LR-Cal FLUID 100-N Operating Manual

Portable temperature calibraon bath
Hide thumbs Also See for FLUID 100-N:
Table of Contents

Advertisement

Quick Links

Portable Temperature Calibra on Bath
MANUAL
FLUID 100 and FLUID 100-N
Rel. 20140812
Opera ng Manual
LR-Cal
FLUID 100
LR-Cal
FLUID 100-N
Portable temperature calibra on bath
DRUCK & TEMPERATUR Leitenberger GmbH • Bahnhofstr. 33 • D-72138 Kirchentellinsfurt • GERMANY
Tel. +49 (0) 7121-90920-0 • Fax +49 (0) 7121-90920-99 • DT-Info@Leitenberger.de •
www.LR-Cal.net
• www.druck-temperatur.de
1

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the FLUID 100-N and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for LR-Cal FLUID 100-N

  • Page 1 Portable Temperature Calibra on Bath MANUAL FLUID 100 and FLUID 100-N Rel. 20140812 Opera ng Manual LR-Cal FLUID 100 LR-Cal FLUID 100-N Portable temperature calibra on bath DRUCK & TEMPERATUR Leitenberger GmbH • Bahnhofstr. 33 • D-72138 Kirchentellinsfurt • GERMANY Tel.
  • Page 2: Table Of Contents

    9.2 Func ons of the 2. Func on Menu Level 9.3 Func ons of the 3. Func on Menu Level 10. Recalibra on of the LR-Cal temperature calibrator 10.1 Calibra on of the internal probe 10.2 Calibra on of the "EXT" and "REF" inputs with a signal calibrator - only-2I versions 10.3 Calibra on of the "REF"...
  • Page 3: Safety Instruc Ons

    Portable Temperature Calibra on Bath MANUAL FLUID 100 and FLUID 100-N 1. Safety instruc ons 1.1 Used symbols WARNING! Hot surface or part. CAUTION! Indicates a poten ally dangerous situa on, that can result in injuries or damage to equipment, personnel or environment, if not avoided.
  • Page 4 Portable Temperature Calibra on Bath MANUAL FLUID 100 and FLUID 100-N WARNING! Due to the fact that the temperature calibrator is a portable instrument, it is very important to ensure that the socket has been earthen correctly when connec ng it to the electricity supply.
  • Page 5 Portable Temperature Calibra on Bath MANUAL FLUID 100 and FLUID 100-N A er any extended use at a temperature below zero: • Bring the Set Point to about 95...100°C le ng the temperature calibrator work ll the water inside the oil has completely evaporated.
  • Page 6: Technical Data

    Portable Temperature Calibra on Bath MANUAL FLUID 100 and FLUID 100-N 2. Technical data 2.1 Purpose and summary of instruc ons This manual contains the use and maintenance instruc ons valid for the following equipment: Portable Temperature Calibra on Bath...
  • Page 7: Name Of The Product

    Portable Temperature Calibra on Bath MANUAL FLUID 100 and FLUID 100-N 2.2 Name of the product Portable temperature calibra on bath FLUID 100 resp. FLUID 100-N , including standard accessories as per chapter 2.9. 2.3 Technical data Environmental temperature range: +5°C…+45°C...
  • Page 8: Intended Use

    Portable Temperature Calibra on Bath MANUAL FLUID 100 and FLUID 100-N Controller Data: Display 2 lines (3.2 x 5.5), with back light Microprocessor: 80C522 (80C51 family, CMOS) Memory: 8K SRAM (ultra-low-power SRAM, crystal, power-fail control and ba ery A/D converter: 24 bit E2PROM-memory, serial communica on RS232 insulated.
  • Page 9: Drawing

    Portable Temperature Calibra on Bath MANUAL FLUID 100 and FLUID 100-N 2.8 Drawing DRUCK & TEMPERATUR Leitenberger GmbH • Bahnhofstr. 33 • D-72138 Kirchentellinsfurt • GERMANY Tel. +49 (0) 7121-90920-0 • Fax +49 (0) 7121-90920-99 • DT-Info@Leitenberger.de • www.LR-Cal.net • www.druck-temperatur.de...
  • Page 10 Portable Temperature Calibra on Bath MANUAL FLUID 100 and FLUID 100-N Electrical drawing DRUCK & TEMPERATUR Leitenberger GmbH • Bahnhofstr. 33 • D-72138 Kirchentellinsfurt • GERMANY Tel. +49 (0) 7121-90920-0 • Fax +49 (0) 7121-90920-99 • DT-Info@Leitenberger.de • www.LR-Cal.net • www.druck-temperatur.de...
  • Page 11: Scope Of Standard Delivery

