10.0 Specification

10.1

Calibration

Vaporizers are calibrated at 21 ° C using an oxygen carrier gas at a flow of 5 litres/minute and they are
temperature, flow and pressure compensated within the specified operating range.
10.2

Resistance To Gas Flow With Vaporizer Out Of Circuit

Resistance
10 -15 cm wg
10.3

Liquid Capacity

The amount of liquid to fully charge a vaporizer with dry wicks = 300 millilitres (nominal).
The amount retained by the wick system = 75 millilitres (nominal).
5% Vaporizer
The amount of liquid to fill the vaporizer from the minimum to the maximum mark = 170 milliliters
(nominal).
8% Vaporizer (Sevotec)
The amount of liquid to fill the vaporizer from the minimum to the maximum mark = 137 milliliters
(nominal).
10.4

Weight And Dimensions

Weight
7.0 Kg
10.5

Environmental Conditions

Operating temperature Range = 18 ° C to 35 ° C
Storage temperature range = –40 ° C to +65° C
Note: The vaporizer packaging should be protected from condensation.
10.6

Performance

Accuracy at 5 litres/minute O
Page 30
August 1999
Vaporizer Setting
Flow rate
ON + 0%
5 litres/minute, Oxygen
Height
Depth
250 mm
164 mm
@ 21 ° C ± 2° C:
2
±0.25% of delivered agent or
Vaporizer 5%
±15% of dial setting (whichever is greater).
±0.4% of delivered agent or
Vaporizer 8%
±15% of dial setting (whichever is greater).
Temperature
21 ± 2° C
Width
110 mm
O & M Manual Part No. 1105-0100-000
Tec 5 Vaporizer