Installation Instructions
for the
Manufacturer
of the End Product
DUOMAT 7
Drive System
CARE L - HOSP L
Issue
10/2007

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the duomat 7 care l and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Dewert duomat 7 care l

  • Page 1 Installation Instructions for the Manufacturer of the End Product DUOMAT 7 Drive System CARE L - HOSP L Issue 10/2007...
  • Page 2: Table Of Contents

    DEWERT or ... the use of spare parts not manufactured or approved by DEWERT - these may not ensure adequate safety! Due to the policy of ongoing product improvement, DEWERT reserves the right to carry out...
  • Page 3: Designated Use

    … for the disabled (CARE L version), for the hospital sector (HOSP L version), for the motorized adjustment of moveable bed parts using original DEWERT fittings and accessories. DUOMAT 7 drive system is not intended for use...
  • Page 4 . For this purpose, in the case of the HOSP L and CARE L drive systems, DEWERT has additionally applied, fully or partially, a number of standards from the medical products sector, in order to facilitate use in medical products, e.g.
  • Page 5: Getting To Know The System

    Getting to Know the System DUOMAT 7 drive system is intended for the German market and complies with the Law applicable in Germany in implementation of relevant EC Directives. Product versions The drive system is supplied ex factory for the power-adjustment of the head and foot ends of a bed DUOMAT 7 ).
  • Page 6 DUOMAT 7 CARE L/HOSP L Technical data Power supply..........: 100 / 110 / 115 / 120 / 230 / 240 VAC 50-60 Hz Power consumption with rated load.... : 0.63 A AC - 3.15 A AC (depending on mains voltage and load) Permissible compressive force....:...
  • Page 7: Fitting

    Fitting Caution! Only ever connect or disconnect electrical components when they are voltage-free. Included in the supply package is the drive depending on the order a handset locking device or a supervisor and one or two slave drives. The components are prewired and receive plugs/ready to plug in.
  • Page 8 DUOMAT 7 CARE L/HOSP L Electrical connection In the Operating Instructions to be issued by you, point out to the operator that if leads, in particular the mains connection lead , are driven over they could sustain damage. Mechanical loads should also be avoided.
  • Page 9 Dismantling Loosen the screws, remove the cap and then the leads from the respective plug sockets. Drive the application in basic position, support the drive before opening the closer - once these are open, the drive is released immediately! Make sure that the closer re-engage so that they cannot get lost.
  • Page 10: Operation

    DUOMAT 7 CARE L/HOSP L Operation For drawing up the Operating Instructions for the end product, you can use the specialist information described herein. Please bear in mind that these instructions are intended for you as a specialist and not for the possibly nonprofessional operator of the end product.
  • Page 11 Emergency Release “ER” (optional) If you have acquired the DUOMAT 7 with Emergency Release , please note the following: Aided by the “Emergency Release” you can mechanically lower the backrest / legrest in an emergency. You must first relieve the backrest / legrest, e.g. by lifting the correctly lever. With the backrest / legrest in the no-load state, operate the “Emergency Release”...
  • Page 12 DUOMAT 7 CARE L/HOSP L Handset (example) ® The IPROXX handset is also offered with an integral blocking device via a magnetic lock ® and is then called IPROXX SE. This can be equipped with up to 10 adjustment keys. The...
  • Page 13 Locking device (Example) Blocking switch blocked released Backrest blocked Foot rest blocked Stroke motors blocked Example of a control box with 3 blocking switches and 2 adjustment keys Adjustment keys for the Trendelenburg- Function Only permissible for DUOMAT 7 HOSP L To guarantee first error security, you can block the drive movements via the locking device: Turn the blocking switch...
  • Page 14 DUOMAT 7 CARE L/HOSP L Supervisor (Example) Blocking switch Adjustment keys Release key Example of a Supervisor with 3 blocking buttons and 3 adjustment functions Trendelenburg The supervisor control system gives the hospital/care staff over the DUOMAT 7 drive system in the CARE L and HOSP L versions.
  • Page 15: Maintenance And Repairs