    Portable Temperature Calibra on Bath MANUAL FLUID 100 and FLUID 100-N 2.9 Scope of standard delivery 1 bo le silikon oil 47V5 at model LR-Cal FLUID 100 1 bo el silikon oil 47V10 at modell LR-Cal FLUID 100-N Cover for transport and emptying...
  • Page 12: Op Onal Accessories

    Portable Temperature Calibra on Bath MANUAL FLUID 100 and FLUID 100-N Cables for thermostat- / switch-test So bag Opera ng manual Test cer ficate (accuracy, stability, final check) 2.10 Op onal Accessories Order-Code: FLUID-INS-9 or FLUID-INS-0 Kit to convert to metal dry block calibrator...
  • Page 13: Calibra On Media

    Portable Temperature Calibra on Bath MANUAL FLUID 100 and FLUID 100-N 3. Calibra on media CAUTION The temperature calibrator has been configured in the factory with the parameters that op mize the performance of the equipment. Modifica on of these parameters (3. Menu Func on Level) can cause malfunc oning or breakage of the equipment with consequent risk of damage to objects or injury to persons.
  • Page 14: Viscosity

    Portable Temperature Calibra on Bath MANUAL FLUID 100 and FLUID 100-N 3.1.2 Viscosity Viscosity is the property of fluids that indicates it‘s resistance to mixing and it depends on the type of fluid and temperature. The more viscous the fluid, the more difficult the mixing of the fluid and consequently the distribu on of a uniform temperature inside the bath.
  • Page 15: S Rrer Adjustment

    Portable Temperature Calibra on Bath MANUAL FLUID 100 and FLUID 100-N Using 47V5, bath performances are op mal in the field from -20° to 120°C, but with a fumes suc on device it can be used up to 125°C, limi ng the running me to the highest values in order to increase the life span, while at a low temperature the viscosity, even if high, allows good func oning of the temperature calibrator, adjus ng the magne c s rrer to maximum speed.
  • Page 16: Recommenda Ons

    Portable Temperature Calibra on Bath MANUAL FLUID 100 and FLUID 100-N 3.1.8 Recommenda ons • Do not close the container when the temperature is very high. Wait for the temperature of the liquid to reach room temperature and then turn off and close using appropriate cap and cover.
  • Page 17: Installa On And Opera On

    Portable Temperature Calibra on Bath MANUAL FLUID 100 and FLUID 100-N 4. Installa on and Opera on 4.1 Installa on 4.1.1 Removal of packaging The temperature calibrator is equipped with suitable packaging for transport and tradi onal shipping systems. Any damage caused during transport must be no fied immediately to the transporters and claim must be made.
  • Page 18: Opera Ng The Calibrator

    Portable Temperature Calibra on Bath MANUAL FLUID 100 and FLUID 100-N 5. Opera ng the calibrator 5.1 Elements of the front panel Supply socket Main switch Protec on fuses Switch test plugs RS232 interface Mixer speed control Display (7.1) LED „hea ng“...
  • Page 19: Descrip On Of The Instrument

    Portable Temperature Calibra on Bath MANUAL FLUID 100 and FLUID 100-N 5.2 Descrip on of the instrument FLUID 100 FLUID 100-N temperature calibrator consists of an aluminium block with one hole into which the sensors to be calibrated are inserted. The capacity of the fluid tank is 500 cm³.
  • Page 20 Portable Temperature Calibra on Bath MANUAL FLUID 100 and FLUID 100-N You have three op ons to calibrate test items in the means of comparison of values: Calibra on with comparison with internal reference sensor Make reference to the temperature value of the display (7).
  • Page 21 Portable Temperature Calibra on Bath MANUAL FLUID 100 and FLUID 100-N Before any calibra ons, follow the instruc ons in chapter 3, and: • Start the calibra on only at ambient temperature. Thermal shock can break the sensi ve element of the probe and cause harm to operator.
  • Page 22: Use Of The Func Ons

