Detector Operations

Activate the detector

Turn-on the detector in a safe area with an atmosphere of
20.9% oxygen and free of hazardous gas.
1. For first time use, charge the battery for up to 8 hours
or until LED light turns green using the charging
adapter provided. Refer to Charge the battery for more
information.
2. Press and hold the button for three seconds.
3. For first time use, a Warming sensors message is
displayed and a 30 minutes countdown is displayed.
In most cases, this countdown only lasts a couple of
minutes.
4. When the detector displays Pump test Block inlet,
block the pump inlet with a finger, and then after a
couple of seconds unblock the pump inlet.
The detector performs a quick pump test. A Pump Test
passed message is displayed.
If you do not block the pump inlet, the detector will turn
off after two minutes.
The detector will then perform a self-test, including
testing the sensors. This process could take several
minutes.
If necessary, you will be instructed by screen prompts to
calibrate newly-installed sensors.
5. When the self-test is complete, press and hold the
button to zero sensors. After the autozero is complete,
the detector then checks the sensors for calibration and
bump test.
If the detector identifies sensors requiring calibration or
bump testing, press the button and follow the on-screen
instructions.

Activate the backlight

To activate the backlight on the screen, press the button.

Deactivate the detector

1. Press and hold the button during the powering-off
countdown.
2. Release the button when OFF is displayed.
6

Navigate the menu

There are four main menu items.
• See Information
• Start Bump Test
• Zero Sensors
• Start Calibration
1. Double pressing the button displays all four options on
the screen. See Information is selected and highlighted
by default.
2. Press the button to transition the selection to the next
choice.
3. Press and hold for three seconds to enter the selected
option.
4. Follow on-screen instructions for your selected
operation. Most of the detector's procedures are
described in this guide.

View detector's general information

1. Double-press the button to enter the main menu.
2. Select See information and press the button to scroll
through the following information:
Peak readings
STEL readings
TWA readings
Bump test intervals
Calibration
BLE information (BLE should be enabled)
LEL Correction Factor
Low Alam setpoint
Hi Alam setpoints
STEL setpoints
TWA setpoints

Reset TWA or STEL readings

Before you begin.
You need to enable TWA/STEL Reset in Fleet Manager II to
reset readings in the detector.
1. Go to the main menu and select > See Information >
TWA readings or STEL readings.
2. Press and hold for 3 seconds to reset readings.
A reset message is displayed.

Reset Peak readings

Before you begin.
You need to enable Peak Reset in Fleet Manager II to reset
readings in the detector.
1. Go to the main menu and select > See Information >
Peak readings.
2. Select Hold to reset peak readings. Press and hold for 3
seconds to reset readings.