Model 247 Product Safety Features - Ranger 247 Operator's Manual

Irrigation fluid warming system
Hide thumbs Also See for 247:
Table of Contents

Advertisement

English
Ranger Irrigation Fluid Warming System

Model 247 Product Safety Features

The following chart describes the safety alarm features of the Ranger Irrigation Fluid warming unit.
ALARM TyPE
ALARM TyPE
Primary over-temperature
Primary over-temperature
alarm system
alarm system
Independent back-up
Independent back-up
over-temperature alarm
over-temperature alarm
system (cut-off).
system (cut-off).
Under-temperature alarm
Under-temperature alarm
:
The difference in over-temperature alarm system cut-off points does not indicate any difference
note
in safety and is simply a result of tighter controller tolerances.
The Ranger Irrigation Fluid Warming System is designed with very tight controls to regulate its temperature
to the 41°C set-point. The microprocessor controller measures the temperature at the fluid pathway inside the
warming plate four times every second, making it very responsive to changes in flow rate.

SET-POINT
SET-POINT
INDICATION
INDICATION
48°C
48°C
LED alternately displays
LED alternately displays
"HI" and temperature,
"HI" and temperature,
alarm sounds, alarm light
alarm sounds, alarm light
flashes.
flashes.
50°C
50°C
LED displays the
LED displays the
temperature or has gone
temperature or has gone
dark, alarm sounds.
dark, alarm sounds.
33°C
33°C
LED displays "LO"
LED displays "LO"
and temperature, alarm
and temperature, alarm
sounds, alarm light
sounds, alarm light
flashes.
flashes.
DESCRIPTION
DESCRIPTION
Primary Alarm: Alarm sounds,
Primary Alarm: Alarm sounds,
controller cuts power to the
controller cuts power to the
heaters until temperature falls
heaters until temperature falls
below 41°C. When heater
below 41°C. When heater
temperature drops below 41°C
temperature drops below 41°C
(in a few seconds), audible and
(in a few seconds), audible and
visual alarms stop.
visual alarms stop.
Primary failure. Temperature
Primary failure. Temperature
will rise to 50°C within a few
will rise to 50°C within a few
seconds. The safety backup
seconds. The safety backup
system is activated at 50°C and
system is activated at 50°C and
power is immediately shut off
power is immediately shut off
to the
to the
heating plates.
heating plates.
Heater temperature has fallen
Heater temperature has fallen
to 33°C. Alarms will stop
to 33°C. Alarms will stop
when temperature rises above
when temperature rises above
33°C. Continue use of unit.
33°C. Continue use of unit.