Lumel PD15 User Manual

For n15, n15z, n17z digital meter series and p15, p16 transducers

Advertisement

PROGRAMMER
FOR N15, N15Z, N17Z
DIGITAL METER SERIES
and
P15, P16 TRANSDUCERS
USER'S MANUAL


Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the PD15 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Lumel PD15

  • Page 1 PROGRAMMER FOR N15, N15Z, N17Z DIGITAL METER SERIES P15, P16 TRANSDUCERS USER’S MANUAL...
  • Page 3: Table Of Contents

    P15, P16 TRANSDUCERS USER’S MANUAL CONTENTS Page 1. APPLICATION ................5 2. PROGRAMMER SET ..............5 3. REQUIREMENTS .................5 4. OPERATIONAL SAFETY .............5 5. SOFTWARE INSTALLATION ............6 6. PROGRAMMER CONNECTION TO THE METER ......6 7. DESCRIPTION OF THE PD15 PROGRAM .........9 7.1. Programming of N15 meter ..........10 7.2. Label of N15 meter .............. 11 7.3. Calibration of N15 meter .............12 7.4. Programming of N15Z and N17Z meters ......12 7.5. Label of N15Z and N17Z meters .........14 7.6. Calibration of N15Z and N17Z meters ........15 7.7. Programming of P15 transducer .........15 7.8. Label of P15 transducer ............16 7.9. Calibration of P15 transducer ..........18 7.10. P rogramming of P16 transducer .........18...
  • Page 5: Application

    1. APPLICATION The PD15 programmer is destined to program meters of N15 , N15Z, N17Z digital meter series, P15 and P16 transducers in the windows 9x/2000/NT/XP environ- ment. The programmer enables: zmianê parametrów mierników i przetworników,  kalibracjê mierników i przetworników.  2. PROGRAMMER SET The PD15 programmer set is composed of: - PD15 programmer 1 pc - Mini CD Diskette with software 1 pc - Adapter plate 1 pc - User’s manual 1 pc - Guarantee card 1 pc When unpacking the programmer, please check whether the type and execution code on the plate correspond to the order code. 3. REQUIREMENTS - Windows 9x/2000/NT/XP - Ca 2 MB of empty place on the disk - Minimum 4 MB RAM memory - LPT port with the possibility of working in the SPP, EPP or EPP/ECP mode 4. OPERATIONAL SAFETY Symbols included in this user’s manual mean: - Particularly important, one must acquaint with it before connecting the programmer. The non-observation of remarks determined by this symbol...
  • Page 6: Software Installation

    5. SOFTWARE INSTALLATION For a correct operation of the PD15 programmer, it is necessary to install the work of the LPT port in the SPP, EPP or EPP/ECP mode. Caution! In case of the LPT port working in the ECP mode or in case of some main plates in EPP/ECP mode, the PD15 programmer will not work correctly. One can modify the setting of the LPT port working mode in the computer BIOS, in the „Integrated peripherals” properties. In order to install the PD15 program, one must: 1. Introduce the installation mini CD diskette into the disk station, e.g. d:\ 2. Click the start key on the windows task bar and choose Start... 3. Write the access path, e.g. d:\setup.exe After starting the installation program, define the in-coming path, e.g. C:\Program Files\LZAE LUMEL S.A.\Programator PD15 and the working group, e.g.: LZAE LUMEL S.A. 6. DESCRIPTION OF THE PD15 PROGRAM The PD15 program realises following functions: - programming - where it is possible to reprogram meter/transducer parameters, - calibration (after giving the calibration password) - which enables the meter/ transducer calibration or its recalibration. Fig. 1. Installation window of the LPT port driver...
  • Page 7 The program start follows after clicking the “PD15” program icon in the “LZAE LUMEL SA” group. After starting, the program works in the edition mode. The main program window is shown on the fig. 2. Caution! In case of the program start for the first time, one must install the LPT “ DL Port I0” parallel port driver. For this aim, one must choose ”LPT …” -> “LPT Driver...” and next proceed according the description. The installation program window of the LPT port driver is shown on the Fig. 1. Fig. 2. Program main window...
  • Page 8 Program menu: - The language enables the work in one of the two language versions: Polish or English. - The “LPT “ program menu enables the selection of the LPT (LPT1, LPT2) parallel port and the installation of the LPT port driver (LPT Driver...). - Type device enables the selection of devices of series N15, N15Z, N17Z, P15, P16 or P17. The setting of programmable parameters or P17 calibration can be carried out after the password write. Readout and recording of parameters into the meter Keys “Read out” and “Write” serve for readout and record- ing of meter parameters. The “Write” key activates itself only when follows the readout of parameters from the meter after the first program start. It secures the meter against the discalibration and recording of er- roneous meter parameters. The correct readout of parameters from the meter is signalled by the message: “ Programming/readout favourably ended”. Caution! The operation failure of meter parameter readout is signalled by the mes- sage about error. See p.9 “Messages about errors” Calculator...
  • Page 9: Programmer Connection To The Meter

