NAUSICAA BLUE WAYUP PLUS User Manual page 17

Stand-up lift with electrical opening legs
Table of Contents

Advertisement

Preventive Maintenance & Security Controls
Control of security points (Frequency: depending on use, recommended every three months)
This is a sight control : any wear trace must lead to a replacement.
1
Connection head of the actuator (bolted)
2
Connection base of the actuator (bolted)
3
Connection head of the mast (bolted)
4
Locking base of the mast
5
Fixing of the castors (bolted)
6
Sliding of the sub-patellar support (adjustable knob)
7
State and sticking of the anti-slip material
8
Fixing of the mast handle
9
OVERALL STATE OF THE DEVICE (WELDS, JOINTS, OXIDATION)
THESE CONTROLS CAN BE CARRIED OUT WITHOUT LOAD
Periodical Greasing
WARNING :
DO NOT USE WITHOUT
SHIN BLOCK
5
User Manual - Stand-Up Lift: BLUE WAYUP PLUS with electrical opening legs
Manufactured by NAUSICAA Médical S.A.S. / Approved by Ghizlane Labrosse (Biomedical Engineer)
7
3
2
6
1
4
5