    Maintenance and Repairs At regular intervals carry out the inspections in accordance with the BGV A3. The inspections must be performed by an electrical specialist. The recommended inspection period is: 6 months In addition to the above, the following checks should be carried out at shorter intervals: Regular visual checks for damage of all kinds Check the housing for cracks and fractures and the mains connection lead for signs of pinching and shearing-off.
  • Page 16 3 or by persons who have taken part and successfully completed the corresponding training programs offered by DEWERT. After correctly changing the mains connection lead, check for safety and proper function. Disconnect the drive from the mains (pull out the mains plug).
  • Page 17: Type Plate And Seal

    Graphical Symbols Conformity marks Protection classification II In order to guarantee the safety of Application part of Type B DEWERT products, a seal is attached IP44 to all DEWERT products. Opening the Protection category product damages the seal, thereby indicating that the drive has been Do not dispose of with the normal altered or tampered with.
  • Page 18: Design Of The Duomat 7 Drive System

    DUOMAT 7 CARE L/HOSP L Design of the DUOMAT 7 CARE L/HOSP L Drive System (Example) (Figure: DUOMAT 7 as QUADROMAT 7 with 2 slave drives) Only connect the components as shown! This can otherwise result in demage to the drive control unit! Pos.
  • Page 19: Trouble-Shooter's Guide

    Trouble-shooter’s Guide to Detect and Eliminate Faults/Errors The following table has been developed to help you detect and eliminate common faults and errors. If you come across a fault/error that is not listed here, please contact your supplier. All of these faults/errors may only be inspected and rectified by specialists holding the qualifications as described on page 3.
  • Page 20: Cleaning And Care

    DUOMAT 7 CARE L/HOSP L Cleaning DUOMAT 7 drive system has been designed to facilitate cleaning for the user, and this has been made even easier thanks to the large number of flat surfaces. DUOMAT 7 drive system should be cleaned with a proprietary cleaning agent suitable for polyamid 6 using a damp cloth.
  • Page 21: Manufacturer's Declaration, Declarations Of Conformity, Additional Information

    EC Manufacturer’s Declaration According to Appendix II B of the EC Machines Directive (98/37/EG) The Manufacturer: DEWERT Antriebs- und Systemtechnik GmbH Weststr. 1 32278 Kirchlengern Germany hereby declares that the drive system described below DUOMAT 7 CARE L/HOSP L is not a ready-to-use machine in keeping with the EC Machines Directive and, therefore,...
  • Page 22 DUOMAT 7 CARE L/HOSP L EC Declarations EC Declaration of Conformity according to appendix IV of the EC Directive on Electromagnetic Compatibility 2004/108/EEC, according to appendix III of EC Low-Voltage Directive 73/23/EWG The Manufacturer: DEWERT Antriebs- und Systemtechnik GmbH Weststr. 1...
  • Page 23 Additional Information DUOMAT 7 drive system (see system overview) For the CARE L und HOSP L versions the following standards were applied Based on EN 60601-1:1990 +A1:1993 +A2:1995, Electromedical Equipment EN60601-1, main section 2 Environmental Conditions EN60601-1, main section Protection against Electrical Shock Hazard EN60601-1, section Mechanical Strength...
  • Page 24 DUOMAT 7 CARE L/HOSP L Additional Information Duomat 7 drive system (cont.) For the HOSP L version the following Standards were applied Based on EN60601-2-38:1996 +A1:2000, Power-operatedHospital Beds EN60601-2-38, section 5.2 Classification of application part EN60601-2-38, section Protection category of the system: >= IPX4...
  • Page 25 DEWERT Antriebs- und Systemtechnik GmbH Weststraße 1 32278 Kirchlengern Germany Phone:+49(0)5223/979-0 Fax: +49(0) 5223/75182 http://www.dewert.de Info@dewert.de ID No. 43698...