    In order to facilitate the user the recount of the resistance or the thermoelectric volt- age into temperature, the program is equipped with a simple recounting calculator for sensor used in N15 meters In case of calculations for thermocouples, the calculator also takes into consideration the sensor temperature of meter terminals. 7. PROGRAMMER CONNECTION TO THE METER The connection of the programmer with meters and the PC computer is presented on fig. 3 and 4. Before programming, one must remove the meter housing. Fig. 3. Programmer connection with N15, N15Z meters and the computer. Fig. 4. Programmer connection with N17Z meter and the computer.
  • Page 10: Programming Of N15 Meter

    W przetworniku P16 gniazdo programatora umieszczone jest wewnątrz obudowy. W celu przeprogramowania należy odkręcić 4 wkręty na pokrywie czołowej i zdjąć obudowę. Na płytce umieszczono 6 pól do podłączenia programatora PD15 z wy- korzystaniem dołączonej przejściówki RJ-piny. 7.1. Programming of N15 meter...
  • Page 11: Label Of N15 Meter

    Decimal point - Setting of the decimal point. Programming of the indication range The programming of the indication range gives the possibility of a quick and easy change of the displayed value on the display. On the base of given co-ordinates of two points by the user, the me- ter assigns a and b. (Y1=a . X1+b, Y2=a . X2+b). In order to reprogram the indi- cation range one must set the indication range programming...
  • Page 12: Calibration Of N15 Meter

    Field Nr 1. One must write the meter working range in, and the unit of the input channel. Field Nr 2. One must write the meter indication range. Field Nr 3. One must write the kind of compensation of the sensor work condition changes (only in the case of a temperature meter). In case of a manual compensation: - for resistance thermometers: in „„, e.g.: 15  (this is the total resistance of two wires connecting the sensor with the meter), - for thermocouples, value in „°C”, np.: 25°C (this is the temperature of terminals), In case of an automatic compensation one must write auto in. In this case the meter will compensate automatically: - in case of resistance thermometers - the wire resistance change, - in case of thermocouples - the terminal temperature change. 7.3. Calibrtion of the N15 meter The N15 calibration window is accessible after selecting the “calibration” tag and writing the calibration password (option accessible only in service workshops). Recovery of calibration coefficients In case of a meter erroneous calibration, there is the possibility to recover calibration coefficients written by the meter Manufacturer during the meter calibration. We can do this by pressing the Save… button. 7.4. Programming of N15Z and N17Z meters Parameter programming Selection of the meter working mode.
  • Page 13 Unit highlighting Switching on or off the highlighting of the unit in the meter N15Z – only for meters with a 5-digit display. N17Z – all versions In N15Z meters with a 4-digit display, the change of this parameter does not influence its work. Measurement time - Measurement averaging time. Decimal point - Setting of the decimal point. Programming of the indication range The programming of the indication range gives the possibility of a quick and easy change of the displayed value on the display. On the base of given co-ordinates of two points by the user, the me- ter assigns a and b. (Y1=a .
  • Page 14: Label Of N15Z And N17Z Meters

    7.5. Label of N15Z and N17Z meters In case of meter reprogramming, one must fill in the label, writing in their fields the appropriate work parameters of the meter. The label of the N15Z and N17Z meters are shown on Fig. 5. and fig. 6. Fig. 6. View of the N15Z meter label Fig. 7. View of the N17Z meter label Field Nr 1. One must write the meter working range in, and the unit of the input channel. Field Nr 2. One must write the meter indication range.
  • Page 15: Calibration Of N15Z And N17Z Meters

    7.6. Calibration of N15Z and N17Z meters The N15Z and N17Z calibration window is accessible after selecting the “calibration” tag. Recovery of calibration coefficients In case of a meter erroneous calibration, there is the pos- sibility to recover calibration coefficients written by the meter Manufacturer during the meter calibration. We can do this by pressing the Save… button. 7.7. Programming of the P15 transducer Parameter programming Input type – Kind of input depending on the version of the P15 transducer (measured Signal) Selection of the type input: J, K, N, E, mV - appears in the case of versions: P15-01X, P15- 02X, P15-03X, P15-04X, P15-09X. Selection of the type input: Tempera- ture, Resistance - appears in the case of versions: P15- 05X, P15-06X, P15-07X, P15-08X.
  • Page 16: Label Of P15 Transducer

    Automatic compensation: - In case of a resistance thermometer and resistance measurement, concerns the compensation of resistance changes of wires linking the sensor with the transducer. It requires a 3-wire line. - In case of a thermocouple, concerns the compensation of reference junction tem- perature changes. Conversion time – Averaging time, conversion. Programming of the indication range The programming of the indication range gives the possibility of a quick and easy change of the displayed value on the display. On the base...
  • Page 17 In case of an automatic compensation one must write auto in. In this case the meter will compensate automatically: - in case of resistance thermometers - the wire resistance change, - in case of thermocouples - the terminal temperature change of terminals. Fig. 8. View of the P15 transducer labels...
  • Page 18: Calibration Of P15 Transducer

