Model 4999-32Vts - Geokon 4999 Instruction Manual

Terminal box
Table of Contents

Advertisement

Connect the patch cord (flying leads) for a GK-403 or GK-404 as follows:
Note: To read positions 9 through 16 the first switch (1-8) must be set to "9-16."
To read positions 17 through 24 the first switch must be set to "17-24."
To read positions 25 through 32 the first switch must be set to "25-32."
Terminal Board Wiring: There are two terminal
boards mounted to the lower plate. Wire the first
16 gages and thermistors to the Board 1 as
indicated in the drawing. Wire gages and
thermistors 17 through 32 to the Board 2.
Attach the gage cable wires to the terminal strips in the order below:
Gage
Gage
Label
Wire
Number
Color
1H
Red
1L
Black
2H
1 or 17
White
2L
Green
S
Shield
3H
Red
3L
Black
4H
2 or 18
White
4L
Green
S
Shield
5H
Red
5L
Black
6H
3 or 19
White
6L
Green
S
Shield
7H
Red
7L
Black
8H
4 or 20
White
8L
Green
S
Shield

Model 4999-32VTS

32 Vibrating Wire Gages with 32 Thermistors
Boot Color Switch Plate Post Label
Red
Black
White
Green
Blue
Gage
Label
Number
9H
9L
10H
5 or 21
10L
S
11H
11L
12H
6 or 22
12L
S
13H
13L
14H
7 or 23
14L
S
15H
15L
16H
8 or 24
16L
S
Front Panel
Gage +
Gage -
Thermistor
Thermistor
Cable Shield
Gage
Gage
Wire
Label
Number
Color
Red
17H
Black
17L
White
18H
9 or 25
Green
18L
Shield
S
Red
19H
Black
19L
White
20H
10 or 26
Green
20L
Shield
S
Red
21H
Black
21L
White
22H
11 or 27
Green
22L
Shield
S
Red
23H
Black
23L
White
24H
12 or 28
Green
24L
Shield
S
17-32
Board 2
1-16
Board 1
Gage
Gage
Wire
Label
Number
Color
Red
25H
Black
25L
White
26H
13 or 29
Green
26L
Shield
S
Red
27H
Black
27L
White
28H
14 or 30
Green
28L
Shield
S
Red
29H
Black
29L
White
30H
15 or 31
Green
30L
Shield
S
Red
31H
Black
31L
White
32H
16 or 32
Green
32L
Shield
S
7
Gage
Wire
Color
Red
Black
White
Green
Shield
Red
Black
White
Green
Shield
Red
Black
White
Green
Shield
Red
Black
White
Green
Shield

Advertisement

Table of Contents
loading