Blood/fluid warming system and irrigation warming system (18 pages)
Summary of Contents for Ranger 247
Page 1
Check the Ranger system website to ensure you have the most recent version of this document. www.rangerfluidwarming.com reorder #247020B Operator's Manual Model 247 Irrigation Fluid Warming System...
Ranger Irrigation Fluid Warming Unit Ranger Irrigation Fluid Warming Disposable Set Model 247 Product Safety Features Instructions For Use Preparation and Setup of the Ranger Irrigation Warming Unit Removing the Irrigation Fluid Warming Disposable Set from the Ranger Irrigation Warming Unit Troubleshooting...
English Ranger Irrigation Fluid Warming System Technical Service and Order Placement Technical Service arizant TEL: +1-952-947-1200 +1-800-733-7775 FAX: +1-952-947-1400 arizant uk ltd. TEL: +44-1924-200-550 FAX: +44-1924-200-518 arizant deutschland TEL: +49-4154-9934-0 FAX: +49-4154-9934-20 arizant france TEL: +33 4 42 98 14 00...
Do not immerse the warming unit. Wipe with a clean, slightly damp cloth. • To prevent tipping, clamp the Ranger Irrigation warming unit to an I.V. pole with a minimum 14" (35.6 cm) radius wheelbase and at a height no higher than 44" (112 cm). Failure to do so may result in...
English Ranger Irrigation Fluid Warming System Product Description The Ranger Irrigation Fluid Warming System consists of a warming unit and a sterile disposable fluid delivery set. Alarm indicator light I.V. pole clamp Power switch Alphanumeric Slot for display warming cassette...
The Ranger Irrigation Fluid Warming System is designed with very tight controls to regulate its temperature to the 41°C set-point. The microprocessor controller measures the temperature at the fluid pathway inside the...
Ranger Irrigation Fluid Warming System Instructions For Use Preparation and Setup of the Ranger Irrigation Fluid Warming Unit 1. Attach the Ranger warming unit to the I.V. pole. Tighten the pole clamp securely. To prevent tipping, clamp the Ranger Irrigation caution warming unit to an I.V.
English Ranger Irrigation Fluid Warming System Troubleshooting CONDITION CAUSE SOLUTION Nothing illuminates on the • Unit is not turned on, plugged in, • Turn unit on. Make sure the warming unit panel. or power cord is not plugged into power cord is plugged into the an appropriate outlet.
Page 9
English Ranger Irrigation Fluid Warming System CONDITION CAUSE SOLUTION • Unit alarms soon after plugging it Test screw on bottom of unit is Make sure test screw is completely in (unit does not have to be turned loose or missing.
3. Wipe with a dry, soft cloth. to clean the heating plates: The Ranger Irrigation hardware cleaning tool is intended to clean both heating plates of the warming unit. It is not necessary to disassemble the warming unit to use the tool.
English Ranger Irrigation Fluid Warming System Specifications Physical Characteristics warming unit 4.5 in. (11 cm) high x 7.5 in. (19 cm) wide x 10 in. (25 cm) long; wt.: 7 lb. 7 oz. (3.4 kg) classification Classified under IEC 601-1 Guidelines (and other national versions of the Guidelines) as Class I, Type B, Ordinary Equipment.
English Ranger Irrigation Fluid Warming System Definition of Symbols Attention - Read accompanying documents Attention - Read accompanying documents Type B Applied Part Type B Applied Part...
Page 13
Wakefield WF2 7AW, United Kingdom Tel (44) 1924 200550 • Fax (44) 1924 200518 *Ranger, and the Bair Hugger and Arizant logos are trademarks of Arizant Healthcare Inc., registered or pending in the U.S. Patent & Trademark Office and in other countries.