Specifications

Biosensor specifications

Hardware
Size (W x H x D)
Weight
Battery
Memory
Robustness
Ingress Protection
Manufactured with Latex
Use
MRI Safe
Singe Use
Disposable
Serviceable
Performance
Heart Rate Measurement Range
Heart Rate Accuracy
Heart Rate Resolution
Heart Rate Calculation
Heart Rate Sampling Rate
Heart Rate Meter Accuracy and
Response to Irregular Rhythm
Response time of heart rate meter to
change in heart rate
34
100mm x 69mm x 6.2mm ±5% (without
the release liners)
12 g ±10%
CR2032, 3V primary cell
1MB non-volatile flash
Survives shock, vibration, free fall,
and bump
IP27
No
No
Yes
Yes
No
30-220 bpm (beats per minute)
10% or ±5bpm (whichever is greater)
1 bpm
Heart Rate is calculated:
• Taking into account last 10 beat-to-
beat intervals
• Excluding the minimum and the
maximum intervals
• Averaging the remaining eight
intervals to compute "mean_interval"
and
• Compute 60/mean_interval (in
seconds) to convert to bpm
250 samples per second
Provides correct heart rates (60, 80,
90, 120 bpm) using test waveforms
as indicated in ANSI/AAMI EC13 Sec.
4.1.2.1(e). All QRS are counted with test
waveforms within HR accuracy defined
above
Inside a reporting interval,
instantaneous heart rate change from
80 bpm to 120 bpm shall be captured
within 8 seconds (margin of +2 sec)