Performance Profi le
Sampling Rate: 2 L/min respirable or thoracic, 4 or 8 L/min respirable
Sample Pump: • Universal XR or AirChek
• Leland Legacy for 8 L/min
Sample Time: Dependent on method used. Note: SKC tests indicate that a particulate
mass of up to 6.8 mg on the four impaction substrates would not aff ect
PPI performance. This amount is equivalent to sampling for 6 hours at
4 L/min in environments where respirable mass concentration is 4.76 mg/m
and equals 50% of total dust. However, labs have reported to SKC that they
prefer no more than 2 mg on the fi lter for analytical reasons. Therefore,
SKC recommends that you work with your lab to determine optimum
sample times for your unique sampling conditions.
Sample Media: Respirable models: 37-mm, 5.0-μm PVC fi lter and cellulose
support pad
Thoracic model: 37-mm, 0.8-μm MCE fi lter and cellulose
support pad or 37-mm Quartz fi lter and cellulose support pad
Impaction
Four 3/8-in diameter pre-oiled porous plastic discs
Substrate:
(Assembled in all Disposable PPIs)
Analysis:
Chemical or gravimetric
Body Material: Conductive ABS plastic
Dimensions:
Height:
Diameter: 1.8 in (4.6 cm)
Depth:
Weight:
1.1 oz (31.2 gm)
Shelf-life:
18 mos from date of manufacture
®
4.25 in (10.8 cm) - clip to exhaust
1.2 in (3 cm)
2
Series for 2 and 4 L/min
3
Need help?