Advertisement

Quick Links

User Manual
2800XL and 4000E series model
Dear customer,
Congratulations on purchasing a product of outstanding quality.
Use of the best materials from renowned suppliers guarantees years of trouble-free operation, provided the
device is handled correctly and as intended in accordance with the conditions described in the user manual.
In the unlikely event you need to make a claim, please contact us.
We welcome suggestions from the users of our products.
Contents
1.
Safety information .................................................................................................................................... 2
1.1
Applied symbols .............................................................................................................................. 2
1.2
Applied standards ............................................................................................................................ 2
1.3
Safety information ............................................................................................................................ 2
1.4
Intended purpose ............................................................................................................................. 2
1.5
Information on setup and use .......................................................................................................... 3
1.6
Commissioning ................................................................................................................................ 3
1.7
Safety notices .................................................................................................................................. 3
1.8
Model designation and type labelling .............................................................................................. 4
1.9
Meaning of the serial number .......................................................................................................... 4
2.
Operating manual .................................................................................................................................... 5
2.1
Table design .................................................................................................................................... 5
2.2
Height adjustment ............................................................................................................................ 5
2.3
Head part adjustment ...................................................................................................................... 5
2.4
Adjustment of other sections ........................................................................................................... 6
2.5
Movability (model dependent or optional) ....................................................................................... 7
2.6
Additional equipment ....................................................................................................................... 8
3.
Additional accessories (for user-specific table configuration), in extracts ............................................... 9
4.
Technical data ....................................................................................................................................... 10
4.1
Individual models ........................................................................................................................... 10
4.2
Technical data for electric motor ................................................................................................... 13
4.3
Technical data for hydraulic system .............................................................................................. 13
5.
Cleaning instructions ............................................................................................................................. 13
6.
Maintenance and technical inspection .................................................................................................. 14
7.
Safety devices ....................................................................................................................................... 15
8.
Reporting obligation ............................................................................................................................... 16
9.
Disposal ................................................................................................................................................. 16
10. Declaration of Conformity ...................................................................................................................... 17
K.H. DEWERT GmbH
Vollmestr. 7
33649 Bielefeld
Germany
Tel. +49 / 521 400 27-0
Fax +49 / 521 400 27-27
info@khdewert.de
www.khdewert.de
Dewert User Manual 2800XL and 4000E Series - Updated 05.2021/1 MDR
Page 1 of 19

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 2800XL Series and is the answer not in the manual?

Questions and answers

Summary of Contents for Dewert 2800XL Series

  • Page 1: Table Of Contents

    10. Declaration of Conformity ........................17 K.H. DEWERT GmbH Vollmestr. 7 33649 Bielefeld Germany Tel. +49 / 521 400 27-0 Fax +49 / 521 400 27-27 info@khdewert.de www.khdewert.de Dewert User Manual 2800XL and 4000E Series – Updated 05.2021/1 MDR Page 1 of 19...
  • Page 2: Safety Information

    The table is used for the ideal positioning of patients for the purpose of curative and disease treatment, examination, massage and health therapy. Table operation and patient positioning on the table may only be Dewert User Manual 2800XL and 4000E Series – Updated 05.2021/1 MDR Page 2 of 19...
  • Page 3: Information On Setup And Use

    Important for the user: When adjusting the upholstery parts, do not reach under the spacers located beneath the upholstery parts. Dewert User Manual 2800XL and 4000E Series – Updated 05.2021/1 MDR Page 3 of 19...
  • Page 4: Model Designation And Type Labelling

    It enables us to identify this table model and trace back assemblies/safety-relevant components at any time. Please always state this serial number when enquiring about spare parts. Dewert User Manual 2800XL and 4000E Series – Updated 05.2021/1 MDR Page 4 of 19...
  • Page 5: Operating Manual

    The head part is adjusted using a gas spring. To operate, press the release lever, which is located at the end of the head part underneath the upholstery part, in the direction of the upholstered surface. The head part is Dewert User Manual 2800XL and 4000E Series – Updated 05.2021/1 MDR Page 5 of 19...
  • Page 6: Adjustment Of Other Sections

    Model 4840E: Alternatively, the roof position can be steplessly adjusted directly using two gas springs. To operate, press the release lever in the direction of the upholstery. The upholstery part is lifted slowly to the Dewert User Manual 2800XL and 4000E Series – Updated 05.2021/1 MDR Page 6 of 19...
  • Page 7: Movability (Model Dependent Or Optional)

