Guided wave radar 4...20 ma/hart - our-wire rod and cable probe -196...+280 c; -196...+450 c
Table of Contents

Advertisement

Menu - Additional adjustments
Menu
Additional adjustments
Additional settings - Copy
instrument settings
Additional settings - Scal-
ing level
PBD-51041064
Menu item
PIN
Date
Time
Time - Format
Probe type
The instrument settings are copied with this function. The following
functions are available:
Read from sensor: Read data from sensor and save in the display
and adjustment module
Write to sensor: Save data from the display and adjustment mod-
ule back into the sensor
The following data or settings for adjustment of the display and ad-
justment module are saved:
All data of the menu "Setup" and "Display"
In the menu "Additional adjustments" the items "Reset, Date/Time"
Special parameters
The copied data are permanently saved in an EEPROM memory in
the display and adjustment module and remain there even in case of
power failure. From there, they can be written into one or more sen-
sors or kept as backup for a possible electronics exchange.
Note:
Before the data are stored in the sensor, a check is carried out to
determine if the data fit the sensor. If the data do not fit, a fault signal
is triggered or the function is blocked. When data are being written
into the sensor, the display shows which instrument type the data
originate from and which TAG-no. this sensor had.
Tip:
We recommend to save the instrument adjustments. In case of an
electronics exchange the saved parameter adjustment data relieve
this process.
Since scaling is very extensive, scaling of the level value was divided
into two menu items.