    Portable Temperature Calibra on Bath MANUAL FLUID 100 and FLUID 100-N 5.4 Use of the func ons 5.4.1 Reading external probes - only versions You can connect up to two external temperature probes to the temperature calibrator (only -2I versions) and show their values on the temperature calibrator‘s display (2nd.
  • Page 23: Switch Test

    Portable Temperature Calibra on Bath MANUAL FLUID 100 and FLUID 100-N NOTE: The temperature calibrator always uses the internal reference sensor to regulate it‘s temperature. Possible Error Messages: • EXT SENSOR FAIL Wrong connec on or configura on of the probe connected to „EXT“.
  • Page 24: Maintenance

    Portable Temperature Calibra on Bath MANUAL FLUID 100 and FLUID 100-N 6. Maintenance 6.1 Periodical Maintenance • Check once a month the level of the liquid inside the fluid container. • Check once a month the liquid and change if it is exhausted.
  • Page 25 Portable Temperature Calibra on Bath MANUAL FLUID 100 and FLUID 100-N Cleaning of the tank (container): To take off the protec on grid, it is necessary to unscrew the insula ng ring nut (A). Remove the spring (B) and then hook the grid with a wire (C). Under the grid, there is the s r bar for mixing the fluid (D).
  • Page 26: Typical Faults

    Portable Temperature Calibra on Bath MANUAL FLUID 100 and FLUID 100-N 7. Typical faults Fault: The calibrator does not work when the power cable is connected and the main switch is turned on. Cause: • The fuse (3) is cut off. Replace the fuse.
  • Page 27: Descrip On Of The Front Panel

    Portable Temperature Calibra on Bath MANUAL FLUID 100 and FLUID 100-N 8. Descrip on of the front panel EXT.: external probe input REF.: external reference probe input Selectable between , selectable between: Pt 100 (4 or 3 wires), Pt 100 (4 or 3 wire),...
  • Page 28: Descrip On Of The Func On Menu Levels

    Portable Temperature Calibra on Bath MANUAL FLUID 100 and FLUID 100-N 9. Descrip on of the Func on Menu Levels Level 1 Level 2 Temp. Value Temp. Value Set Point P. B. new value new value SP.2 new value T. i.
  • Page 29: Func Ons Of The 1. Func On Menu Level

    Portable Temperature Calibra on Bath MANUAL FLUID 100 and FLUID 100-N LR-Cal temperature calibrator has three func on menu levels (see previous page): • Level 1: Func ons for the con nuous usage • Level 2: More specific func ons and for the regula on of the calibrator •...
  • Page 30 Portable Temperature Calibra on Bath MANUAL FLUID 100 and FLUID 100-N Ramp procedure effec ve example: Let‘s say, that the set temperature is the ambient one and that it is necessary to reach 400°C with a gradient of 2°C/min. • Press „F“-key, set „SP2" to „400°C“ using the arrow-keys. Press „E“-key to confirm the choice.
  • Page 31: Func Ons Of The 2. Func On Menu Level

    Portable Temperature Calibra on Bath MANUAL FLUID 100 and FLUID 100-N 9.2 Func ons of the 2. Func on Menu Level This func ons can be selected by pressing the „F“-key together with the „arrow-up“-key at the same me. Once the 2. Func on Menu Level is intered, these func ons can be selected only by pressing the „F“-key.
  • Page 32: Func Ons Of The 3. Func On Menu Level

    Portable Temperature Calibra on Bath MANUAL FLUID 100 and FLUID 100-N • DEF. Default Parameters: this func on allows choosing to set the thermoregulator with the P.B., T.i., T.d. parameters either as a default or as a customisable adjustment. By selec ng the „OFF“ parameter and confirming by the „E“-key, it is possible to modify the adjustment parameters, which will keep...
  • Page 33 Portable Temperature Calibra on Bath MANUAL FLUID 100 and FLUID 100-N • ADRESS Communica on adress: the value of this parameter is necessary to communicate from the PC to many LR-Cal temperature calibrators. The admi ed values are from 1 to 32 and once the value is set by using the arrow-keys, it is necessary to confirm the choice by pressing the „E“-key.
  • Page 34 Portable Temperature Calibra on Bath MANUAL FLUID 100 and FLUID 100-N • P1 First calibra on point. Press arrow-keys to set the value read with a precision reference thermoemter and press „E“-key to accept. • P2 Second calibra on point. Press arrow-keys to set the value read with a precision reference thermometer and press „E“-key to accept.
  • Page 35: Recalibra On Of The Lr-Cal Temperature Calibrator

