Dimetix DLS-C 15 Technical Reference Manual

Distance laser sensor
Hide thumbs Also See for DLS-C 15:
Table of Contents

Advertisement

Distance Laser Sensor
DLS-C 15 / DLS-C 30
DLS-CH 15 / DLS-CH 30
FLS-C 10 / FLS-C 30
FLS-CH 10 /FLS-CH 30
Technical Reference Manual v2.03
print rev_
Distributed and represented exclusively in the USA by:
DIMETIXUSA
www.DIMETIX-USA.com
ionville, PA 19353
Tel: (484) 212 0636
Website: www.dimetix-usa.com
.
Email: info@dimetix-usa.com
- 1 -

Advertisement

Table of Contents
loading

Summary of Contents for Dimetix DLS-C 15

  • Page 1 Distance Laser Sensor DLS-C 15 / DLS-C 30 DLS-CH 15 / DLS-CH 30 FLS-C 10 / FLS-C 30 FLS-CH 10 /FLS-CH 30 Technical Reference Manual v2.03 print rev_ Distributed and represented exclusively in the USA by: DIMETIXUSA www.DIMETIX-USA.com ionville, PA 19353 Tel: (484) 212 0636 Website: www.dimetix-usa.com...
  • Page 3: Table Of Contents

    ECHNICAL EFERENCE ANUAL Table of Contents 1 Introduction........................................4 1.1 Product identification...................................5 1.2 Components......................................5 1.3 Validity.........................................5 1.4 Measuring range....................................6 1.5 Prevention of erroneous measurements..............................6 1.5.1 Rough surfaces...................................6 1.5.2 Transparent surfaces...................................6 1.5.3 Wet, smooth, or high-gloss surfaces............................6 1.5.4 Inclined, round surfaces................................6 1.5.5 Multiple reflections..................................6 2 Device setup........................................7 2.1 Connection......................................7 2.2 Controlled mode....................................8...
  • Page 4 ECHNICAL EFERENCE ANUAL 8.2.1 STOP/CLEAR command (sNc)..............................21 8.2.2 Distance measurement (sNg)..............................21 8.2.3 Signal measurement (sNm)...............................21 8.2.4 Temperature measurement (sNt)...............................21 8.2.5 Laser ON (sNo)..................................22 8.2.6 Laser OFF (sNp)..................................22 8.2.7 Single sensor Tracking (sNh)..............................23 8.2.8 Tracking with buffering – Start (sNf)............................23 8.2.9 Read out - Tracking with buffering(sNq)...........................23 8.3 Configuration commands...................................24 8.3.1 Set/Get communication parameter (sNbr)..........................24 8.3.2 Enhanced measuring Modes (sNuc)............................25...
  • Page 5: Introduction

    • IP65 (protected against ingress of dust and water) • 4 LEDs for status signaling • Complementary configuration software available at Dimetix web site (www.dimetix.com) • Optional: Internal heater for module operation down to -40°C • Laser class II (<0.95mW) •...
  • Page 6: Product Identification

    The product is identified by the label on the top of the sensor: DLS-C Version with a measurement rate of up to 6 Hz Typical Accuracy 1.5mm 3.0mm Standard version DLS-C 15 DLS-C 30 Part No.: 500622 Part No.: 500621 Extended temperature range DLS-CH 15 DLS-CH 30 Part No.: 500624...
  • Page 7: Measuring Range

    ECHNICAL EFERENCE ANUAL 1.4 Measuring range The DLS-C(H)/FLS-C(H) is an optical instrument, whose operation is influenced by environmental conditions. Therefore, the measuring range achieved in use may vary. The following conditions may influence the measuring range: Factors increasing range Factors reducing range Target surface Bright and reflective surfaces such as the target plates Matt and dark surfaces...
  • Page 8: Device Setup

    To be able to configure the DLS-C(H)/FLS-C(H), it must be powered and connected to a PC. Figure 2 shows the necessary connections. On the PC, any terminal program can be used to communicate with the module. A configuration utility is also available on the web page www.dimetix.com. DLS-C(H)
  • Page 9: Controlled Mode

    Host software is required for operation of the DLS-C(H)/FLS-C(H) in controlled mode. When connecting multiple devices to a single serial line (RS422), strict Master-Slave communication must be implemented (DLS-C(H)/FLS-C(H) operates as slave). For software sample-code or application-notes please consult our web site www.dimetix.com. Careful testing of the host software together with the devices prior to installation is strongly recommended.
  • Page 10: Automatic Mode

