Harvest Healthcare SALISBURY HS310 User Manual

Active overlay system mattress

Advertisement

Quick Links

General User/ Safety Guide
SALISBURY
ACTIVE OVERLAY
SYSTEM
www.harvesthealthcare.co.uk
E sales@harvesthealthcare.co.uk
www.harvesthealthcare.co.uk
1

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the SALISBURY HS310 and is the answer not in the manual?

Questions and answers

Summary of Contents for Harvest Healthcare SALISBURY HS310

  • Page 1 General User/ Safety Guide SALISBURY ACTIVE OVERLAY SYSTEM www.harvesthealthcare.co.uk E sales@harvesthealthcare.co.uk www.harvesthealthcare.co.uk...
  • Page 2 T +44 (0)1709 377172 F +44 (0)1709 377173...
  • Page 3: Table Of Contents

    CONTENTS WARNINGS & CAUTIONS GENERAL INFORMATION 6-10 DEFINITION OF THE GROUPS MENTIONED NON-COMPLIANT USE SAFETY INSTRUCTIONS GENERAL SAFETY INSTRUCTIONS SAFETY INFORMATION FOR THE OPERATOR SAFETY INFORMATION FOR THE USER SYMBOLS USED CLEANING & DISINFECTION SERVICING & MAINTENANCE SERVICE LIFE & DISPOSAL TECHNICAL SPECIFICATION OVERVIEW 12-13...
  • Page 4: Warnings & Cautions

    Do not place device on or near a heat source or cover pump with bedding. • Harvest Healthcare advise against smoking whilst the system is in use, to prevent the accidental ignition of associated items which may be flammable, such as bed linen.
  • Page 5: Warnings & Cautions

    WARNINGS & CAUTIONS • Do no expose the pump to liquids. • Do not use with hot water bottles or electric blankets. • Wireless equipment such as mobile phones should be kept at least 10ft / 3m away from the system. •...
  • Page 6: General Information

    Read through this instruction manual conscientiously from start to finish. Please note that the various safety instructions must be observed. Harvest Healthcare products bear the CE mark and meet all safety and functionality requirements. These safety requirements can only be met if the user is satisfied with the proper condition of the product (including accessories) before use.
  • Page 7: Non-Compliant Use

    / care personnel. Programming of the system should be carried out by competent trained personnel. Use only spares and accessories supplied or approved by Harvest Healthcare. Only suitably trained personnel are allowed to operate the system. E sales@harvesthealthcare.co.uk...
  • Page 8: Safety Information For The Operator

    GENERAL INFORMATION The mains cable must be free and not be allowed to be caught up in the bed’s moving mechanisms. The mains cable may be damaged or it may be pulled out of its socket and electric leads exposed as a result. If the mains cable or the mains plug are damaged, the relevant part must be replaced.
  • Page 9: Symbols Used

    GENERAL INFORMATION SYMBOLS USED This symbol indicates general hazards. There is a danger to life and health. Conformity mark in accordance with the Medical Device Directive (93/42 EEC). The symbol for Protection Class II device, double insulated. The symbol for type B device according to EN 60601-1. Handle with care This way up Keep dry...
  • Page 10: Cleaning & Disinfection

    Any defects, damage or signs of wear must be rectified without delay. Only original spare parts from Harvest Healthcare may be used, otherwise all guarantees or warranties will be excluded. See pages 20-21 for Routine Maintenance and Servicing.
  • Page 11: Technical Specification

    TECHNICAL SPECIFICATION SALISBURY TECHNICAL SPECIFICATION Product Code HS310 Pump Model No. HUP1 Pressure Sore Risk Level High Risk Minimum Patient Weight 5 Stone / 32 kg Maximum Patient Weight 24 Stone / 152 kg Inflated Mattress Dimensions 1970 x 900 x 120 mm Mattress Weight 6 kg Operating Cycle...
  • Page 12: Overview

    OVERVIEW SYSTEM OVERVIEW CPR Valve The mattress can be rapidly deflated through the use of the CPR valve, allowing emergency personnel to begin resuscitation. Salisbury Cells The Salisbury has 17 independent cells which can be individually removed and replaced to allow for cost effective repair and in-depth cleaning.
  • Page 13 OVERVIEW PUMP OVERVIEW Feed Tube Connection Visual Low-Pressure Warning w/ Alarm Mute Manual Comfort Control Dial Bed Hooks Power Switch Non-slip Feet Static Mode Static Mode is only to be used for patient care. This must be manually switched off after use. E sales@harvesthealthcare.co.uk www.harvesthealthcare.co.uk...
  • Page 14 INSTALLATION INSTALLING THE SALISBURY SYSTEM The Salisbury is an alternating overlay system to be used in conjunction with a foam underlay. This mattress should NOT be used as a replacement system. Remove the mattress from its packaging and lay the parts out on the floor. You should have the following items: •...
  • Page 15: Installation