    With this movability option, access to the patient is improved during examination and treatment thanks to easy positioning within the room. Transporting patients is not considered as intended use. Dewert User Manual 2800XL and 4000E Series – Updated 05.2021/1 MDR Page 7 of 19...
  • Page 8: Additional Equipment

    (an extension bar is optionally available). The inserted extension bar is fixed with a knurled screw. Please make sure that the extension bar is not pulled out too far and that it is securely clamped. Dewert User Manual 2800XL and 4000E Series – Updated 05.2021/1 MDR Page 8 of 19...
  • Page 9: Additional Accessories (For User-Specific Table Configuration), In Extracts

    1x pair of attachment clamps • Belt pad • Safety belt • Armpit and pelvic fixation support • Fixation point, upholstered • Shoulder fixation supports with holder Dewert User Manual 2800XL and 4000E Series – Updated 05.2021/1 MDR Page 9 of 19...
  • Page 10: Technical Data

    Foot part adjustment range +30° / -90° +30° / -90° Leg part adjustment range Adjustment range, inclined position Max. patient weight (kg) 250*** 250*** 250*** 250*** Dewert User Manual 2800XL and 4000E Series – Updated 05.2021/1 MDR Page 10 of 19...
  • Page 11 +20° / -45° Adjustment range, inclined position 0° / +20° 0° / -20° 0° / -20° 0° / -20° Max. patient weight (kg) 250* 250* 250* 250* Dewert User Manual 2800XL and 4000E Series – Updated 05.2021/1 MDR Page 11 of 19...
  • Page 12 (DC: 25 s/400 s) is exceeded. The table should be ready for use again after a 15-minute rest period. There is, therefore, no risk of overloading the motor. Dewert User Manual 2800XL and 4000E Series – Updated 05.2021/1 MDR Page 12 of 19...
  • Page 13: Technical Data For Electric Motor

    For cleaning and disinfection, a selection of agents from various manufacturers has been tested for compatibility. Please refer to the separate enclosed sheet. The table is equipped as standard with the Dewert hygiene standards, which enable optimal cleaning and disinfection: •...
  • Page 14: Maintenance And Technical Inspection

    Lubricate the roller guides with a little bearing grease or similar using a brush, if necessary. Insufficient lubrication is noticeable through noise development. A checklist for maintenance/technical inspection can be found in the appendix. Dewert User Manual 2800XL and 4000E Series – Updated 05.2021/1 MDR Page 14 of 19...
  • Page 15: Safety Devices

    In addition, the table can be equipped with another option: Foot switch cover The top of the foot switch is fitted with a protective cover. This makes unintentional or unauthorised operation more difficult. Dewert User Manual 2800XL and 4000E Series – Updated 05.2021/1 MDR Page 15 of 19...
  • Page 16: Reporting Obligation

    8. Reporting obligation All serious incidents occurring in connection with the product shall be reported to the manufacturer (K.H. Dewert GmbH) and to the competent authority of the Member State in which the user and/or the patient resides. Member State...
  • Page 17: Declaration Of Conformity

    We hereby declare conformity with the aforesaid directive. As the manufacturer, we bear sole responsibility for issuing this EU Declaration of Conformity Mark: Bielefeld/Germany, 26 May 2021 K.-H. DEWERT GmbH Management Dewert User Manual 2800XL and 4000E Series – Updated 05.2021/1 MDR Page 17 of 19...
  • Page 18 Checklist for maintenance/technical inspection to IEC 62353 and DGUV Regulation 3 Device Model designation Serial no. Location Responsible person Date, inspector Actuator designation Inspections FAIL Description of defects Visual inspection Overall impression of the table OK? Labels, CE mark, type plate present? Manufacturer’s operating manual available and accessible? Sufficient space available when carrying out all adjustment functions? Mechanical structure undamaged:...
  • Page 19 FAIL Description of defects Height retained? Hydraulic pump leaking? Smooth lowering possible? Check by operating the pedals to lower the table Strong noise development? Wipe the piston rod with a cloth Lying surface adjustment functions: Metal ratchets - arrester - gas spring Metal ratchet inspection: Lifting the lying surface segment: Do the two metal ratchets engage properly and securely? Do they engage evenly?

Table of Contents