    ECHNICAL EFERENCE ANUAL 2.3 Automatic mode An automatic mode is provided for host-less operation of the DLS-C(H)/FLS-C(H). The analog and digital outputs are updated according to the configuration described below as soon as the unit is powered off and on. Analog Output The analog output is configurable and works with two ranges: –...
  • Page 11: Display Mode

    DLS-C(H) 14,15 FLS-C(H) RS232 or RS422 For a detailed description of this mode, please contact Dimetix or your dealer. 2.5 External Trigger The DLS-C(H)/FLS-C(H) includes the option of triggering measurements with an external switch or push button on Digital Input 1 (DI 1). Using the Digital Input DI 1 disables the Digital Output DO 1.
  • Page 12: Installation

    ECHNICAL EFERENCE ANUAL 3 Installation 3.1 Mounting Three M4 threaded holes in the bottom of the DLS-C(H)/FLS-C(H) make it easy to mount the device. Always obey all applicable safety regulations and never use the device outside the specifications stated under 4 Technical data on page 14. 3.2 Device wiring 3.2.1 Power Supply For trouble-free operation use a separate power supply for the DLS-C(H)/FLS-C(H).
  • Page 13: Controlled Mode

    ECHNICAL EFERENCE ANUAL 3.2.4 Controlled mode RS232 Only point-to-point communication is possible when using the RS232 interface. Never connect multiple DLS-C(H)/FLS-C(H)s on a RS232 serial line DLS-B(H) DLS-C(H) Host FLS-C(H) (PC or PLC) 9..30VDC 24..30VDC with Heating 14,15 Fig. 4 Point-to-point connection with RS232 RS422 It is possible to connect multiple devices on a single RS422 line.
  • Page 14: Automatic Mode

    ECHNICAL EFERENCE ANUAL 3.2.5 Automatic mode The analog interface of the DLS-C(H)/FLS-C(H) is isolated from the rest of the device. When using the analog interface, connect the analog ground (AGND). Make sure, that the total resistance in the analog path is lower than 500 Analog Input AGND 0..20mA...
  • Page 15: Technical Data

    ECHNICAL EFERENCE ANUAL 4 Technical data 4.1 Measuring accuracy The measuring accuracy corresponds to the ISO-recommendation ISO/R 1938-1971 with a σ, 99.7% statistical confidence level of 95.4% (i.e. ± twice the standard deviation refer to diagram on the right). The typical measuring accuracy relates to average conditions for measuring.
  • Page 16: Specifications

    ECHNICAL EFERENCE ANUAL 4.2 Specifications Typical measuring accuracy for σ DLS-C 15 / DLS-CH 15 ± 1.5 mm @ 2 σ DLS-C 30 / DLS-CH 30 ± 3.0 mm @ 2 1,3) σ FLS-C 10 / FLS-CH 10 ± 1.0 mm @ 2 1,3) σ...
  • Page 17: Electrical Components

    ECHNICAL EFERENCE ANUAL 5 Electrical components 5.1 ID switch This switch is used to set the module ID and can be set from 0 to 9. The default setting is 0. 5.2 Reset switch To reset the module to factory settings do the following: •...
  • Page 18: Connector

    ECHNICAL EFERENCE ANUAL 5.6 Connector 5.6.1 D-Sub connector Designator Description RS232 receive line RS232 send line RS422 send line negative SSI data output negative (Only for FLS-C and if configured for SSI) RS422 send line positive SSI data output positive (Only for FLS-C and if configured for SSI) RS422 receive line negative SSI clock input negative (Only for FLS-C and if configured for SSI) RS422 receive line positive...
  • Page 19: Physical Dimensions

    ECHNICAL EFERENCE ANUAL 6 Physical dimensions All dimensions in mm 88.5 38.5 3 x M4 x 5 150.0 M16 x 1.5 37.3 12.2 152.5 Laser Beam Measuring reference 80.0 Distance Laser Sensor Page 18/41...
  • Page 20: Factory Settings

