T Y P E 1398-A PULSE G E N E R A T O R
Overshoot and ringing.
Not greater than 10% of
amplitude.
5
10%
Jitter-oscilloscope limitation.
The amount of jitter
present in the oscilloscope must be taken into account
when duration and period jitter are measured. Typical
. .
jitter for most delaying sweep oscilloscopes is 0.02%
of the amount of delay. T o determine a c t u a l jitter, con-
nect
a
s t a b l e time-mark signal t o the oscilloscope,
set
the oscilloscope for the amount of delay required,
and measure any jitter that may be present. If the time-
mark signal i s s t a b l e , any jitter present i s due to the
oscilloscope.
Duration jitter.
L e s s than or equal to 0.04%. Set
the Type 1398-A controls a s follows:
A .
. . . . . . . . . . . . . . . . . . . . . . . . . .
centered
P R F
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3 0 c / s
PULSE DURATION d i a l
. . . . . . . . . . . . . . . .
1 0
PULSE DURATION RANGE.
. . . . .
1 t o 1 0 ms
. . . . . . . . .
AMPLITUDE
. 1 0 (fully clockwise
)
input of the oscilloscope via a 10-megohm, 12-pF probe.
Set the Type 1398-A controls a s follows:
AF
. . . . . . . . . . . . . . . .
HIGH ( f u l l y clockwise
)
. . . . . . . . . . . . . . . . . . . . . . . . . . . .
P R F
1.2 Mc/s
. . . . . . . . . . . . . . . .
PULSE DURATION d i a l
..5
. . . .
PULSE DURATION RANGE.
.O.1 to 1 p s
. . . . . . . . .
AMPLITUDE
. 1 0 (fully clockwise
)
+SYNC output.
The characteristics and specifi-
cations for both the + and the
-
SYNC outputs are the
same, except f or polarity. Amplitude: Equal to or greater
than 8 v o l t s , peak-to-peak. Duration: Equal to or l e s s
than 0.2
ps.
D E L A Y E D SYNC output.
Amplitude: Equal to or
greaterthan g v o l t s , peak-to-peak. Duration: Equal t o or
l e s s than 0.3
ps.
DELAYING SWEEP
INTENSIFIED BY
,
5.8.6 EXTERNAL DRIVE.
PRF
jitter.
L e s s than or equal t o 0.04%. Set the
Type 1398-A controls a s follows:
L F . .
. . . . . . . . . . . . . .
HIGH (fully c l o c k w i s e )
P R F .
. . . . . . . . . . . . . . . . . . . . . . . . . . . .
100 c / s
. . . . . . . . . . . . . . . .
P U I S E DURATION dial.
.1
. . . .
PULSE DURATION RANGE.
0.1 t o 1
ps
Setup.
Connect a sine-wave generator t o the EXT
DRIVE binding posts. Set the generator for an output
of l 0 k c / s , 0.1V, rms. Set the Type 1398-A controls a s
follows:
. . . . . . . . . . . . . . . . . . . . . . . . . . . .
A F .
.centered
. . . . . . . . . . . . . . . . . . . . . . . .
P R F .
EXT DRIVE
. . . . . . . . . . . . . . . .
PULSE DURATION d i a l .
-1
. . .
PULSE DURATION RANGE.
.0.1 t o 1
p s
. . . . . . . . .
AMPLITUDE.
10 ( f u l l y clockwise
)
AMPLITUDE
1 0 (fully clockwise
)
R107
adjustment.
Connect a 10-megohm probe from
the oscilloscope to V102, pin 1. Adjust R107 for a sym-
loms4--.
metrical square wave.
DELAYING SWEEP
_TV"\_
DELAYED
SIFIED SWEEP
BY
2.2-Mc
sensitivity.
Set the generator for an output
of' 2.2 Mc/s, 0.5
V,
rms. Set the AF control for a trigger-
'.--'
ed oscilloscope display. After the
AF
control has been
/ r 7
s e t for a triggered display, it must b e within 30 degrees
,
I
of center. T o check i f the display i s actually triggered,
-
I
disconnect the sine-wave generator; if the display was
JITTER'
5
0.04%
( 4 ~ s )
triggered, the display will disappear. Reconnect the
5.8.5 SYNC-OUTPUT CHARACTERISTICS.
generator.
Setup.
Connect the OUTPUT PULSE binding p o s t s
Trigger pulling.
Vary the PULSE DURATION dial
to the external trigger input of .the oscilloscope and con-
from 1 to
6;
the pulse repetition frequency must remain
nect the appropriate SYNC binding post to the vertical
2-2 Mc/s.