INSTRUCTIONS FOR USE
INSTRUMENT CLEANING AND STERILIZATION INSTRUCTIONS
WARNING: READ GENERAL WARNINGS
LIMITATIONS ON INSTRUMENT REUSE
Many of MicroAire's Instruments and Accessories are reusable; however, they have a limited life span. Prior to and after each use,
the instruments must be inspected where applicable for sharpness, wear, damage, proper cleaning, corrosion and integrity of the
connecting mechanisms. Particular care should be paid to rigid endoscope; these instruments are fragile and require special care.
1. DISASSEMBLY
1. Disconnect the light cable from the endoscope.
2. Carefully detach the endoscope/camera from the
and pushing against the
endoscope straight out, taking care not to bend the endoscope lens.
3. Detach the endoscope from the camera.
4. Loosen the blade lock screw. Pull up and hold the blade lock screw while pulling out the Disposable Blade Assembly.
5. Blade assembly should be properly disposed of in appropriate container. DO NOT re-sterilize the Blade Assembly – it is
designed for single use only.
2. PREPARATION FOR DECONTAMINATION
1. Excess tissue or debris should be wiped off to prevent it from drying onto the surface. Body fluids and tissue should not be
allowed to dry on instruments prior to cleaning.
2. Decontamination of reusable instruments or accessories should occur immediately after completion of the surgical procedure.
3. Blade assembly should be properly disposed of in appropriate container. DO NOT re-sterilize the Blade Assembly – it is
designed for single use only.
4. Disassemble endoscope from handpiece and disassemble light-post adapters from endoscope.
CLEANING OF INSTRUMENTATION
INSTRUMENTS: REF 81014, REF 81025, REF 81025A, REF 81151, REF 81152,
REF 81061, REF 81026, REF 81027, REF 81029, and REF 81040.
To avoid contact between devices utilize the MicroAire sterilization tray.
REF: 81040 (Ref: IMAGE 10)
CLEANING: MANUAL
1. Clean the
SmartRelease®
detergent, and a soft brush. Scrub the Handpiece with the brush, paying close attention to instrument crevices.
2. Scrub the endoscope and adapters with a soft brush in a solution of mild soap and water. Rinse and dry thoroughly.
CAUTION: Do not ultrasonically clean the endoscope or use solvents other than alcohol.
NOTE: Clean the lenses of the endoscope at the tip, camera-mount, and fiber-optic light ends with a cotton swab moistened in
isopropyl alcohol to assure a clear endoscopic view.
NOTE: Do not use hard brushes, abrasives or strong chemical solutions to clean the Handpiece or instruments.
3. Rinse all items thoroughly under running (< 50°C /122°F) water for a minimum of 2 minutes. If possible, use distilled water for
the final rinse.
™
SMARTRELEASE
INSTRUCTIONS FOR USE
SmartRelease®
Handpiece with the thumb and index finger until the endoscope is released. Pull the
SmartRelease®
Handpiece, Dilators, and Elevator thoroughly with warm (> 60°C / 140°F) water, pH-neutral enzymatic
IM-SMARTRELEASE REV E
per ISO 17664:2004 & AAMI ST 81:2004