Instructions For Use

Preparation and Setup of the Ranger Irrigation Fluid Warming Unit
1. Attach the Ranger warming unit to the I.V. pole. Tighten the pole clamp securely.
:
To prevent tipping, clamp the Ranger Irrigation
caution
warming unit to an I.V. pole with a minimum 14" (35.6 cm)
radius wheelbase and at a height no higher than 44" (112 cm).
Failure to do so may result in damage to the product (see figure 1).
2. Slide the irrigation fluid warming cassette into the slot in the
warming unit. The cassette can only fit into the device one way
(see figure 2).
3. Prime the disposable set. For more information about priming
the set, refer to instructions provided with the disposable sets.
4. Plug the power cord into an appropriate outlet. Turn the unit
ON. In a few seconds the alphanumeric display will illuminate.
It takes less than 2 minutes to warm up to the 41
temperature.
5. Begin infusion. When infusion is completed, remove the
disposable set and discard according to institutional protocol.
Removing the Irrigation Fluid Warming Disposable
Set from the Ranger Irrigation Warming Unit
1. Close the inlet clamp proximal to the fluid warming cassette
and open all clamps distal to the cassette.
2. Disconnect the disposable set from the fluid source, if
applicable.
3. Allow a small amount of fluid to drain from the end of the
patient line to reduce pressure in the cassette.
4. Remove the irrigation fluid warming cassette from the warming
unit and discard according to institutional protocol.
Ranger Irrigation Fluid Warming System
44" (112 cm)
C set point
°
Figure 1
Figure 2
English
14" (35.6 cm)
7