Troubleshooting Guide - NAUSICAA BLUE WAYUP PLUS User Manual

Stand-up lift with electrical opening legs
Table of Contents

Advertisement

Problem
The product, without patient, does
not go down.
Mobile parts are too hard, pro-
duct is difficult to handle.
The actuator does not work but
the user can hear a "click" in the
control box when the remote
control button is pushed.
The actuator does not work and
the user does not hear a "click" in
the control box when the remote
control button is pushed.
The actuator stops when lifting.
The actuator does not work at all
22
Manufactured by NAUSICAA Médical S.A.S. / Approved by Ghizlane Labrosse (Biomedical Engineer)

Troubleshooting Guide

Our products need some
Mobile parts need to be greased.
1 . Batteries are unloaded.
2 . The actuator jack is not plugged.
3 . The actuator jack is deteriorated.
4 . There are damages on the
actuator or on the control box.
1 . Emergency button is activated.
2 . There is no contact between
the battery and the control box.
3 . Batteries are unloaded.
4 . The remote control jack
5 . The remote control jack
6 . There are damages on the
remote control or on the control
1 . Batteries are unloaded
(sound system + "low battery"
written on the screen).
2 . The product has been
3 . The weight lifted is too heavy.
1 . The control box is broken down.
2 . The remote control is broken
User Manual - Stand-Up Lift: BLUE WAYUP PLUS with electrical opening legs
Origin
weight to go down.
is not plugged.
is deteriorated.
box.
used for too long.
down.
Solutions
Push down on the lifting arm while
pushing the lowering button on
the remote control.
Grease the mobile parts.
1 . Load the device.
2 . Plug the actuator jack.
3 . The actuator has to be repaired.
4 . The entire electric system has
to be checked.
1 . Turn the button to
unlock the device.
2 . Take off the battery
and put it back.
3 . Check the charging cord.
4 . Plug the remote control.
5 . Change the remote control.
6 . The entire electric system has to
be checked.
1 . Check the charging cord
then load battery.
2 . Let the device rest a
moment (as a protection
for the actuator lifespan).
3 . Reduce the weight lifted.
1 . The control box has to be changed.
2 . The remote control has to
be changed.

Advertisement