    ECHNICAL EFERENCE ANUAL 7 Factory settings 7.1 Operation Mode: Controlled Mode 7.2 Communication parameters Baud: 19200 Data bit: Parity: Even Stop bit: (Setting 7) 7.3 Analog outputs Min output: Range min: Range max: Error output: 7.4 Module ID ID Number: 7.5 Digital output 1 (DOUT1) 2m + 5mm 2005mm...
  • Page 21: Command Set

    ECHNICAL EFERENCE ANUAL 8 Command set 8.1 General 8.1.1 Command termination <trm> All commands for the DLS-C(H)/FLS-C(H) are ASCII based and terminated <trm> with <cr><lf> 8.1.2 Module identification N Since the module can be addressed with the ID switch, the ID is represented in the commands by .
  • Page 22: Operation Commands

    ECHNICAL EFERENCE ANUAL 8.2 Operation commands 8.2.1 STOP/CLEAR command (sNc) Stops the current execution and resets the status LEDs as well as the digital outputs. Command Command sNc<trm> Return successful gN?<trm> Return Error gN@Ezzz<trm> Parameters Module ID Error code 8.2.2 Distance measurement (sNg) Triggers simple measurement of distance.
  • Page 23: Laser On (Sno)

    ECHNICAL EFERENCE ANUAL 8.2.5 Laser ON (sNo) Switches laser the beam ON for easy adjustment. Command Command sNo<trm> Return successful gN?<trm> Return Error gN@Ezzz<trm> Parameters Module ID Error code 8.2.6 Laser OFF (sNp) Switches the laser OFF. Command Command sNp<trm> Return successful gN?<trm>...
  • Page 24: Single Sensor Tracking (Snh)

    ECHNICAL EFERENCE ANUAL 8.2.7 Single sensor Tracking (sNh) Triggers continuous measurement of the distance. The measurements are made as fast as possible (Measurement speed depends on target conditions) This command is not to be used with more then one DLS-C(H)/FLS-C(H) on a RS-232/RS-422 line. The measurements are continued until the STOP/CLEAR command is issued.
  • Page 25: Configuration Commands

    ECHNICAL EFERENCE ANUAL 8.3 Configuration commands 8.3.1 Set/Get communication parameter (sNbr) Sets the communication parameters for the serial interface. This command saves all configuration parameters to Flash. The changed baud rate is activated after the next power on. Bold = default parameters (first use or after reset) Command Command sNbr+y<trm>...
  • Page 26: Enhanced Measuring Modes (Snuc)

    ECHNICAL EFERENCE ANUAL 8.3.2 Enhanced measuring Modes (sNuc) Multiple measuring modes available in the FLS-C measuring devices meet different requirements for various applications. With these measuring modes measuring rate and accuracy can be optimized for special requirements. The FLS-C measuring device has with factory settings an accuracy of ±1 mm (FLS-C(H) 10) respectively ±3 mm (FLS-C(H) 30) at 2σ. Measuring rate depends on environmental conditions such as target surface, distance, background light (e.g.
  • Page 27: Set Automatic Mode (Sna)

    ECHNICAL EFERENCE ANUAL 8.3.3 Set automatic mode (sNA) This command activates the automatic mode of the DLS-C(H)/FLS-C(H). It triggers continuous measurement of the distance. The analog, digital and SSI outputs are updated according to the measured distance values. The sampling time defines the measurement rate. If the sampling time is set to zero, the measurements are executed as fast as possible (depending on target conditions).
  • Page 28: Set/Get Distance Range (Snv)

    ECHNICAL EFERENCE ANUAL 8.3.6 Set/Get distance range (sNv) Sets the minimum and maximum distances corresponding to the minimum and maximum analog output current levels. 0...20mA 4...20mA DIST −D DIST −D Aout = ∗20mA Aout = ∗16mA4mA −D −D Aout Analog current output DIST Actual measured distance Distance programmed for the minimum output current...
  • Page 29: Set/Get Digital Output Levels (Snn)

    ECHNICAL EFERENCE ANUAL 8.3.7 Set/Get digital output levels (sNn) Sets the distance levels at which the digital outputs are switched ON and OFF with a hysteresis. Two different situations are possible: ON level > OFF level The ON level of the hysteresis is larger than the OFF level. With an increasing distance, the digital output is switched on (open drain output is closed) when the closed distance exceeds the ON level.
  • Page 30: Set Configuration Parameters To Factory Default (Snd)

