Page 1
ACIONAL OBADELA CelCulture CO Incubator ESCO CCL-170B-8 CelCulture Control System DECON Button & DECON LED To activate 90oC moist heat decontamination procedure During decontamination procedure, the yellow DECON LED will light up MENU Button When the main LCD shows normal display – MENU button will activate the menu options ...
Page 2
Edição: 01 Á ABORATÓRIO DE NSAIOS ECNOLÓGICOS EM REAS IMPAS Revisão: 00 Data: 2014 10, 2695-066 B STRADA ACIONAL OBADELA HEAT LED HEAT LED will light up whenever a heating process is activated INJECT LED INJECT LED will light up whenever a gas injection process is activated oC/%RH Display ...
Page 3
Edição: 01 Á ABORATÓRIO DE NSAIOS ECNOLÓGICOS EM REAS IMPAS Revisão: 00 Data: 2014 10, 2695-066 B STRADA ACIONAL OBADELA voltmeter. If voltage is not within +/-2% of the socket voltage, replace the power cord. Check the fuse of the power cord. CO2 level is below or above CO2 set-point Possible Cause Corrective Actions...
Page 4
Edição: 01 Á ABORATÓRIO DE NSAIOS ECNOLÓGICOS EM REAS IMPAS Revisão: 00 Data: 2014 10, 2695-066 B STRADA ACIONAL OBADELA Inner door opened for an extended period of Close the door and let the unit stabilize. time Access port stopper at the rear of cabinet Ensure access port stopper and filter are was removed or not install installed.
Edição: 01 Á ABORATÓRIO DE NSAIOS ECNOLÓGICOS EM REAS IMPAS Revisão: 00 Data: 2014 10, 2695-066 B STRADA ACIONAL OBADELA Air leakage through Inner door gasket. Check installation of the gasket ensure it is properly secure to the cabinet and that there is no gap.
Page 6
5. Re-assemble the unit by reversing s Decontamination Decontamination in CelCulture uses 90°C moist heat to kill common organisms that contaminates workspace including all installed components. The entire program run of the decontamination routine takes approximately 15 hours (overnight process. The chamber is clean...