Measuring accuracy (according to EN 60770-2)
Zero point and measuring
range
Long-term stability
Zero point and measuring
range
Effect of auxiliary power supply
Rated conditions
Installation conditions
Ambient conditions
Ambient temperature
Altitude
Relative humidity
Note
Storage temperature
Degree of protection
according to EN 60529
Electromagnetic
compatibility
Interference emission and
interference immunity
Process medium conditions
Process medium
temperature
Pressure transmitter construction
Weight
Pressure transmitter
Cable
Material
Wetted parts materials
Housing
Sensor
Cable
Protective cover
Sealing material
Electrical connection
Torque for cable gland nut
made of
Cable box 7MF1575-8AA construction (accessory)
Field of application
Weight
Electrical connection
SITRANS LH300 (7MF1575..) Transmitter for hydrostatic level
A5E37897183-AA, 09/2016
0.05% of measuring range per 10K
0.15% of full-scale value/year
0.01% per 1 V voltage change
-10 ... +80 °C (-4 ... +176 °F)
Max. 2 000 m above sea level
0 to 100%
Observe the temperature class in hazardous areas.
-20 to +80 °C (-4 to +176 °F)
IP68
To EN 61326-1 and EN 61326-2-3
-10 ... +80 °C (-4 ... +176 °F)
Approx. 0.4 kg (0.88 lb)
0.08 kg/m (approx. 0.059 lb/ft)
Stainless steel material No. 1.4404/316L or 1.4539/904L
Ceramic AI
O
(99.6%)
2
3
PE-HD, FEP
PPE, ETFE, stainless steel (1.4404/316L or 1.4539/904L)
FPM, EPDM (for drinking water)
Lengths: 3, 5, 10, 15, 20, 30, 40, 50, ..., 1000 m
The maximum permissible cable length for devices with type of protection "Intrinsic
safety Ex i" is 300 m.
Plastic
2.5 Nm (1.8 ft lb)
For connecting the transmitter cable
0.2 kg (0.44 lb)
2 x 3-way (cable diameter from 1 mm
to 1.5 mm
)
2
2
17