    ECHNICAL EFERENCE ANUAL 8.3.9 Set configuration parameters to factory default (sNd) This command restores all configuration parameters to their factory default values. The parameters are written to the Flash Memory and therefore permanently saved. The communication parameters are also reset to factory settings. Command Command sNd<trm>...
  • Page 31: Digital Input

    ECHNICAL EFERENCE ANUAL 8.4 Digital Input The following commands configure the terminal DO1. This port can also be used as digital input. For safety reasons, always use a resistor to protect the connection terminal. DLS-C(H) FLS-C(H) DO/DI 1 8.4.1 Configure digital input (sNDI1) The digital output 1 of the DLS-C can also be used as digital input.
  • Page 32: Ssi (Synchronous Serial Interface)

    ECHNICAL EFERENCE ANUAL 8.5 SSI (Synchronous serial interface) The SSI Interface is only implemented in the FLS type. Connect the device by RS232 to do the configuration and use the following commands to activate the SSI interface. The application note “SSI Interface” contains all detailed information to operate the SSI Interface. 8.5.1 Change from RS422 to SSI output and back If the SSI interface should be used at the same port where normally the RS422 resides, it can be switched.
  • Page 33: Special User Commands

    ECHNICAL EFERENCE ANUAL 8.6 Special User Commands The special user commands are user configurable and are an extension to the standard commands. Use these commands carefully and only if you really understand their impact. The syntax of the user commands differ to the standard commands as follows: Standard Commands: sNxx User Commands:...
  • Page 34: User-Configured Single Sensor Tracking (Snuh)

    ECHNICAL EFERENCE ANUAL 8.6.3 User-configured single sensor tracking (sNuh) Triggers continuous measurement of the distance and outputs the result immediately to the serial interface. The distance measurement is executed as fast as possible (depending on target conditions). The measured distance is corrected with the user offset as set with the command 8.6.1 Set/Get user distance offset (sNuof).
  • Page 35: Read Out - User-Configured Tracking With Buffering (Snuq)

    ECHNICAL EFERENCE ANUAL 8.6.5 Read out – User-configured tracking with buffering (sNuq) After starting “User-configured fast tracking with buffering” with the command , the last measurement can be read out from the DLS- sNuf C(H)/FLS-C(H). This command takes the user distance offset into account Command Command sNuq<trm>...
  • Page 36: Error Codes

    It takes too long to measure according distance Received signal too weak @E255 (Use different target and distances, if the problem persists, please contact Dimetix) Received signal too strong @E256 (Use different target and distances, if the problem persists, please contact Dimetix)
  • Page 37: Safety Instructions

    Opening of the equipment, except to open the cover for access to the screw terminal Carrying out modification or conversion of the product Operation after failure in operation Use of accessories from other manufacturers without the express approval of Dimetix Aiming directly into the sun Deliberate dazzling of third parties; also in the dark 10) Inadequate safeguards at the surveying location (e.g.
  • Page 38: Areas Of Responsibility

    Responsibilities of the manufacturer of non-Dimetix accessories: The manufacturers of non-Dimetix accessories for the DLS-C(H)/FLS-C(H) are responsible for developing, implementing and communicating safety concepts for their products. They are also responsible for the effectiveness of these safety concepts in combination with the Dimetix equipment.
  • Page 39: Laser Classification

    ECHNICAL EFERENCE ANUAL Precautions: Only use this product as a measuring sensor, not as a control device. The system must be configured and operated in such a way that no damage will occur the event of an erroneous measurement, malfunction of the device or power failure due to installed safety measures (e.g. safety limit switch).
  • Page 40: Electromagnetic Compatibility (Emc)

    9.7 Producer Standards Dimetix hereby certifies that the product has been tested and complies with the specifications as stated in this 'Technical Reference Manual'. The test equipment used is in compliance with national and international standards. This is established by our Quality Management System.
  • Page 41: Labeling

    Clean the optical parts only with a clean soft glasses cleaning cloth and avoid scratching of the optical parts. 9.12 Service If you need to service the device, please contact Dimetix for instructions. CAUTION: The warranty is void if the device is opened except the cover of the screw terminal. Removing the label is also understood as opening.
  • Page 42 Customers should obtain the latest relevant information before placing orders and should verify that such information is up to date and complete. All products are sold subject to Dimetix terms and conditions of sale supplied at the time of order acknowledgment.