    Portable Temperature Calibra on Bath MANUAL FLUID 100 and FLUID 100-N 10. Recalibra on of the LR-Cal temperature calibrator CAUTION: This sec on is applicable for authorised personnel only. Personnel must be able to operate the calibrator and must be familiar with the menu item selec on etc. Deep knowledge about calibra on of temperature instruments is necessary.
  • Page 36: Calibra On Of The "Ext" And "Ref" Inputs With A Signal Calibrator - Only-2I Versions

    Portable Temperature Calibra on Bath MANUAL FLUID 100 and FLUID 100-N • Enter the 3. Func on Menu Level and select „P2". Use the arrow-keys to enter the temperature value, indicated by the pecision reference thermometer and press „E“-key to confirm. Your input is confirmed by symbol „*“...
  • Page 37 Portable Temperature Calibra on Bath MANUAL FLUID 100 and FLUID 100-N • Enter the 3. Func on Menu Level (see instruc ons) and select Cal_chnl = REF. Press E key to „ “ „ “- confirm. • Select P1 and press „...
  • Page 38: Communica On Protocol Rs232/C

    Portable Temperature Calibra on Bath MANUAL FLUID 100 and FLUID 100-N 11. Communica on protocol RS232/C This chapter is needed only, if you like to communicate with the temperature calibrator with your OWN so ware. This instruc ons are intended to be used by so ware engineers only.
  • Page 39 Portable Temperature Calibra on Bath MANUAL FLUID 100 and FLUID 100-N Each command string is an ASCII character succession. First is $ character, the next must indicate the instrument adress (default = 1) and than follows the commanc (4 characters):...
  • Page 40 Portable Temperature Calibra on Bath MANUAL FLUID 100 and FLUID 100-N If the temperature unit is not °C, but e.g. °F, you first should write the variable no. 10 (units) to the required value, see table of variables. Data wri ng (INTEGER variables) Above is shown how to write into variables of type „float“.
  • Page 41: Instruc Ons For Op Onal Accessories

    Portable Temperature Calibra on Bath MANUAL FLUID 100 and FLUID 100-N 12. Instruc ons for op onal Accessories 12.1 Instruc ons for Metal Dry Block Insert Kit Order-Code for kit with metal block blank, without drilling(s): FLUID-INS-0 Order-Code for kit with metal block with 9 holes: FLUID-INS-9 (Drillings with 4 / 4 / 4.5 / 5.5 / 6.5 / 6.5 / 8.5 / 10.5 / 12.5 mm diameter)
  • Page 42: Instruc Ons For The Extension Tube For Increasing The Inser On Depth

    Portable Temperature Calibra on Bath MANUAL FLUID 100 and FLUID 100-N 12.2 Instruc ons for the Extension Tube for increasing the inser on depth Order-Code: FLUID100-ER With this op onal accessory the useable inser on depth of the reservoir is increased.
  • Page 43: Declara On Of Conformity

    Portable Temperature Calibra on Bath MANUAL FLUID 100 and FLUID 100-N 13. Declara on of Conformity DRUCK & TEMPERATUR Leitenberger GmbH, Bahnhofstr. 33, D-72138 Kirchentellinsfurt, GERMANY, declares that the • Portable temperature calibrator - temperature calibra on bath FLUID 100...
  • Page 44: Druck & Temperatur Leitenberger Gmbh Bahnhofstr. 33, D-72138 Kirchentellinsfurt, Germany

    Portable Temperature Calibra on Bath MANUAL FLUID 100 and FLUID 100-N DRUCK & TEMPERATUR Leitenberger GmbH Bahnhofstr. 33, D-72138 Kirchentellinsfurt, Germany Tel.: +49 7121-90920-0 Fax: +49 7121-90920-99 E-Mail: DT-Export@Leitenberger.de www.LR-Cal.net DRUCK & TEMPERATUR Leitenberger GmbH • Bahnhofstr. 33 • D-72138 Kirchentellinsfurt • GERMANY Tel.