    7.9. Calibration of P15 transducer The calibration window of P15 transducer is accessible after selecting the „Calibra- tion” tag and writing and acceptation the calibration password (option accessible only in authorized workshops). Recovery of calibration coefficients In case of a transducer erroneous calibration, there is the possibility to recover calibration coefficients written by the transducer Manufacturer during the transducer calibration. We can do this by pressing the Save… button. 7.10. Programming of the P16 transducer Parameter programming Output OC type – Accessible only in the case of a version with OC. OC type – allows to select: Normal, Enabled, Disabled, Manually ena- bled, Manually disabled. Input type - Temperature or humidity PrL OC – temperature or humidity lower threshold, depending on the selected input type. PrH OC – temperature or humidity upper threshold, depending on the selected input type.
  • Page 19: Label Of P16 Transducer

    In order to reprogram the conversion range, one must set to which values meas- ured on the input (X1, X2) have to correspond values displayed on the display (Y1, Y2). The programming of the Humidity conversion Range is accessible only in the ver- sion P16-W 7.11. Label of the P16 transducer Labels of the P16 transducer are shown on the Fig. 9. Fig. 9. View of the P16 transducer labels...
  • Page 20: C Alibration Of P16 Transducer

    7.12. Calibration of the P16 transducer The calibration window of the P16 transducer is accessible after the selection of the “Calibration” tag, write and acceptation of the correct calibration password (This option is accessible only in service workshops). Recovery of calibration coefficients In case of a transducer erroneous calibration, there is the possibility to recover calibration coefficients written by the transducer Manufacturer during the transducer calibration. We can do this by pressing the Save… button. 8. TECHNICAL DATA Galvanic isolation 4000 V d.c. Rated operation conditions: - supply voltage collect from the meter - ambient temperature -20…23…60 - storage temperature -20…+85 - air relative humidity <95% (condensation inadmissible) - working position Communication parameters: - data transmission through LPT link Assured protection degree IP 00 Dimensions...
  • Page 21: Messages About Errors

    9. MESSAGES ABOUT ERRORS In case of incorrect symptoms please acquaint with the following table Item Error message Message reason Solution The programmer This message appears One must check whether or the supply- in the case of meter the programmer is correctly ing leads are parameter readout connected to the meter or not connected. trial when the PD15 to the PC computer and The sequence programmer is not con- whether the meter is con- programming/ nected or the meter is nected to the network. readout did not not supplied. succeed. The verification During the parameter Renew the recording opera- did not succeed. recording into the meter, tion again and in case of the programmer found repeated error contact the out that the E prom manufacturer.
  • Page 22: Maintenance And Guarantee

    “Eeprom.hex: file being name and carry out the rea- processor or one in the folder in which dout and then the recording of the BLB fuses the PD15 program was of parameters into the meter installed. and in case of a repeated can not be re- corded. blunt of error contact the manufacturer.
  • Page 24 MEASUREMENT SALES PROGRAM DIGITAL and BARGRAPH PANEL METERS CONTROL MEASURING TRANSDUCERS RECORDING ANALOG PANEL METERS (DIN INSTRUMENTS) ANALOG and DIGITAL CLAMP-ON METERS pkt pkt INDUSTRIAL and HOUSEHOLD CONTROLLERS CHART AND PAPERLESS RECORDERS POWER CONTROL UNITS and INVERTERS WATT-HOUR METERS AUTOMOTIVE DASHBOARD INDICATORS ACCESSORIES FOR MEASURING INSTRUMENTS (SHUNTS) MEASURING SYSTEMS (ENERGY, HEAT, CONTROL) CUSTOM-MADE MEASURING ELECTRONIC DEVICES. WE ALSO OFFER OUR SERVICES IN THE PRODUCTION OF: ALUMINIUM ALLOY PRESSURE CASTINGS PRECISION ENGINEERING AND THERMOPLASTICS PARTS PRESSURE CASTING DIES AND OTHER TOOLS VARIOUS ELECTRONIC SUB-ASSEMBLIES (MSD TECHNOLOGY) QUALITY PROCEDURES: According to ISO 9001 and ISO 14001 international requirements. All our instruments have CE mark. For more information, please write to or phone our Export Lubuskie Zak³ady Aparatów Elektrycznych LUMEL S.A. ul. Sulechowska 1, 65-022 Zielona Góra, Poland Tel.: (48-68) 3295 100 (exchange) Fax: (48-68) 3295 101 e-mail:lumel@lumel.com.pl http://www.lumel.com.pl Export Department: Tel.: (48-68) 329 53 02 or 53 04 Fax: (48-68) 325 40 91 e-mail: export@lumel.com.pl...