    INSTALLATION The pump has integral bed hooks for hanging the pump on the foot end of the bed. The pump and feed tubes should be at the foot end of the bed. Check that the CPR valve is set to the CLOSED setting as indicated below. The mattress will not inflate if the CPR valve is open.
  • Page 16: Operation

    OPERATION INFLATING THE MATTRESS Press the power On/Off switch to the On position: the switch should illuminate to show it has a power supply. The low pressure warning light will now illuminate to indicate that the mattress is being inflated. Once the mattress is up to the correct pressure, the low pressure warning light will go off.
  • Page 17 OPERATION SETTING PROCEDURE It is important to follow the correct setting procedure to ensure the patient receives sufficient support whilst achieving maximum pressure relief and comfort. Failure to follow this procedure could result in the patient being put at risk. Position the patient in a lying position on the mattress.
  • Page 18: Cleaning & Care

    Eye protection, gloves and protective clothing should be worn when carrying out cleaning and disinfection procedures When disinfecting the system, Harvest Healthcare recommends the following guidelines which have been developed to comply with recognised infection control procedures. These procedures are also to be used to prevent cross infection when transferring the system between patients.
  • Page 19: Cleaning & Care

    CLEANING & CARE PUMPS For general cleaning wipe with a soft cloth dampened with a mild detergent and water solution. This may be followed by either wiping with a sodium hypochlorite solution to a dilution of 1000ppm (parts per million) or by using alcohol wipes. DO NOT USE HYPERCARBONATE, PHENOL-BASED CLEANING SOLUTIONS, ABRASIVE COMPOUNDS OR CLEANING PADS.
  • Page 20: Routine Maintenance

    • The individual cells can be wiped clean with a mild antiseptic solution. • All cells are replaceable and can be sourced from Harvest Healthcare. POWER UNIT Disconnect the power unit from the electricity supply before carrying out maintenance, repairs or cleaning.
  • Page 21: Routine Maintenance

    WEEE requirements, which apply to the pump and SMPS only. SERVICING YOUR SYSTEM The Salisbury system should be serviced every 12 months by Harvest Healthcare approved personnel using genuine Harvest Healthcare spare parts. Due to the risk of electric shock, do not attempt to open the main unit. Please refer any repairs to a qualified dealership or service technician.
  • Page 22: Troubleshooting

    T +44 (0)1709 377172 F +44 (0)1709 377173...
  • Page 23 E sales@harvesthealthcare.co.uk www.harvesthealthcare.co.uk...
  • Page 24 T +44 (0)1709 377172 F +44 (0)1709 377173...
  • Page 25 E sales@harvesthealthcare.co.uk www.harvesthealthcare.co.uk...
  • Page 26: Parts List

    PARTS LIST APPLIED PARTS HS314 Salisbury Mattress (only) *HPU1 Harvest 1 Pump (only) HS312 Salisbury Top Cover and Base HS306 Salisbury Inner (Complete) CPR Valve HP574/2 Seat Cushion HH21 REPLACEMENT PARTS Cell HS3104/90 Salisbury Inner Base Sheet HS3101/90 HP597 CPC (Female) 10mm HP586 CPC Male) 10mm Pump Outlet Pipes...
  • Page 27: Guarantees & Warranties

    GUARANTEES & WARRANTIES MATTRESS (COVER AND INTERIOR COMPONENTS) All Harvest Healthcare Ltd Mattress are covered by warranty for a period of 12 months from date of purchase. Damage through incorrect use and penetration by sharp instruments will invalidate this warranty.
  • Page 28: Declaration Of Conformity

    DECLARARTION OF CONFORMITY Declaration of Conformity Annex VII EU Directive 93/42/EEC Harvest Healthcare Limited We, as company: Sheaf House Bradmarsh Way Bradmarsh Business Park Rotherham, S60 1 BW Confirm on our own behalf that the medical product: SALISBURY OVERLAY ACTIVE MATTRESS SYSTEM Complies with all applicable requirements in Appendix I of the EU directive 93/42/EEC.
  • Page 29 NOTES ............................................................................................................................................................................................................................................................................................................................................................................................................ E sales@harvesthealthcare.co.uk www.harvesthealthcare.co.uk...
  • Page 30 NOTES ............................................................................................................................................................................................................................................................................................................................................................................................................ T +44 (0)1709 377172 F +44 (0)1709 377173...
  • Page 31 NOTES ............................................................................................................................................................................................................................................................................................................................................................................................................ E sales@harvesthealthcare.co.uk www.harvesthealthcare.co.uk...
  • Page 32 Harvest Healthcare Limited. Company No: 07210261 Our Full Terms & Conditions are available by request or can be found on our Website. Harvest Healthcare reserves the right to alter or amend this document without notice. T +44 (0)1709 377172 F +44 (0)1709 377173...

Table of Contents