water and
analyzed by ion chromatography/electrochemical detector. The HCN Passive
Sampler is designed for single use.
Collar Clip
Cover
Front View
of Housing
Performance Profi le
Sorbent/Amount:
Housing Material:
Diameter:
Length
(including clip):
Measuring Range:
Analysis:
Sample Storage:
Sample Time:
Sampling Rate:
Cat. No. 590-400
Diffusion Barrier
Vial
Adapter
Back View
of Housing
Soda Lime, 600 mg (supplied in glass vial)
Nylon
1.4 in (3.5 cm)
2.5 in (6.3 cm)
0.44 to 20 ppm
Solvent extraction; ion chromatography (IC) with electro-
chemical detector (ELCM)
3 weeks at 68 F (20 C) in storage studies
Note: OSHA Method 1015 recommends that samples be sent
to a laboratory as soon as possible. If delay is unavoidable, store
samples in a refrigerator as a precaution.
Validated for 15 min and 8 hrs
28.4 ml/min (4.40% RSD)
www.skcinc.com
Operating Instructions
863 Valley View Road, Eighty Four, PA 15330 USA
Tel: 724-941-9701 • www.skcinc.com
Collar
Clip
Plug
Vial containing
soda lime sorbent
(sorbent is white in
appearance)